Atomic absorption spectrophotometer with flame and graphite furnace atomisation, background correction and optional hydride / cold-vapour mercury generation

An atomic absorption spectrophotometer (AAS) measures metals such as lead, cadmium, arsenic, mercury, zinc and iron by how much light of a specific wavelength their atoms absorb in a flame or a heated graphite tube. In Pakistan it is bought by pharma QC and drug testing labs (metal assays and elemental impurities), food and water labs (heavy metals in spices, rice, fish and drinking water) and university and environmental labs. The key choices are flame versus graphite furnace, the background correction method, hydride and mercury vapour generation, and data-integrity software.

What this equipment will do

  • Samples: liquid solutions and acid digests of medicines, herbal and unani products, foods (spices, rice, honey, fish), drinking water and wastewater, soil, plant and biological materials (Ph. Eur. 2.2.23, USP <852>).
  • Flame measurement of metals at mg/L levels, such as sodium, potassium, calcium, magnesium, zinc, iron, copper and manganese, with deuterium background correction (APHA 3111, AOAC 999.11).
  • Graphite furnace measurement of trace metals at ug/L levels, such as lead, cadmium, chromium and nickel in water, food and medicines, with an automatic furnace autosampler (ISO 15586, APHA 3113, USP <233>). (with the optional module)
  • Correction of structured background in salty or complex furnace samples by Zeeman-effect or self-reversal background correction. (with the optional module)
  • Arsenic, selenium, antimony, bismuth and tin by hydride generation, for example arsenic in drinking water (APHA 3114). (with the optional module)
  • Mercury by cold-vapour generation, for example mercury in fish, medicines and water. (with the optional module)
  • Calibration by multi-point curve or by standard addition for difficult matrices, with mean, SD and RSD statistics.
  • Sensitivity and precision proven at installation with certified standards, as part of instrument qualification (USP <852>, USP <1058>).
  • Unattended flame runs of 50 or more samples with programmed rinsing. (with the optional module)
  • Several elements (not less than 6) measured on each flame sample in one aspiration by fast sequential switching. (with the optional module)
  • Automatic QC checks in each run: blanks, check standards, spike recovery and duplicates, with automatic actions on failure. (with the optional module)
  • Electronic records for GMP labs: user roles, audit trail, electronic signatures and protected raw data (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)

Not covered: Fast simultaneous analysis of many elements, ultra-trace (ng/L) levels and non-metals such as phosphorus, sulphur or boron need an ICP-OES or ICP-MS.; Dissolving solid samples before measurement needs a microwave digestion system or hot-block digestion; the AAS only reads liquid solutions.; The chemical form of a metal (for example inorganic versus organic arsenic, or methylmercury) needs speciation by HPLC coupled to ICP-MS.; Organic contaminants such as pesticides, aflatoxins and residual solvents need HPLC, GC or GC-MS.

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety requirements for laboratory equipment, covering the spectrometer, furnace power supply and accessories.
  • IEC 61326-1 Electromagnetic compatibility (EMC) of laboratory equipment, important beside high-current furnace supplies and other lab instruments.
  • Ph. Eur. 2.2.23 European Pharmacopoeia general chapter on atomic absorption spectrometry (also reproduced in the BP), defining apparatus, background correction and calibration methods.
  • USP <852> USP general chapter on atomic absorption spectroscopy: instrument qualification (wavelength, sensitivity, precision) and procedure validation for AAS.
  • USP <232> / USP <233> USP elemental impurity limits and the procedure-validation requirements that an AAS method for Pb, Cd, As and Hg must meet.
  • ICH Q3D(R2) International guideline setting permitted daily exposures for elemental impurities in medicines, which DRAP and export markets expect manufacturers to control.
  • USP <1058> Framework for analytical instrument qualification (DQ, IQ, OQ, PQ).
  • 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and e-signature requirements for software in GMP labs.
  • ISO 15586 International standard method for trace elements in water by graphite furnace AAS, used by water quality labs.
  • APHA Standard Methods 3111 / 3113 / 3114 Standard Methods for the Examination of Water and Wastewater: flame AAS (3111), graphite furnace AAS (3113) and hydride AAS for As and Se (3114).
  • AOAC 999.11 Official method for lead, cadmium, copper, iron and zinc in foods by AAS after dry ashing.
  • ISO 17034 / ISO/IEC 17025 Accreditation of reference material producers and calibration labs supplying the certified standards used for calibration and qualification.

Common mistakes in tenders for this equipment

  • Copying one maker's datasheet figures (an exact number of lamp positions, a specific furnace maximum temperature such as 3,000 deg C, a particular magnet type or a named heating geometry) so only that model complies; state the minimum performance the methods need.
  • Using a maker's trade names for furnace platforms, background correction systems or vapour generators instead of describing the measurable function.
  • Buying a flame-only AAS for a lab that must report lead, cadmium, arsenic or mercury at ug/L levels, then discovering that furnace and vapour generation modules cost almost as much again.
  • Leaving out graphite tubes, lamps, certified standards, gas regulators, chiller, compressor and exhaust, so the instrument is delivered but cannot be operated safely or at all.
  • Insisting on Zeeman background correction for routine pharmacopoeial or water work, excluding self-reversal and deuterium systems that pass the same validation.
  • Accepting 'Part 11 ready' software without requiring audit trail, unique logins and e-signatures to be demonstrated at installation, or finding later that the compliance module is a separately priced licence.
  • Ignoring site readiness: furnace power supplies need a dedicated high-current circuit with good earthing, and acetylene cylinders need a ventilated, secured area outside the lab.
  • Ignoring post-warranty cost; graphite tubes, lamps and argon can cost more each year than the service contract, so ask bidders to price a year of consumables.