Upright compound laboratory microscope (binocular / trinocular) with infinity plan optics and LED illumination

An upright compound microscope with plan objectives up to 100x oil immersion and LED illumination, used to read stained smears, blood films, urine sediment and cultures. It is the basic instrument of every hospital clinical lab, TB microscopy centre, drug testing laboratory microbiology section and teaching lab. Options add phase contrast, darkfield, LED fluorescence for auramine-stained TB smears, polarisation and a camera with measurement software.

What this equipment will do

  • Sample types: Gram- and Ziehl-Neelsen-stained smears, blood films, urine sediment, stool preparations and smears of isolates from microbial limit and sterility tests.
  • Bacterial shape and staining at 1000x with a 100x oil-immersion objective; plan objectives keep the whole field sharp (ISO 19012-1).
  • Systematic scanning of the slide with a mechanical stage (75 x 50 mm travel) and position scales to find a field again.
  • Unstained live cells and motility by phase contrast. (with the optional module)
  • Thin organisms such as spirochaetes by darkfield. (with the optional module)
  • Auramine-stained TB sputum smears by LED fluorescence, without a darkroom (WHO LED fluorescence microscopy policy, IEC 62471). (with the optional module)
  • Crystals (for example in urine sediment) by simple polarisation. (with the optional module)
  • Images saved as records, with calibrated length and area measurement, counting and annotation. (with the optional module)
  • Live image shared to several tablets or PCs for teaching. (with the optional module)
  • Size measurement traceable to a certified stage micrometer (ISO/IEC 17025). (with the optional module)
  • Reading slides during load-shedding on battery power. (with the optional module)

Not covered: Counting colonies on agar plates needs a colony counter.; Low-power 3D viewing of large objects (colonies, insects, tablet surfaces, dissection) needs a stereo microscope.; Naming the species or testing antibiotic susceptibility needs a microbial identification / susceptibility system; the microscope shows only shape and staining.; Full blood counts need a haematology analyser, and molecular TB and drug-resistance testing needs a rapid PCR (nucleic acid) system.

9 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety for laboratory equipment, covering the LED illuminator, fluorescence module and their power supplies.
  • IEC 61326-1 EMC requirements for laboratory equipment, so the LED drivers and camera do not interfere with or suffer from other lab instruments.
  • ISO 19012-1:2013 Defines when an objective may be called 'plan' (flatness of field), so plan claims can be checked.
  • ISO 19012-2:2013 Defines chromatic correction classes (achromat, semi-apochromat, apochromat) used to compare objectives.
  • ISO 8039:2014 Sets the tolerances on stated magnification of objectives and eyepieces, which matters for measurement.
  • ISO 9345:2019 Interfacing dimensions (parfocal distance, threads, tube length) that keep objectives and cameras interchangeable.
  • ISO 8036:2015 Specifies immersion oil properties so the 100x objective delivers its rated resolution.
  • IEC 62471 Photobiological safety classification of lamps and LED systems, relevant to the blue LED fluorescence module.
  • WHO policy: LED fluorescence microscopy for TB (2011) WHO recommends LED fluorescence microscopy as an alternative to conventional fluorescence and to Ziehl-Neelsen microscopy, and sets expectations for LED microscopes in TB programmes.
  • ISO 15189:2022 Medical laboratory accreditation standard; requires equipment to be fit for purpose, maintained and with documented performance checks.
  • ISO/IEC 17025:2017 Testing laboratory standard; requires traceable calibration of measuring equipment, such as the stage micrometer used for image measurement.

Common mistakes in tenders for this equipment

  • Copying one brand's numbers (an exact tilt range, a field number of 22 only, a named LED wavelength or a specific stage size) so only that model complies; state minimums with 'not less than' instead.
  • Writing proprietary names of optics or illumination systems instead of measurable features such as 'infinity-corrected plan achromat, NA not less than 1.25'.
  • Calling an objective 'plan' or 'Koehler illumination' without asking for the datasheet proof, then receiving achromats with blurred field edges and a simple Abbe condenser.
  • Buying LED fluorescence microscopes for TB without auramine reagents, training or a light shield, so they sit unused while staff return to Ziehl-Neelsen.
  • Leaving out immersion oil, cleaning kit and a desiccator box; within two monsoon seasons fungus clouds the objectives and prisms.
  • Asking for a camera without a trinocular head, C-mount adapter matched to the sensor, computer or calibrated measurement software.
  • Accepting 1-year warranty with no cleaning visits; most microscope faults in Pakistan are fungus and misalignment that a yearly service prevents.
  • Demanding DRAP medical-device registration for a general laboratory microscope; ask for the DRAP status statement instead, so lab-only models are not wrongly excluded.