Specification builder · Tier A

Colony counter for microbiology QC (manual digital and automated image-based, with zone-of-inhibition measurement)

A colony counter helps analysts count bacterial and fungal colonies on agar plates and membrane filters. Simple manual counters register a count each time a colony is touched with a pen under a magnifier, while automated counters photograph the plate, count colonies in seconds, measure antibiotic zones and keep the image as evidence. They are used in drug testing laboratories, pharma QC, medical-device bioburden labs, food and water labs and teaching labs.

What this equipment will do

  • Sample types: standard 90 mm Petri dishes from pour-plate, spread-plate and surface counts (USP <61>, Ph. Eur. 2.6.12, ISO 4833-1).
  • Sample types: 55-65 mm contact plates from environmental monitoring and 47-50 mm membrane filters from water testing and device bioburden (ISO 11737-1). (with the optional module)
  • Sample types: large dishes up to 150 mm and square dishes of 100 x 100 mm or more. (with the optional module)
  • Colony count per plate for total aerobic microbial count and total yeast and mould count in microbial limit tests (USP <61>, Ph. Eur. 2.6.12) and for food and water counts (ISO 7218).
  • Manual counting: each colony is touched with a pen and added to a digital count (0-999) under a magnifier on a lit base. (with the optional module)
  • Automatic counting of the whole plate from a camera image, with every counted colony marked on screen for the analyst to check and correct; the plate image is saved with the result. (with the optional module)
  • Very small colonies: detects colonies down to 0.1 mm diameter (0.05 mm micro-colonies with the high-resolution counter). (with the optional module)
  • Zone of inhibition diameters around antibiotic discs, wells or cylinders, +/-0.5 mm, for antibiotic potency assays (USP <81>). (with the optional module)
  • Separate counts of up to 4 colony colours on the same chromogenic plate. (with the optional module)
  • Spiral-plated dishes: counts by ring or sector and calculates CFU/mL automatically. (with the optional module)
  • Tamper-evident records: user log-ins, audit trail of every manual colony edit and electronic review of results (21 CFR Part 11, EU GMP Annex 11). (with the optional module)

Not covered: Growing the colonies: plates must first be incubated at the right temperature in a microbiological incubator.; Identifying what the organisms are (Gram reaction, species) or testing antibiotic susceptibility needs a microscope and a microbial identification / susceptibility system.; Making the plates: pouring media needs a laminar air-flow bench or media preparator, and spiral plates need a spiral plater; the spiral module only counts them.; Counting cells in liquids (cell suspensions, turbidity) needs a haemocytometer with a microscope, a cell counter or a spectrophotometer.

9 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety standard for laboratory equipment, including the counter, its lighting and power supply.
  • IEC 61326-1 EMC requirements for laboratory equipment so the camera and electronics work reliably next to other lab instruments.
  • USP <61> Microbial enumeration tests (plate count and membrane filtration) on non-sterile products; defines the plates the counter must count and the countable range.
  • Ph. Eur. 2.6.12 European Pharmacopoeia microbial enumeration tests, harmonised with USP <61>; used with BP monographs in Pakistan.
  • USP <81> Microbial assays of antibiotics; the cylinder-plate method relies on accurate zone-of-inhibition diameter measurement.
  • ISO 7218 General requirements for microbiological examinations of food, including rules for counting colonies and calculating results.
  • ISO 4833-1 Horizontal method for enumerating microorganisms by colony count at 30 deg C using the pour-plate technique.
  • ISO 11737-1 Determination of microbial population (bioburden) on medical devices, including membrane filtration and plate counting.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures; applies to counting software whose results are used for release.
  • EU GMP Annex 11 EU/PIC/S GMP requirements for computerised systems such as automated counters used in QC.
  • USP <1058> Analytical instrument qualification framework (DQ/IQ/OQ/PQ) used to accept the counter.
  • ISO/IEC 17025:2017 Accreditation standard for testing labs; requires verified equipment and controlled, traceable records.

Common mistakes in tenders for this equipment

  • Copying one brand's numbers (exact camera megapixels, a specific number of colours or lighting combinations) so only that brand complies; specify the minimum detectable colony size and the plate formats you use instead.
  • Accepting a "Part 11 compliant" claim on paper without a live demonstration that manual colony additions/removals are audit-trailed and original images cannot be deleted.
  • Buying an automated counter without verification plates, dish holders for contact plates and membranes, or a PC that meets the software requirement, so it cannot be qualified or used on day one.
  • Ordering zone-of-inhibition measurement without a traceable reference scale or calibration, so potency results cannot be defended.
  • Using a manual pen counter for GMP or DTL release counts, leaving no image to re-check when results are challenged.
  • Leaving out yearly verification and software updates, so the counter runs on an unsupported operating system and drifts without anyone noticing.
  • Asking for DRAP medical-device registration or ISO 13485: a colony counter is laboratory equipment, so these demands only confuse bidders.