Container closure integrity tester (CCIT): non-destructive vacuum decay leak tester for vials, prefilled syringes and sterile device packs (ASTM F2338, USP <1207>, ISO 11607), with optional pressure decay, high voltage leak detection and dye / bubble test kits (ASTM F1929, F2096)

A container closure integrity tester checks that sterile packages - injectable vials, ampoules, prefilled syringes and sterile device pouches - have no leaks that could let microbes in. It places the pack in a vacuum chamber and measures the tiny pressure rise from escaping gas or liquid, giving a non-destructive pass / fail result. It is bought by drug and device testing laboratories, pharmaceutical QC labs and local device makers; dye and bubble kits cover the simpler tests used for porous device pouches.

What this equipment will do

  • Product types: rigid containers - glass and plastic vials (liquid-filled and freeze-dried), ampoules, prefilled syringes, cartridges and bottles (ASTM F2338-24, USP <1207>).
  • Product types: non-porous flexible packs - sterile device pouches, sachets and bags including IV bags, using a flexible-package chamber (ASTM F2338-24). (with the optional module)
  • Non-destructive vacuum decay leak test with automatic pass / fail; the tested pack can be returned to stock or retested (ASTM F2338-24).
  • Detection of holes down to about 10 um (gross and mid-size leaks). (with the optional module)
  • Detection of holes down to 5 um, the usual target for injectables (USP <1207>). (with the optional module)
  • Detection of holes down to 1-2 um for high-value products. (with the optional module)
  • Method validation and daily checks with certified positive and negative controls (USP <1207>). (with the optional module)
  • Pressure decay leak test for containers not suited to vacuum. (with the optional module)
  • High voltage leak detection of pinholes and cracks in liquid-filled syringes, vials and ampoules (USP <1207>). (with the optional module)
  • Seal channel leaks in porous (paper or spunbonded polyolefin) device pouches by dye penetration (ASTM F1929-23). (with the optional module)
  • Gross leaks in pouches and trays by internal pressurisation bubble test (ASTM F2096-11(2019)). (with the optional module)
  • Audit-trailed, e-signed electronic results for batch release (21 CFR Part 11). (with the optional module)

Not covered: Seal strength (peel) and burst / creep strength of pouches need the package seal strength tester (ASTM F88, ASTM F1140).; Accelerated ageing of packs before integrity testing needs the environmental test chamber (ASTM F1980).; Sterility, bacterial endotoxins and microbial ingress (challenge) tests need the sterility testing isolator, endotoxin reader and microbiology lab.; Leak tests of the devices themselves (infusion sets, catheters, connectors) need the device pressure and vacuum leak tester.

7 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • ASTM F2338-24 Defines the non-destructive vacuum decay leak test this instrument performs on rigid, semi-rigid and flexible packages, including use of reference defects to show detection capability.
  • USP <1207> USP chapters on package integrity evaluation of sterile products: choice of deterministic methods (vacuum decay, pressure decay, high voltage), method validation with positive controls, and the move away from dye and bubble tests for injectables.
  • ISO 11607-1:2019 Requirements for packaging of terminally sterilized medical devices, including demonstration that the sterile barrier system stays intact.
  • ASTM F1929-23 Dye penetration test for seal leaks in porous medical packaging (medical paper or spunbonded polyolefin pouches).
  • ASTM F2096-11(2019) Internal pressurisation bubble test for gross leaks in packaging, sensitive to holes of about 250 um.
  • EU GMP Annex 1 (2022) EU / PIC/S rules for sterile medicinal products, requiring validated container closure integrity testing of sterile containers.
  • IEC 61010-1 Electrical safety standard for laboratory and test equipment.
  • IEC 61326-1 EMC requirements for laboratory and test equipment.
  • ISO/IEC 17025:2017 Accreditation standard for testing and calibration laboratories; requires calibrated, traceable equipment and controlled records.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures, used as the benchmark for audit trails and e-signatures in test software.

Common mistakes in tenders for this equipment

  • Accepting a sensitivity figure from a brochure without requiring a demonstration on the buyer's own containers with certified positive controls.
  • Buying a tester with a chamber for one container size only; every vial, syringe and pouch format needs its own close-fitting chamber or insert.
  • Specifying one maker's exact vacuum level, chamber volume or test time so only that brand can comply; state the required detection limit and package types instead.
  • Buying only dye or bubble kits and calling it container closure integrity testing; USP <1207> prefers deterministic methods such as vacuum decay for sterile injectables.
  • Expecting vacuum decay to test porous device pouches directly; porous lids need the dye penetration test or a method validated for them.
  • Asking for high voltage, pressure decay and 1 um sensitivity 'just in case', which narrows the field to one or two makers and can double the price.
  • Forgetting method development on the buyer's products, positive controls and clean dry air, so the instrument sits unused after installation.
  • Asking for DRAP medical-device registration or ISO 13485 for the tester itself: it is laboratory equipment, so this only confuses bidders.