Portable airborne particle counter for cleanroom and clean-air device classification and monitoring (ISO 14644-1:2015, EU GMP Annex 1), 0.5 and 5.0 um channels, 28.3 to 100 L/min, ISO 21501-4 calibrated, with data-integrity options

A portable airborne particle counter pulls a fixed flow of air through a laser sensor and counts particles by size, mainly at 0.5 um and 5.0 um, to classify and monitor cleanrooms and clean-air devices against ISO 14644-1 and EU GMP / WHO grade limits. Buyers are sterile pharma plants and their QC labs, drug testing laboratories with sterility-testing suites, hospital pharmacies and operating theatres, cleanroom certification contractors and local device manufacturers. It counts total particles only and does not replace microbial air sampling.

What this equipment will do

  • Areas handled: pharmaceutical cleanrooms Grade A to D and ISO Class 5 to 8 rooms, including filling areas and sterility-testing suites (EU GMP Annex 1, WHO TRS 1044 Annex 2).
  • Areas handled: clean-air devices such as laminar air-flow benches, biosafety cabinets and isolators during qualification and recertification (ISO 14644-3).
  • Areas handled: hospital pharmacy compounding rooms and hoods (USP <797>), operating theatres and device-manufacturing cleanrooms.
  • Counts airborne particles per cubic metre at 0.5 um and 5.0 um and at least four other sizes, with ISO 21501-4 calibrated counting efficiency (ISO 21501-4).
  • Counts particles down to 0.3 um for ISO Class 5 and below and for clean-air device checks. (with the optional module)
  • Cleanroom classification and routine monitoring against ISO 14644-1:2015 classes and EU GMP grade limits, with pass/fail per location. (with the optional module)
  • Samples 1 cubic metre in about 10 minutes (100 L/min) or 20 minutes (50 L/min) for Grade A/B volumes. (with the optional module)
  • Daily zero-count check with a purge filter; false count not more than 1 per 5 minutes. (with the optional module)
  • Secure GMP records: unique user log-ins, audit trail and unalterable raw data (EU GMP Annex 11, ALCOA+). (with the optional module)
  • Electronic signatures on records (21 CFR Part 11). (with the optional module)
  • Printed result at the sampling point. (with the optional module)
  • Air temperature and relative humidity recorded with each sample. (with the optional module)
  • Particle counts in compressed air and process gases such as nitrogen. (with the optional module)

Not covered: Living microorganisms in air or on surfaces (viable monitoring) need an active microbial air sampler, settle plates and contact plates.; Air velocity, airflow volume and room pressure differences need an anemometer or balometer and a differential-pressure manometer.; Routine HEPA filter leak scanning is normally done with an aerosol generator and aerosol photometer; a particle counter can do it only with a suitable challenge aerosol and test method (ISO 14644-3).; Sub-visible particles in injections (USP <788>) need a liquid (light-obscuration) particle counter, and outdoor PM2.5 / PM10 needs a particulate sampler.

7 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Electrical safety of laboratory and measuring equipment, including the counter, its battery charger and power supply.
  • IEC 61326-1 Electromagnetic compatibility (EMC) of laboratory electrical equipment.
  • ISO 21501-4:2018 Calibration and verification of light-scattering airborne particle counters for clean spaces: counting efficiency, size resolution, flow rate, false count and concentration limits.
  • ISO 14644-1:2015 Classification of cleanroom and clean-zone air cleanliness by particle concentration; sets class limits, number of sampling locations and minimum sample volume.
  • ISO 14644-2:2015 Monitoring plan to give evidence of continued cleanroom performance related to air cleanliness by particle concentration.
  • ISO 14644-3:2019 Test methods for cleanrooms and clean-air devices, including airborne particle count and installed filter leak testing with a particle counter.
  • EU GMP Annex 1 (2022) Manufacture of sterile medicinal products: Grade A-D particle limits at 0.5 and 5.0 um for classification and monitoring, and expectations for monitoring systems and data.
  • WHO TRS 1044, Annex 2 (2022) - WHO GMP for sterile pharmaceutical products WHO good manufacturing practices for sterile products, aligned with EU GMP Annex 1 grade limits and monitoring.
  • USP <797> Pharmaceutical compounding of sterile preparations: ISO Class 5 primary engineering controls and ISO 7/8 rooms certified with total particle counts at least every 6 months.
  • 21 CFR Part 11 US FDA rule for electronic records and electronic signatures; reference for the counter's audit trail, user access and e-signature functions.
  • EU GMP Annex 11 Computerised systems in GMP: validation, audit trails, access control and data integrity of the counter software.
  • ISO/IEC 17025:2017 Competence of testing and calibration laboratories; the counter's calibration certificate should come from an accredited laboratory with SI traceability.

Common mistakes in tenders for this equipment

  • Copying one brochure's figures (exact flow such as 3.53 CFM only, 8 channels, a specific battery hours figure, polycarbonate or a named software) so only one brand complies; specify minimum flow, required channels (0.5 and 5.0 um) and ISO 21501-4 instead.
  • Buying a low-flow handheld counter (2.83 L/min) for GMP grading; it cannot collect the large sample volumes needed for 5.0 um limits in Grade A/B in reasonable time.
  • Leaving out the 5.0 um channel or accepting a counter calibrated only by the maker's own method rather than ISO 21501-4, so counts are not comparable with GMP limits.
  • Forgetting the isokinetic probe, zero-count filter, spare battery and printer paper, so sampling in unidirectional flow and daily zero checks cannot be done correctly.
  • No firm annual calibration price, turnaround or loan unit; the counter goes abroad for months and monitoring stops or continues with an expired certificate.
  • Accepting 'GMP compliant' or 'Part 11 ready' without a written description of user levels, audit trail and export functions, then failing computerised-system validation.
  • Treating total particle counts as proof of sterility; viable air sampling and surface monitoring are still required.