Specification builder · Tier D

Extracorporeal membrane oxygenation (ECMO) system for adult and paediatric VV and VA support, with centrifugal pump console, emergency back-up drive, heater, gas blender and transport capability

An ECMO system takes blood out of the body, pumps it through a membrane oxygenator that adds oxygen and removes carbon dioxide, and returns it, supporting the lungs (VV ECMO) or both heart and lungs (VA ECMO) for days to weeks. It is bought by tertiary cardiac centres, teaching-hospital ICUs and children's hospitals in Pakistan for severe respiratory failure, cardiogenic shock and after cardiac surgery. The console is only part of the cost: maker-specific circuits and oxygenators, a trained team and 24/7 clinical support decide whether the programme works.

What this equipment will do

  • Patients: adults on VV ECMO for severe respiratory failure, or VA ECMO for cardiogenic shock, cardiac arrest (ECPR) and after heart surgery, with blood flow up to 7 L/min or more from a centrifugal pump (ISO 18242; ELSO guidelines).
  • Children from about 10 kg: paediatric circuits with stable flow down to 0.5 L/min. (with the optional module)
  • Newborns from about 2 kg: neonatal circuits with flow down to 0.1 L/min. (with the optional module)
  • Gas exchange and warming: membrane oxygenator with heat exchanger, air-oxygen blender (FiO2 21-100%) with sweep gas up to 10 L/min, and water heater 35-39 deg C (ISO 7199, ISO 15676).
  • Circuit safety monitoring: blood flow, air-bubble detection, at least 2 circuit pressures and temperature, with prioritised alarms and an alarm log (IEC 60601-1-8).
  • Oxygenator watch: 4 circuit pressures with transmembrane (delta) pressure alarm and 24-hour trends of flow, speed and pressures. (with the optional module)
  • Continuous in-line venous oxygen saturation and haemoglobin/haematocrit on the console. (with the optional module)
  • Cooling down to 15 deg C for controlled hypothermia and temperature management after cardiac arrest. (with the optional module)
  • Power and pump failure: battery for 60 minutes or more of full operation and an independent emergency back-up drive or hand crank to restore flow within 60 seconds (ELSO circuit guidelines).
  • Moving the patient within the hospital: compact console of 12 kg or less, or a transport frame/bed mount carrying the whole system. (with the optional module)
  • Road and air ambulance transfer between hospitals with 12/24 V DC power and ambulance mounting (IEC 60601-1-12). (with the optional module)
  • Low-flow CO2 removal (blood flow 0.3-1.5 L/min) on the same console. (with the optional module)

Not covered: Full cardiopulmonary bypass during open-heart surgery (venous reservoir, cardiotomy suction, cardioplegia) needs a heart-lung machine.; The patient still needs an ICU ventilator; ECMO takes over gas exchange but does not ventilate the lungs.; Blood gases, lactate and clotting tests on patient blood need a blood gas analyser and a coagulation or ACT analyser (an ACT analyser can be ordered as an add-on).; Kidney support (CRRT or dialysis) needs a CRRT or haemodialysis machine.

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of the console, pump drive and heater as medical electrical equipment.
  • IEC 60601-1-2 Electromagnetic compatibility so the pump and alarms are not disturbed by diathermy, defibrillators and phones.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and signals for flow, pressure and bubble alarms.
  • IEC 60601-1-12 Requirements for equipment used in the emergency medical services (ambulance) environment, for transport consoles.
  • ISO 7199 Performance and safety of blood-gas exchangers (oxygenators): gas transfer, blood damage, leaks and heat exchange.
  • ISO 18242 Performance and safety of centrifugal blood pumps, including haemolysis and flow testing.
  • ISO 15676 Requirements for single-use tubing packs for cardiopulmonary bypass and ECMO.
  • IEC 62304 Software life-cycle for the console control and alarm software.
  • ISO 13485 Quality management system of the console and circuit manufacturers.
  • DRAP Medical Devices Rules, 2017 Registration/enlistment required to import and sell the console and its consumables in Pakistan.
  • ELSO Guidelines for Adult and Pediatric ECMO Circuits (2022) International ECMO society guidance on circuit components, monitoring, back-up and safety features.
  • ELSO Narrative Guideline on ECMO Training and Continuing Education (2025) International guidance on training and maintaining competence of the ECMO team.

Common mistakes in tenders for this equipment

  • Writing the console specification around one maker's integrated design (exact weight, sensor count, in-line saturation measurement or proprietary control modes) so only one bidder complies.
  • Buying the console without pricing consumables: circuit sets, oxygenators and pump heads are maker-specific and cost USD 2,500-6,000 each, so a cheap console can become the most expensive choice and sit idle when sets run out.
  • Not checking that oxygenators, circuit sets and cannulae (not just the console) are registered with DRAP, so life-saving sets are held at customs.
  • Leaving out the emergency hand crank or back-up drive, spare flow sensor and a loan-console clause; a pump stop without back-up is fatal within minutes.
  • Treating training as a one-day demonstration: ECMO needs a trained multidisciplinary team, water-drill simulation and clinical support for the first runs, or complications and an unused console follow.
  • Ignoring heater-cooler water hygiene; contaminated heater water has caused mycobacterial infections after cardiac surgery worldwide.
  • Assuming support lasts for the life of the console: one maker wound down its ECMO console business in 2024 with service ending within a year, so demand a 10-year parts and set-supply undertaking with notice before end of support.
  • Demanding the longest labelled oxygenator duration that only one maker's coating claims, instead of a stated minimum and the validated duration in the instructions for use.