Specification builder · Tier A
Microbiological incubator: heating-only bacteriological incubator (ambient +5 to 70 deg C) or cooled (compressor or thermoelectric) incubator for 20-25 deg C, below-ambient and 20 deg C BOD work
A microbiological incubator is a heated (and often cooled) cabinet that holds cultures at a fixed temperature, for example 30-35 deg C and 20-25 deg C for pharmacopoeial microbial limit and sterility tests, 37 deg C for clinical bacteria and 20 deg C for the 5-day BOD test of water. Buyers are pharma QC and drug testing laboratories, medical-device manufacturers, water and environmental labs and teaching institutions. In Pakistan, hot summer rooms and load-shedding decide the choice: heating-only units cannot hold temperatures near or below room temperature, so cooled incubators, uniformity proven to DIN 12880, temperature mapping and power backup matter as much as size.
What this equipment will do
- Sample types: agar plates, broth tubes, sterility-test media containers and environmental-monitoring plates, on stainless-steel shelves.
- Incubation at 30-35 deg C and 37 deg C: total aerobic counts (USP <61>/<62>, Ph. Eur. 2.6.12/2.6.13), device bioburden (ISO 11737-1), clinical and teaching cultures. A heating-only unit needs the room at least 5 deg C cooler than the set point.
- Incubation at 20-25 deg C whatever the room temperature, for yeast and mould counts (USP <61>, Ph. Eur. 2.6.12). (with the optional module)
- Sterility test: media held for 14 days at both 30-35 deg C and 20-25 deg C (USP <71>, Ph. Eur. 2.6.1). (with the optional module)
- 5-day BOD test of water and wastewater: 300 mL BOD bottles held in the dark at 20 +/- 1 deg C (Standard Methods 5210 B). (with the optional module)
- Cold incubation below room temperature: down to +5 deg C, or down to -5 deg C (including 4 deg C) with the low-temperature model. (with the optional module)
- Holds 20 deg C even when the room reaches 35 deg C or more (no air conditioning in summer). (with the optional module)
- Automatic two-temperature programme, for example 20-25 deg C then 30-35 deg C for environmental-monitoring plates. (with the optional module)
- Even temperature on every shelf, with uniformity and fluctuation measured by the DIN 12880 method, and an independent over-temperature cut-out that protects the cultures.
- Temperature record for test reports: 30 days or more of chamber temperature and alarm history, exported as CSV or PDF. (with the optional module)
- Remote alarms by SMS or e-mail (or to a building alarm system) at night and on holidays. (with the optional module)
- Tamper-evident electronic incubation records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)
Not covered: CO2 incubation for cell culture or CO2-loving organisms needs a CO2 incubator.; Anaerobic culture needs an anaerobic jar or anaerobic workstation.; Sterilising media or drying glassware needs an autoclave or hot-air oven; this incubator only reaches about 60-70 deg C.; Stability testing of medicines at controlled humidity (for example 25 deg C / 60% RH, 40 deg C / 75% RH) needs a stability (climatic) chamber.
Loading the builder...
Standards this specification draws on
- IEC 61010-1:2010+AMD1:2016 Basic electrical, mechanical and fire safety for laboratory equipment; the safety baseline for any incubator.
- IEC 61010-2-010:2019 Particular safety requirements for laboratory equipment for the heating of materials, including over-temperature protection; cooled models are assessed to IEC 61010-2-011 or IEC 61010-2-012 where those parts' selection rules apply.
- IEC 61326-1:2020 Electromagnetic compatibility (EMC) of laboratory equipment, so the controller and alarms work reliably near other equipment.
- DIN 12880:2007 German standard for laboratory heating cabinets and incubators that defines how temperature uniformity and fluctuation are measured and the over-temperature protection classes quoted on datasheets.
- USP <61> / <62> (Ph. Eur. 2.6.12 / 2.6.13) Harmonised microbial enumeration and specified-microorganism tests for non-sterile products, incubated at 30-35 deg C and 20-25 deg C.
- USP <71> (Ph. Eur. 2.6.1) Sterility test: media incubated for not less than 14 days at 30-35 deg C and 20-25 deg C.
- ISO 11737-1:2018+A1:2021 Determination of the population of microorganisms (bioburden) on medical devices, with validated incubation conditions.
- Standard Methods 5210 B (APHA/AWWA/WEF) 5-day BOD test: samples incubated in the dark at 20 +/- 1 deg C for 5 days.
- WHO TRS 961 Annex 2 (2011) WHO good practices for pharmaceutical microbiology laboratories, which expect incubators to be qualified, temperature-mapped, monitored and calibrated.
- 21 CFR Part 11 / EU GMP Annex 11 Requirements for trustworthy electronic records, audit trails and signatures for incubation records.
- ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories, including traceable calibration of incubation equipment.
- ISO 9001:2015 Quality management system of the manufacturer.
Common mistakes in tenders for this equipment
- Buying a heating-only incubator for 20-25 deg C work (yeast and mould counts, SCDM sterility media) or for BOD at 20 deg C; it physically cannot cool below room temperature.
- Ignoring the room: heating-only and thermoelectric units fail in a 40 deg C summer room when the AC is off during load-shedding, and nobody notices until results are questioned.
- Copying one maker's exact figures (a fixed litre size such as 108 L or 256 L, a particular temperature range or a named software) so that only one brand qualifies; use 'not less than' sizes and ranges.
- Accepting controller 'accuracy' or 'setting accuracy' instead of spatial uniformity and fluctuation measured to DIN 12880 at the working temperatures.
- Buying one incubator for a sterility test that needs two temperatures at once for 14 days.
- Skipping IQ/OQ, temperature mapping and an accredited calibration certificate, then failing an ISO/IEC 17025 assessment or GMP inspection.
- Relying only on the incubator's own display, with no independent calibrated probe and no night-time alarm, so failures are found the next working day.
- Leaving out the voltage stabiliser and UPS, so voltage swings damage controllers and compressors within the first year.