Specification builder · Tier A
Automated external defibrillator (AED), semi-automatic or fully automatic, biphasic, with CPR coaching and paediatric capability, for hospitals, ambulances and public access
A small battery-powered defibrillator that analyses the heart rhythm through two sticky pads and tells the user, by voice, whether to shock and how to do CPR; a fully automatic model gives the shock itself. Hospitals place them on wards and OPDs, ambulance and rescue services carry them, and airports, malls, schools and offices mount them on walls for bystanders. The key tender decisions are semi or fully automatic, CPR feedback, child mode, pad and battery shelf life, and a readiness service that keeps pads in date.
What this equipment will do
- Adults in cardiac arrest: automatic heart-rhythm analysis through two pads and a biphasic shock adjusted to the patient's impedance (IEC 60601-2-4).
- Children under 8 years or 25 kg: reduced energy of about 50 J using paediatric pads, a key or a child-mode switch, with child CPR prompts (IEC 60601-2-4).
- Semi-automatic shock: the AED charges and the rescuer presses the shock button when told to. (with the optional module)
- Fully automatic shock: the AED warns stand clear, counts down and shocks by itself, for rescuers who might hesitate. (with the optional module)
- Step-by-step voice prompts and CPR coaching with a 100-120 per minute metronome, designed for untrained lay rescuers (IEC 60601-1-11, IEC 62366-1).
- Real-time feedback on the rescuer's actual compression rate (and depth where a sensor is used), with push faster / push harder prompts. (with the optional module)
- Rhythm analysis while chest compressions continue, keeping the pause before a shock to 10 seconds or less. (with the optional module)
- Live ECG rhythm, heart rate and rescue time on a colour screen, for doctors and nurses. (with the optional module)
- Use in ambulances and aircraft, tested for shock and vibration and working up to 50 deg C (IEC 60601-1-12). (with the optional module)
- Readiness self-tests of battery and pads, with pad and battery expiry shown by the device.
- Remote readiness reports for a fleet of AEDs over Wi-Fi or cellular, with e-mail or SMS alerts when a unit is not ready. (with the optional module)
- Storage and download of the rescue record (ECG, shocks, energy, CPR times) for clinical review. (with the optional module)
Not covered: Manual defibrillation, synchronised cardioversion, external pacing and monitoring of SpO2, CO2 or a 12-lead ECG need a manual defibrillator-monitor.; Continuous heart monitoring of ward or ICU patients needs a patient monitor; diagnostic ECGs need a 12-lead ECG machine.; Breathing support during resuscitation (bag-valve-mask, oxygen, airway devices, suction) needs the crash cart resuscitation kit.; Long or transport resuscitations with continuous chest compressions need a mechanical CPR device.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General basic safety and essential performance of medical electrical equipment.
- IEC 60601-2-4 Particular requirements for cardiac defibrillators, including AED rhythm recognition performance, energy accuracy, charge time and operator protection.
- IEC 60601-1-2 Electromagnetic compatibility so the AED analyses correctly near phones, radios and other equipment.
- IEC 60601-1-11 Requirements for medical electrical equipment used by lay operators outside professional healthcare facilities.
- IEC 60601-1-12 Requirements for equipment used in the emergency medical services environment (ambulances, aircraft).
- IEC 62366-1 Usability engineering, so that untrained rescuers can follow the prompts correctly under stress.
- IEC 60529 Degrees of protection by enclosures (IP code) against dust and water.
- IEC 62304 Software life-cycle for the rhythm analysis and control software.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Requiring 360 J, a named waveform or a specific escalation sequence: this copies one brand and excludes equally effective validated 150-200 J biphasic units.
- Naming a proprietary CPR feedback or rhythm-analysis feature instead of the measurable function (compression rate or depth feedback, analysis pause time).
- Buying AEDs without spare pads, paediatric pads and a fixed 5-year consumable price; some units accept only costly single-source pads or combined pad-battery cartridges.
- Accepting pad and battery life counted from manufacture with stock that is already near expiry on delivery; specify minimum remaining shelf life at delivery.
- Placing AEDs in locked offices or without signage and trained staff, so they cannot be reached within 3 minutes; ask for a placement survey and training.
- Mounting AEDs outdoors or in vehicles without heat protection; Pakistan summer heat ruins pads and batteries.
- Accepting DRAP registration of a different model or of the device without its pads and batteries.
- Accepting refurbished, monophasic or discontinued models; require new, current-production biphasic units with 8-year parts and pads support.