Side-viewing video duodenoscope for ERCP with forceps elevator (reusable with removable or single-use distal component, or fully single-use), compatible video processor, and validated duodenoscope reprocessing
A duodenoscope is a side-viewing flexible video endoscope with a small lever (elevator) at the tip, used by gastroenterologists for ERCP to remove bile duct stones, place stents in blocked bile and pancreatic ducts and take samples. Teaching, tertiary, larger DHQ and private hospitals buy it as a scope for an existing GI video processor or as a complete system with processor and reprocessing equipment. The elevator is hard to clean and has been linked to outbreaks, so the key decisions are the tip design (removable cap, single-use tip part or fully single-use scope) and validated reprocessing; reusable scopes work only with the same maker's processor.
What this equipment will do
- Patients: adults needing ERCP, with a side-viewing duodenoscope having a channel of 4.2 mm or more for stents, balloons and baskets (IEC 60601-2-18, ISO 8600-1).
- Patients in whom a standard duodenoscope does not pass, with a slim duodenoscope having a 3.2 mm channel. (with the optional module)
- Side view of the papilla and steering of catheters and guidewires into the bile or pancreatic duct with the elevator at the scope tip.
- Holding the guidewire in place with a guidewire lock while catheters and stents are exchanged. (with the optional module)
- Therapeutic ERCP: bile duct stone removal with balloons and baskets, mechanical lithotripsy, and plastic stent placement, with the first-year accessory set. (with the optional module)
- Sphincterotomy (cutting the papilla) with an electrosurgical generator (IEC 60601-2-2). (with the optional module)
- Brush cytology sampling of bile duct strictures.
- Still images and video of the procedure recorded, with a printed ERCP report.
- Patient worklist from the hospital system and image storage in PACS (DICOM). (with the optional module)
- Throw-away tip part (distal cover or elevator cap) replaced for every patient, so the elevator can be fully cleaned. (with the optional module)
- Single-use duodenoscopes for selected patients with no reprocessing needed. (with the optional module)
- Automated cleaning and high-level disinfection of reusable duodenoscopes with leak test and a record of every cycle (ISO 15883-4, ISO 17664-1). (with the optional module)
- Dry storage of cleaned scopes in a filtered-air cabinet (EN 16442). (with the optional module)
- Periodic culture tests of reprocessed duodenoscopes to check that cleaning works. (with the optional module)
Not covered: X-ray screening of the ducts during ERCP needs a C-arm or fluoroscopy unit with radiation protection; the duodenoscope system does not include it.; Looking directly inside the bile duct (cholangioscopy) needs a separate cholangioscope system passed through the duodenoscope.; Endoscopic ultrasound (EUS) needs an echoendoscope with an ultrasound processor; a duodenoscope has no ultrasound.; Gastroscopy and colonoscopy need forward-viewing gastroscopes and colonoscopes; a duodenoscope is not used for routine examinations.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including the processor, light source and trolley system.
- IEC 60601-1-2 Electromagnetic compatibility, so the image is not disturbed by the electrosurgical generator used for sphincterotomy.
- IEC 60601-2-18 Particular safety requirements for endoscopic equipment: light and tip heating, insulation of the insertion tube and use with high-frequency surgery.
- IEC 60601-2-2 Particular safety requirements for the electrosurgical generator used for sphincterotomy.
- ISO 8600-1 General requirements for medical endoscopes, so channel width, outer diameters and viewing direction are stated and measured the same way by every maker.
- ISO 15883-4 Requirements and tests for washer-disinfectors using chemical disinfection for thermolabile endoscopes (read with ISO 15883-1).
- ISO 17664-1 Information the manufacturer must give for processing critical and semi-critical devices, so the cleaning of the elevator and distal end is validated and written down.
- EN 16442 Controlled-environment storage and drying cabinets for processed thermolabile endoscopes.
- ANSI/AAMI ST91 Guidance on processing flexible endoscopes in health care facilities: point-of-use treatment, leak testing, cleaning, high-level disinfection or sterilisation, storage and quality control.
- ISO 14971 Risk management for medical devices, covering cross-infection, loosening of tip parts, image loss and electrical hazards.
- ISO 13485 Quality management system of the device manufacturer.
- DRAP Medical Devices Rules 2017 Legal requirement for registration (Class B-D) or enlistment (Class A) of medical devices imported or sold in Pakistan.
Common mistakes in tenders for this equipment
- Accepting an old duodenoscope design with a fixed, glued end cap, or refurbished scopes of such designs. The US FDA asked makers in 2019 to withdraw fixed-endcap models; require a removable or single-use distal component.
- Copying one brand's brochure: an exact distal-end diameter to 0.1 mm, a trademarked cap or elevator name, or a 4.5 mm channel only one model has. Specify ranges such as 'channel not less than 4.2 mm, insertion tube not more than 11.6 mm'.
- Leaving out the running cost of single-use distal components or single-use scopes, so the scope sits idle when the first stock runs out. Fix 5-year unit prices and evaluate cost per procedure.
- Buying a washer-disinfector that the duodenoscope maker has not validated for that duodenoscope model and its elevator, and no elevator brushes or flushing adaptors, so the hardest part to clean is never cleaned properly.
- Buying a duodenoscope without checking that it runs on the hospital's existing processor, or buying processor and scopes in separate tenders from different makers.
- Forgetting the rest of the ERCP room: a C-arm or fluoroscopy unit with lead protection, an electrosurgical generator for sphincterotomy and first-year accessories (sphincterotomes, guidewires, balloons, baskets, stents).
- Warranty and CMC that exclude reusable duodenoscopes or treat every elevator or tip fault as misuse, with no fixed price list for minor, medium and major repairs.
- Accepting DRAP registration of the processor only; every duodenoscope, single-use scope, single-use tip part and washer-disinfector must be registered or enlisted.