Specification builder · Tier C

Intensive care ventilator for adult and paediatric patients (invasive and non-invasive), with optional neonatal capability

A critical-care ventilator breathes for, or supports the breathing of, patients who cannot breathe adequately on their own, through a breathing tube or a mask. It is bought by public and private hospitals for ICUs, HDUs and emergency departments. A good tender fixes the patient range, modes, gas supply (turbine or wall air), battery time, monitoring and the service backing for 7-10 years.

What this equipment will do

  • Patients: adults and children, ventilated invasively (breathing tube or tracheostomy) or non-invasively by mask (ISO 80601-2-12).
  • Patients: newborns and premature babies, invasive ventilation and nasal CPAP with tidal volumes down to 2 mL. (with the optional module)
  • Standard ventilation modes: volume and pressure control, SIMV, CPAP with pressure support, apnoea back-up and non-invasive ventilation with leak compensation.
  • Oxygen 21-100 % with continuous measured FiO2, and PEEP up to at least 35 cmH2O for stiff (ARDS) lungs.
  • Monitoring and alarms: pressure, flow and volume waveforms and loops, 72-hour trends, and three-priority alarms for pressure, volume, apnoea, oxygen, gas and power failure (IEC 60601-1-8).
  • Keeps ventilating on internal battery during power cuts for at least 30 minutes (120-240 minutes with larger battery options).
  • Advanced modes: volume-targeted pressure control, BiLevel and APRV, and automatic tube compensation. (with the optional module)
  • Automated (closed-loop) mode that adjusts support to the patient and helps weaning. (with the optional module)
  • Full ventilation without piped medical air, using a built-in turbine or compressor. (with the optional module)
  • Lung mechanics: compliance, resistance, auto-PEEP, driving pressure and RSBI. (with the optional module)
  • High-flow oxygen therapy by nasal cannula up to 60 L/min. (with the optional module)
  • Exhaled CO2 (capnography) monitoring (ISO 80601-2-55). (with the optional module)

Not covered: High-frequency oscillatory ventilation (HFOV) for newborns needs a dedicated high-frequency ventilator.; Anaesthesia with volatile agents in the operating theatre needs an anaesthesia workstation.; Ventilation in ambulances and during transfer between hospitals needs a transport ventilator.; Arterial blood gases (pH, PaO2, PaCO2) need a blood gas analyser; ECG and blood pressure need a patient monitor.

9 international manufacturers checkedMarket checked 2026-1013 standards referenced

Loading the builder...

Standards this specification draws on

  • ISO 80601-2-12 Particular safety and essential performance standard for critical care ventilators.
  • IEC 60601-1 General electrical and mechanical safety for medical electrical equipment.
  • IEC 60601-1-2 Electromagnetic compatibility, so the ventilator is not disturbed by other ICU equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and signals.
  • ISO 80601-2-55 Respiratory gas monitors (O2 and CO2 measurement).
  • ISO 80601-2-61 Pulse oximeters, if SpO2 is integrated.
  • ISO 80601-2-74 Heated humidifiers supplied with the ventilator.
  • ISO 5356-1 Standard 15/22 mm conical connectors so any circuit and filter fits.
  • IEC 62304 Software life-cycle for medical device software.
  • IEC 62366-1 Usability engineering, reducing use errors.
  • ISO 13485 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules, 2017 Registration/enlistment required to import and sell the ventilator in Pakistan.
  • WHO technical specifications for invasive and non-invasive ventilators (2020) Reference minimum specification widely used for public procurement.

Common mistakes in tenders for this equipment

  • Naming proprietary modes or tools (for example a single maker's adaptive, closed-loop or neural-trigger mode) instead of describing the function, which locks the tender to one brand.
  • Copying one brochure's exact numbers (such as a tidal volume range, screen size to the inch or battery minutes) so only that model complies.
  • Requiring a specific sensor technology (for example paramagnetic O2 or hot-wire flow sensor) rather than the needed accuracy and life.
  • Buying compressed-air-only ventilators for hospitals without a reliable medical air plant, so ventilators stop when the compressor fails.
  • Not pricing consumables (flow sensors, O2 cells, circuits) and post-warranty CMC, so a cheap ventilator becomes expensive to run.
  • No written 7-10 year spare-parts guarantee; Medtronic announced in 2024 that it would exit the ventilator business and phase out its Puritan Bennett line, showing support can end.
  • Ignoring altitude: hospitals in Quetta (about 1,700 m) and other highland areas need ventilators with barometric compensation.
  • Bundling neonatal and adult needs into one tender when only a few beds need neonatal capability, raising the cost of every unit.
Medicine brands A–Z:ABCDEFGHIJKLMNOPQRSTUVWXYZ0-9