Benchtop laboratory centrifuge, general purpose (swing-out rotor for blood and 15 / 50 mL tubes, optional fixed-angle microtube rotor, non-refrigerated or refrigerated)
A benchtop centrifuge separates solids from liquids by spinning tubes at a set force, time and temperature. Hospital labs use it to separate serum and plasma from blood, pharma QC and drug testing labs to clarify sample extracts before HPLC or UV analysis, and microbiology labs to collect pellets. The real buying decisions are the rotors and tube adapters, refrigeration, sealed (aerosol-tight) buckets for infectious samples and local calibration and service.
What this equipment will do
- Blood collection tubes (13 x 75 mm, 13 x 100 mm and similar) in a swing-out rotor for serum and plasma separation (CLSI GP44-A4).
- Conical tubes of 15 mL and 50 mL in a swing-out rotor at not less than 3,000 x g. (with the optional module)
- Microtubes of 1.5 / 2.0 mL in a fixed-angle rotor at not less than 15,000 x g. (with the optional module)
- 50 mL conical tubes in a fixed-angle rotor at not less than 8,000 x g for faster pelleting. (with the optional module)
- Clarifying sample solutions and extracts (removing undissolved excipients) before HPLC, UV and dissolution analysis.
- Pelleting bacteria, yeasts and cells from broth or suspension, held at +4 deg C. (with the optional module)
- Spinning infectious or toxic samples in sealed buckets and rotor lids that are opened inside a biological safety cabinet (WHO Laboratory Biosafety Manual). (with the optional module)
- Runs set by RCF (x g) or speed (rpm), with timer, short-spin and automatic stop on imbalance.
- Gentle acceleration and braking curves for loose pellets and layered samples. (with the optional module)
- Safety to IEC 61010-1 and IEC 61010-2-020: lid locked while the rotor turns, emergency lid release, imbalance cut-off.
Not covered: Separation of whole-blood donation bags into red cells, plasma and platelets needs a refrigerated blood component centrifuge.; Haematocrit (packed cell volume) in capillary tubes needs a haematocrit centrifuge or a haematocrit rotor; cytology slide preparation needs a cytocentrifuge.; Forces above about 25,000 x g or large bottle volumes (litres per run) need a high-speed floor-standing centrifuge or an ultracentrifuge.; Drying or concentrating samples by evaporating solvent needs a vacuum centrifugal concentrator.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical, mechanical and fire safety of laboratory equipment.
- IEC 61010-2-020 Particular safety requirements for laboratory centrifuges: lid interlock, containment of rotor failure, imbalance and rotor energy (current edition published 2026, replacing the 2016 edition).
- IEC 61326-1 Electromagnetic compatibility of laboratory electrical equipment.
- WHO Laboratory Biosafety Manual, 4th edition (2020) Risk-based guidance on safe centrifugation of infectious material, including sealed buckets opened in a biological safety cabinet.
- CLSI GP44-A4 Procedures for handling and processing blood specimens, including centrifugation force, time and temperature for serum and plasma separation.
- ISO 15189:2022 Medical laboratory accreditation requirements, including equipment calibration and maintenance records for hospital labs.
- ISO 15190:2020 Safety requirements for medical laboratories, including centrifuge use and spill handling.
- ISO/IEC 17025:2017 Testing and calibration laboratory requirements; centrifuge speed, time and temperature calibration must be traceable.
- USP <1058> Framework for analytical instrument qualification (IQ/OQ/PQ) of equipment used in GMP QC labs.
Common mistakes in tenders for this equipment
- Specifying speed in rpm only instead of the force (RCF, x g) the method needs; rpm gives different force on rotors of different radius and favours one maker's rotor.
- Copying one maker's brochure: exact maximum RCF figures, named bucket or lid systems and unusual temperature limits (for example -20 deg C) that only one or two makers quote.
- Leaving out the tube adapters actually used in the lab (blood tube sizes, 15 / 50 mL skirted and non-skirted conicals), so the centrifuge arrives but cannot run the samples.
- Not asking for sealed (aerosol-tight) buckets and rotor lids for blood and microbiology samples, or accepting 'biosafe' claims with no test report.
- Citing IEC 60601 (medical electrical equipment) instead of IEC 61010-1 and IEC 61010-2-020 for a laboratory centrifuge, or asking for DRAP medical device registration, which wrongly excludes valid bidders.
- No calibration plan: speed, timer and temperature need accredited checks at installation and every year for ISO 15189, ISO/IEC 17025 and GMP labs.
- Ignoring rotor life and corrosion: no written rotor life limit, no inspection in the service contract, and disinfectants that attack aluminium rotors.
- Buying a refrigerated unit without an online UPS or stabiliser and without room ventilation, so the compressor fails early in hot rooms with unstable power.