Specification builder · Tier B
Tablet and capsule disintegration tester per USP <701> / Ph. Eur. 2.9.1 / BP, 1-4 basket stations, manual or automatic end-point detection
A disintegration tester checks whether tablets and capsules break up within the time set in the pharmacopoeia by moving them up and down in warm medium (37 deg C) in a 6-tube basket. Pharma QC labs use it for routine batch release, drug testing labs use it to check products against USP and BP monographs, and teaching labs use it for practicals. Units range from simple timer-based testers to models that detect the end-point in each tube automatically.
What this equipment will do
- Dosage forms: tablets and hard and soft capsules, 6 units per basket (USP <701>, Ph. Eur. 2.9.1 Test A).
- Large tablets and capsules too big for the standard tubes: 3-tube Test B basket (Ph. Eur. 2.9.1 Test B). (with the optional module)
- Dietary supplement, herbal and nutraceutical tablets and capsules (USP <2040>).
- Disintegration time test: basket moved through a 53-57 mm stroke at 29-32 cycles per minute in medium at 37 +/- 2 deg C, timed and judged by the analyst (USP <701>, BP Appendix XII A1; interchangeable per ICH Q4B Annex 5).
- Tests with discs where the monograph asks for them (USP <701> / Ph. Eur. 2.9.1 discs).
- Several products or batches at the same time: 2 to 4 baskets, each in its own beaker. (with the optional module)
- Independent start, stop, timing and basket lift at each station. (with the optional module)
- Medium temperature measured in the test beaker and recorded on the test report. (with the optional module)
- Automatic end-point: disintegration time detected and recorded for each tube. (with the optional module)
- Disintegration curve: progress of each tube recorded during the test. (with the optional module)
- Printed test report with date, product / batch, frequency, temperature and result. (with the optional module)
- Electronic records with audit trail and electronic signatures (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)
Not covered: How fast the drug dissolves (dissolution rate and profiles) needs a dissolution tester.; Tablet hardness, friability and thickness need a tablet hardness tester and a friabilator.; Disintegration of suppositories and pessaries (Ph. Eur. 2.9.2) needs a separate suppository disintegration apparatus.; Assay, content uniformity and impurities need an HPLC or UV-Vis spectrophotometer.
Loading the builder...
Standards this specification draws on
- USP <701> Disintegration Defines the basket-rack assembly, discs, beaker, stroke 53-57 mm, 29-32 cycles per minute and 37 +/- 2 deg C medium.
- Ph. Eur. 2.9.1 / BP Appendix XII A1 European and British pharmacopoeia disintegration test, harmonised with USP for Test A and adding Test B for large tablets and capsules.
- ICH Q4B Annex 5(R1) Confirms the USP, Ph. Eur. and JP disintegration chapters are interchangeable, so one tester serves all three.
- USP <2040> Disintegration and Dissolution of Dietary Supplements Disintegration conditions for dietary supplement and herbal tablets and capsules.
- USP <1058> Analytical Instrument Qualification Framework for IQ / OQ / PQ of the instrument.
- 21 CFR Part 11 / EU GMP Annex 11 / PIC/S PI 041 Electronic records, audit trail and data integrity expectations for GMP labs inspected by DRAP.
- IEC 61010-1 Electrical safety of laboratory equipment.
- IEC 61326-1 Electromagnetic compatibility (EMC) of laboratory equipment.
- ISO 9001 Quality management system of the manufacturer.
Common mistakes in tenders for this equipment
- Copying one brand's brochure: proprietary names for wireless or sensor baskets, an exact stroke frequency range or a basket count that only one maker offers.
- Writing only 'USP compliant' without requiring stroke, frequency, medium temperature and timer to be verified at every station at installation with certified tools.
- Accepting bath temperature accuracy as if it were medium temperature; the requirement is 37 +/- 2 deg C inside each beaker during the test.
- Forgetting the Ph. Eur. Test B basket for large tablets, then finding that BP / Ph. Eur. monographs cannot be followed for those products.
- Not asking for spare discs, mesh screens and beakers; these are maker-specific and a broken beaker or worn discs can stop testing for weeks.
- Asking for full 21 CFR Part 11 software on a simple disintegration tester when printouts are the GMP record, which cuts competition to one or two makers.
- Awarding to an offer whose country of origin cannot currently be imported into Pakistan (Indian-origin equipment has faced trade restrictions since 2019), causing cancellation and re-tendering.
- Asking for DRAP medical device registration, which does not apply to a laboratory disintegration tester and wrongly excludes valid bidders.