Colposcope (binocular optical, optical with integrated camera, or digital video) with LED illumination, stepped or zoom magnification, green filter, mobile stand and image documentation
A colposcope is a lighted magnifier on a stand that lets a gynaecologist examine the cervix closely after a positive VIA/VILI, HPV test or Pap smear, choose where to take a biopsy and guide treatment such as LEEP. In Pakistan it is bought for DHQ and teaching hospital gynaecology departments, cancer hospitals, cervical screening programmes and private clinics. The key tender decisions are optical versus video, magnification range and working distance, LED light and green filter, image capture and reporting, and the specula and biopsy instruments needed to use it.
What this equipment will do
- Women referred after positive VIA/VILI, HPV test or abnormal Pap smear: magnified examination of the cervix, vagina and vulva (WHO 2020 technical guidance for cervical precancer devices).
- Magnified view in steps or zoom from low power (about 2-4.5x) to at least 12x, at a working distance that leaves room for instruments.
- Green (red-free) filter to show abnormal blood vessel patterns.
- Assessment of the cervix after acetic acid and Lugol's iodine to decide on biopsy or treatment.
- Higher magnification of at least 15x for fine vessel detail. (with the optional module)
- Bright LED light of at least 20,000 lux at 300 mm for optical viewing and image capture. (with the optional module)
- Live image on a monitor for teaching and patient counselling. (with the optional module)
- Saved still images and video, patient records and printed reports. (with the optional module)
- Structured colposcopy report with lesion map and scoring. (with the optional module)
- Export of images and reports to PACS or the hospital information system (DICOM / HL7). (with the optional module)
- Guided punch biopsy and endocervical sampling with the supplied instrument set. (with the optional module)
- Colposcopic guidance of loop excision (LEEP / LLETZ) with insulated specula. (with the optional module)
- Battery-powered screening in outreach camps with wireless image transfer. (with the optional module)
Not covered: HPV testing and Pap smear reading are not done by the colposcope; they need an HPV DNA test system (real-time PCR) and a cytology microscope.; Treatment of lesions needs a separate electrosurgical unit with smoke evacuator for LEEP, or a thermal ablation or cryotherapy unit; the colposcope only guides it.; The final diagnosis of a biopsy is made in a histopathology laboratory, not by the colposcope.; Looking inside the uterus needs a hysteroscope; the colposcope sees only the cervix, vagina and vulva.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including the LED power supply, camera and stand stability.
- IEC 60601-1-2 Electromagnetic compatibility, so the camera and light work near electrosurgical (LEEP) units and do not disturb other equipment.
- IEC 60601-2-41 Particular requirements for surgical luminaires and luminaires for diagnosis (2021 edition), declared by leading makers for the colposcope light source.
- ISO 14971 Risk management, covering hazards such as excessive light, stand tipping, wrong patient images and data loss.
- ISO 13485 Quality management system of the manufacturer.
- IEC 62304 Software life-cycle for the image capture, documentation and reporting software of camera-equipped and video colposcopes.
- IEC 62353 Electrical safety testing at installation, after repair and at preventive maintenance.
- WHO technical guidance and specifications of medical devices for screening and treatment of precancerous lesions (2020) WHO reference technical specifications for colposcopes and related cervical screening and treatment devices for public procurement.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Copying one brand's brochure: exact magnification values (for example 3.75x/7.5x/15x), a named filter technology or a fixed number of LEDs. Use ranges such as 'from not more than 4.5x to not less than 12x' and 'not less than 20,000 lux at 300 mm'.
- Mixing optical and digital magnification: video colposcopes may quote very high total magnification that is mostly digital enlargement; ask for optical zoom range and working distance separately.
- Buying a video colposcope without computer, monitor, licensed software and printer, so it cannot save or print images when it arrives.
- Leaving out specula and biopsy instruments, so the clinic cannot start work; these cost little but delay use by months in public hospitals.
- Accepting halogen light sources that need regular bulb changes and heat the tissue; LED is standard on current colposcopes.
- Ignoring training: a colposcope is only as good as the colposcopist; require hands-on training and link the purchase to the provincial cervical screening training programme.
- Not checking local service for cameras and software; a broken camera with no local engineer stops image recording for months.