Cardiotocograph (CTG / fetal monitor) for antepartum and intrapartum fetal heart rate and uterine activity monitoring, with twin, maternal-parameter and internal-monitoring options

A cardiotocograph (CTG) records the unborn baby's heart rate by ultrasound and the mother's contractions on a continuous paper trace, for antenatal non-stress tests and monitoring during labour. District and teaching-hospital labour rooms, antenatal clinics and private maternity hospitals buy them. In Pakistan the key tender decisions are twin monitoring, recording the maternal heart rate to avoid confusing it with the baby's, paper supply at the right speed and scale, and keeping the trace running through load-shedding.

What this equipment will do

  • Patients: pregnant women from about 28 weeks for antenatal non-stress tests and during labour, singleton pregnancies, with external transducers on the abdomen (IEC 60601-2-37).
  • Patients: twin pregnancies, each baby's heart rate on its own channel with an alert when both channels pick up the same heart. (with the optional module)
  • Patients: triplet pregnancies on three heart-rate channels. (with the optional module)
  • Fetal heart rate by Doppler ultrasound, 50 to 210 bpm or wider, accuracy +/-2 bpm, with heart sounds on a loudspeaker and FHR high, low and signal-loss alarms (IEC 60601-2-37; IEC 60601-1-8).
  • Uterine contractions (frequency, duration, relative strength) by an external toco pressure transducer, and fetal movements marked by the mother.
  • Printed trace at 1, 2 or 3 cm/min and 20 or 30 bpm/cm, matching the scale staff are trained on (FIGO guidelines 2015).
  • Automatic marking of fetal movements detected by the ultrasound transducer. (with the optional module)
  • Maternal heart rate on the same trace, so recording the mother's pulse instead of the baby's heart can be recognised. (with the optional module)
  • Maternal blood pressure and oxygen saturation on the same screen and record (IEC 80601-2-30; ISO 80601-2-61). (with the optional module)
  • Internal monitoring in labour: beat-to-beat fetal heart rate from a scalp electrode, and contraction pressure in mmHg from an intrauterine catheter. (with the optional module)
  • Computerised measurement of baseline, variability, accelerations and decelerations as decision support. (with the optional module)
  • Live traces of all labour beds at a central midwives' station with electronic archiving of complete traces. (with the optional module)

Not covered: Checking the baby's heartbeat quickly at antenatal visits or intermittent listening in labour is done with a handheld fetal Doppler; a CTG is for continuous recording.; Fetal growth, position, placenta, amniotic fluid and umbilical artery blood flow need an obstetric ultrasound scanner with colour/pulsed Doppler.; Fetal ECG ST-segment analysis and fetal blood sampling for pH or lactate need dedicated fetal ECG analysis equipment or a blood gas/lactate analyser.; Full maternal monitoring after caesarean section or in eclampsia (continuous ECG, respiration, invasive pressure) needs a multiparameter patient monitor.

7 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment.
  • IEC 60601-2-37 Particular safety and essential performance of ultrasonic diagnostic and monitoring equipment (edition 3, 2024), covering the Doppler FHR transducer and its acoustic output.
  • IEC 60601-1-2 Electromagnetic compatibility, so traces stay clean near diathermy, phones and other labour-room equipment.
  • IEC 60601-1-8 Alarm system design for FHR, signal-loss and maternal alarms.
  • IEC 80601-2-30 Safety and performance of automated non-invasive blood pressure measurement, where maternal NIBP is offered.
  • ISO 80601-2-61 Safety and accuracy of pulse oximetry, where maternal SpO2 is offered.
  • FIGO intrapartum fetal monitoring guidelines (2015) International consensus on CTG recording and interpretation, including paper speeds, FHR scales, twin and maternal heart-rate monitoring.
  • IEC 62304 Medical device software life-cycle for alarm, analysis and networking software.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Electrical safety testing at installation and after repair.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: a named coincidence-detection or pulse-sensing feature, an exact screen size or a particular number of crystals in the transducer. Specify performance (FHR range and accuracy, twin capability, maternal heart rate on the trace) instead.
  • Not fixing paper speed and FHR scale: monitors delivered at 3 cm/min and 30 bpm/cm to staff trained on 1 cm/min and 20 bpm/cm lead to misread traces; FIGO warns against unfamiliar scales.
  • Forgetting recorder paper and transducer belts: the CTG stands idle or prints on the wrong grid within months. Buy a first-year supply and fix 5-year prices for the maker's paper.
  • Buying single-channel monitors for labour rooms that see twins, so one twin is not monitored or both traces come from the same heart.
  • Accepting computerised analysis as a substitute for training: staff still need FIGO-based CTG interpretation training and regular review meetings.
  • Ignoring power cuts: monitors without battery on non-UPS sockets lose the trace during load-shedding, creating gaps in the legal record.
  • Buying a central station from one maker and monitors from another; most central systems only accept their own maker's monitors.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.