Specification builder · Tier A

Benchtop conductivity / resistivity / TDS / salinity meter with temperature compensation (switchable OFF) for pharmaceutical water testing to USP <645> / Ph. Eur. 2.2.38

A conductivity meter measures the dissolved ionic content of water and solutions with a cell of known cell constant, and reports conductivity, resistivity, TDS or salinity at a reference temperature. Pharma QC labs, drug testing labs, water and environment labs, hospitals and universities use it for purified water and WFI, RO and raw water, cleaning-validation rinse water and buffers. The buying decisions are the measuring range and accuracy, the cell types (a certified low-constant cell for pure water), whether temperature compensation can be switched off for the pharmacopoeial test, and how results are recorded.

What this equipment will do

  • Samples: purified water, water for injection (WFI), RO and deionised water, cleaning-validation rinse water, raw and drinking water, buffers, reagents and other aqueous solutions.
  • Brines, seawater and concentrated process solutions up to about 500 mS/cm or higher. (with the optional module)
  • Conductivity and TDS with temperature, corrected to 25 deg C by automatic temperature compensation (ISO 7888, ASTM D1125).
  • Salinity, and resistivity (MOhm.cm) or 0.001 uS/cm resolution for high-purity water. (with the optional module)
  • Pharmaceutical water conductivity test Stage 1 and Stage 2: uncompensated conductivity read with the temperature and compared with the pharmacopoeial table (USP <645>, Ph. Eur. 2.2.38). (with the optional module)
  • Low-conductivity water measured with a certified 0.1 /cm cell in a flow-through or closed vessel, away from air. (with the optional module)
  • Stage 2 measurement with the sample held at 25 +/- 1 deg C. (with the optional module)
  • Guided three-stage test with automatic pass/fail report. (with the optional module)
  • Ultrapure (Type I) water down to about 0.055 uS/cm with a 0.01 /cm cell. (with the optional module)
  • Cell constant calibration with certified KCl standards (1 point; 3 or more points with multi-point calibration).
  • Verification of the meter electronics with certified resistors, as the pharmacopoeia requires. (with the optional module)
  • Checks at water-system sampling points with a portable meter. (with the optional module)
  • Continuous Stage 1 monitoring of a purified water or WFI loop with alarm and 4-20 mA output. (with the optional module)
  • GLP records of results and calibrations with date, time and settings, printed or exported.
  • Electronic records with user login, audit trail and electronic signatures (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)

Not covered: pH, including Stage 3 of the pharmacopoeial water conductivity test (which needs a pH reading): this needs a pH meter.; Total organic carbon (TOC) of purified water and WFI (USP <643>, Ph. Eur. 2.2.44) needs a TOC analyser.; Microbial count and bacterial endotoxins of water need the microbiology lab and an endotoxin (LAL) test system.; Individual ions such as chloride, nitrate or heavy metals are not identified; these need ion chromatography, a photometer or AAS/ICP.

9 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical and fire safety of laboratory measuring equipment, including the meter, stirrer, bath and any monitor.
  • IEC 61326-1 Electromagnetic compatibility of laboratory electrical equipment; low-level conductivity readings are sensitive to interference.
  • USP <645> Pharmacopoeial water conductivity test (three stages) for purified water and WFI, with requirements for cell constant (+/-2%), temperature accuracy, meter verification and uncompensated readings.
  • Ph. Eur. 2.2.38 European Pharmacopoeia conductivity method used with the purified water and WFI monographs (BP equivalent), harmonised in approach with USP <645>.
  • USP <1644> USP general information chapter on theory and practice of conductivity measurement of solutions, cell constants and temperature compensation.
  • USP <1231> USP guidance on water for pharmaceutical purposes, including how conductivity fits into water-system monitoring.
  • ISO 7888 Water quality method for electrical conductivity, including temperature correction for natural water, used by water and environment labs.
  • ASTM D1125 Standard test methods for electrical conductivity and resistivity of water, including cell constant calibration with KCl solutions.
  • ISO 17034 Competence of reference material producers; certified KCl conductivity standards should come from an accredited producer.
  • ISO/IEC 17025 Competence of calibration laboratories; calibration certificates for the meter, cell and thermometer should be accredited.
  • USP <1058> Framework for analytical instrument qualification (IQ/OQ/PQ) of the meter in GMP labs.
  • 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and electronic signature expectations where conductivity results are kept electronically for GMP release.

Common mistakes in tenders for this equipment

  • Copying one brand's figures (a very large memory size, an exact range top such as 3000 mS/cm, a named intelligent-sensor or guided-test feature) so that only one model qualifies; use 'not less than' limits.
  • Asking for a 'USP <645> certified meter'. No such certification exists; ask for compensation that can be switched OFF, a certified 0.1 /cm cell, certified resistors and a 25 deg C bath.
  • Buying only a general 1 /cm cell and measuring purified water in an open beaker, so air carbon dioxide and cell polarisation give false failures.
  • Leaving out certified KCl standards, a low-level standard for the water cell and certified resistors, so the meter cannot be qualified after installation.
  • Treating the in-line monitor and the portable meter as part of the bench meter; they are separate items with different makers and should be optional lines.
  • Not asking for a cell price list and 7-year cell supply; the cell is the part that breaks and some models use cells with special connectors.
  • Demanding DRAP medical device registration or ISO 13485 for a laboratory conductivity meter, which does not apply and wrongly excludes bidders.
  • Accepting calibration certificates from non-accredited labs for an ISO/IEC 17025 drug testing laboratory.