Neonatal bubble CPAP system with air-oxygen blending, heated humidification, bubble pressure generator and nasal prongs/masks for SNCU, NICU and district hospitals
Bubble CPAP gives newborns with breathing difficulty a gentle, continuous pressure through nasal prongs or a mask by bubbling the expired gas through water, with blended air and oxygen that is warmed and humidified. It is the main non-invasive breathing support in special care newborn units and NICUs and greatly reduces the need for ventilators. In Pakistan the key tender decisions are where medical air comes from (pipeline or built-in compressor), accurate oxygen blending with an analyser and oximeter, heated humidification, enough prongs in every size, power back-up and nurse training.
What this equipment will do
- Patients: spontaneously breathing preterm and term newborns up to about 10 kg, from below 1 kg to above 2.5 kg, using soft nasal prongs in at least 3 sizes with bonnets.
- Gas supply where there is no piped medical air: air from a built-in or separate oil-free compressor, so only an oxygen pipeline or cylinder is needed. (with the optional module)
- Fully self-contained gas supply with a built-in oxygen concentrator (ISO 80601-2-69), needing no pipeline or cylinder. (with the optional module)
- CPAP pressure set from 3 to 10 cmH2O by the depth of the tube in the water-seal bottle, with bubbling visible to the nurse and pressure relief at not more than 20 cmH2O.
- Oxygen mixing: delivered oxygen (FiO2) set from 21 % to at least 90 % and checked with an oxygen analyser, never 100 % oxygen only.
- Precision oxygen control: blender 21-100 % with gas-failure alarm (ISO 11195), or electronic blending with continuous FiO2 display and high/low FiO2 alarms. (with the optional module)
- Warming and humidifying the gas: servo heated humidifier with heated tube, gas at about 34-37 deg C, at least 33 mg/L humidity and cut-off before 41 deg C (ISO 80601-2-74). (with the optional module)
- Power cuts: CPAP keeps running for at least 30 minutes on a dedicated online UPS or 60 minutes on an internal battery. (with the optional module)
- Nasal masks to alternate with prongs, to reduce injury to the nose. (with the optional module)
- Digital airway pressure display with high-pressure and disconnection alarms. (with the optional module)
- High-flow nasal cannula therapy for weaning, using the same blender and heated humidifier. (with the optional module)
- Oxygen saturation (SpO2) and pulse monitoring to titrate oxygen, built in or as a standalone oximeter (ISO 80601-2-61), with 24-hour trends where ordered. (with the optional module)
Not covered: Babies who are not breathing, have repeated apnoea or need mandatory breaths through a tube need a neonatal ventilator.; CPAP or BiPAP for older children and adults needs a non-invasive ventilator (BiPAP/CPAP).; Resuscitation at birth (bag-mask or T-piece breaths) needs a neonatal resuscitation unit; keeping the baby warm needs an infant incubator or radiant warmer.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical, mechanical and thermal safety of the powered parts (humidifier, compressor, electronic blender).
- ISO 80601-2-74 Particular safety and essential performance of respiratory humidifying equipment: delivered humidity, gas temperature limits, over-temperature protection and alarms.
- IEC 60601-1-2 Electromagnetic compatibility, so humidifier control and alarms work near phones, monitors and other NICU equipment.
- IEC 60601-1-8 Alarm system design: priorities, sounds and visual signals for humidifier, blender and pressure alarms.
- ISO 11195 Gas mixers (air-oxygen blenders) for medical use: accuracy, supply-failure alarm and prevention of cross-flow between gases.
- ISO 5367 Breathing sets and tubes: leakage, connector security and resistance of the patient circuit.
- ISO 80601-2-61 Pulse oximeter safety and accuracy, for built-in SpO2 or the standalone oximeters supplied to titrate oxygen.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer(s) of the CPAP system, humidifier and blender.
- WHO-UNICEF Technical specifications and guidance for oxygen therapy devices (2019) Reference specifications for flowmeters, humidifiers, oxygen analysers, pulse oximeters and interfaces used in low-resource oxygen and CPAP programmes.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Copying one maker's brochure: a named prong or mask system, a proprietary self-levelling generator, an exact CPAP scale or a single-make heated circuit. This locks the hospital into one maker's consumables for years; write performance limits and let any compatible interface be offered.
- Buying the CPAP unit without first-year prongs, masks, bonnets, circuits and humidifier chambers in every size, or without a fixed 5-year price list; units then stand idle within months or consumables cost more than the machine.
- Allowing 100 % oxygen only, or a blender without an oxygen analyser and pulse oximeter, so preterm babies get far more oxygen than they need and develop retinopathy of prematurity.
- Specifying a pipeline-air system for a district nursery that has oxygen but no piped medical air, or a compressor-based system without a UPS in an area with daily load-shedding.
- Accepting unheated humidification for small preterm babies, or a heated humidifier without heated tubes and temperature probes, leading to cold dry gas, blocked prongs and nasal damage.
- Gas hoses and probes that do not fit the hospital's wall outlets; ask for a site survey and matching probes in the bid.
- No hands-on nurse training and skills check: bubble CPAP fails mostly because of poor prong sizing, leaks through an open mouth and loss of bubbling that nobody notices.
- Accepting DRAP registration of a different model or of the brand only; the exact offered CPAP system and humidifier must be registered or enlisted.