Specification builder · Tier A
Stand-alone capnograph (bedside or handheld) for end-tidal CO2 monitoring with real-time capnogram, sidestream or mainstream, for intubated and non-intubated patients in theatre, ICU, emergency and procedural sedation
A capnograph measures the carbon dioxide in each breath and shows end-tidal CO2, respiratory rate and the CO2 waveform, giving the fastest warning of a misplaced tube, stopped breathing or failing circulation. Hospitals buy stand-alone units for theatres and recovery rooms without capnography, ICUs, emergency and resuscitation teams, ambulances and procedural sedation suites. The key tender decisions are handheld or bedside, sidestream or mainstream sampling, neonatal capability, integrated SpO2, and the long-term cost of sampling lines.
What this equipment will do
- Patients: adults and children, with patient-category selection and matching default alarm limits.
- Newborns, with neonatal sampling lines at not more than 50 mL/min or infant airway adapters of not more than 1 mL dead space (ISO 80601-2-55). (with the optional module)
- Intubated and ventilated patients, measured through an airway adapter on the tracheal tube or mask (ISO 80601-2-55).
- Non-intubated, spontaneously breathing patients (sedation, recovery) through nasal or oral-nasal sampling cannulas, including cannulas that give oxygen at the same time. (with the optional module)
- End-tidal CO2, inspired CO2, airway breathing rate and real-time capnogram, range 0-99 mmHg with accuracy +/-2 mmHg up to 40 mmHg (ISO 80601-2-55).
- Confirming tube placement, guiding CPR and catching hypoventilation, rebreathing or stopped breathing, with high/low EtCO2, breathing rate and no-breath (apnoea) alarms (IEC 60601-1-8).
- Wide range up to 150 mmHg CO2 and breathing rates up to 120 breaths/min, for severe CO2 retention. (with the optional module)
- Use on the move: at least 2 hours on battery; hand-held units under 600 g and transport bag or ambulance bracket where ordered (IEC 60601-1-12). (with the optional module)
- Oxygen saturation (SpO2) and pulse rate with plethysmogram on the same unit (ISO 80601-2-61). (with the optional module)
- Composite respiratory status index and hourly apnoea and desaturation event counts for post-operative and sedation patients. (with the optional module)
- Non-invasive blood pressure on the same unit (IEC 80601-2-30). (with the optional module)
- Records and remote alarms: 24-hour trend memory, USB/PC export, nurse-call output or link to a central station or EMR (HL7). (with the optional module)
Not covered: Anaesthetic agent, oxygen and nitrous oxide concentrations need a multigas monitor on the anaesthesia workstation or patient monitor.; Arterial blood CO2, oxygen and pH (PaCO2, PaO2) need an arterial blood gas analyser; end-tidal CO2 does not replace it.; ECG, temperature and invasive pressures, and full multi-parameter monitoring, need a patient monitor.
Loading the builder...
Standards this specification draws on
- ISO 80601-2-55 Particular safety and essential performance standard for respiratory gas monitors, covering CO2 accuracy, response time, sampling-flow effects and alarms.
- IEC 60601-1 General electrical and mechanical safety for medical electrical equipment.
- IEC 60601-1-2 Electromagnetic compatibility, so diathermy, defibrillators and phones do not disturb readings.
- IEC 60601-1-8 Alarm system design: priorities, sounds and signals.
- ISO 80601-2-61 Safety and performance of pulse oximeters, for units with integrated SpO2.
- IEC 60601-1-12 Requirements for medical equipment used in the emergency medical services environment (ambulance and field use).
- ISO 5356-1 Standard 15/22 mm conical connectors so airway adapters fit any tracheal tube, filter and circuit.
- IEC 62304 Software life-cycle for medical device software that calculates readings and runs alarms.
- ISO 14971 Risk management applied by the manufacturer across design and production.
- ISO 13485 Quality management system of the manufacturer.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell the capnograph in Pakistan.
Common mistakes in tenders for this equipment
- Copying one brand's sidestream technology name, sampling-line name or exact sample flow and tolerance, which only that maker can meet; specify the flow range and patient groups instead.
- Not pricing sampling lines, nasal cannulas and airway adapters, which are often specific to one maker, so a cheap capnograph becomes expensive to run or stands unused when lines run out.
- Buying a numeric-only capnometer when a capnogram is needed; without the waveform, staff cannot recognise oesophageal intubation, rebreathing, bronchospasm or poor CPR.
- Asking for a composite respiratory index, apnoea-per-hour reports or integrated blood pressure in a general tender, which cuts the field to one or two makers.
- Choosing mainstream sensors for sedation suites, where patients are not intubated, or high-flow sidestream for neonates.
- Leaving the CO2 sensor/module and battery out of the warranty, or accepting units that need yearly factory calibration with no local calibration gas or service.
- Accepting refurbished, demonstration or discontinued models without a written undertaking that consumables will be available for at least 5-7 years.