Specification builder · Tier B
Fully automatic benchtop ion-selective electrode (ISE) electrolyte analyser for sodium, potassium and chloride, with options for ionised calcium, pH, lithium and total CO2; manual sample presentation as basic, autosampler tray as standard
A small benchtop analyser that measures sodium, potassium and chloride (and, in some models, ionised calcium, pH and lithium) in blood, serum, plasma and urine within about a minute, using ion-selective electrodes. It is bought by THQ and DHQ hospital labs, ICUs, dialysis centres and private labs in Pakistan, often as the main electrolyte instrument or as a backup to the chemistry analyser. It is an IVD medical device; its electrodes and reagent packs fit only the maker's own analyser, so running cost matters more than the purchase price.
What this equipment will do
- Samples: heparinised whole blood, serum and plasma from syringes, tubes, cups or capillaries, 200 uL or less, with clot and air-bubble detection.
- Urine (and other fluids the maker has validated) for Na, K and Cl, with separate urine measuring ranges. (with the optional module)
- Sodium, potassium and chloride from one aspiration by direct ISE: Na 100-180, K 1.5-10.0, Cl 60-150 mmol/L, result within 90 seconds, precision checked at installation (CLSI EP15-A3, CLSI EP09c).
- Ionised calcium 0.3-3.0 mmol/L in the same aspiration. (with the optional module)
- pH and pH-normalised calcium (nCa at pH 7.4). (with the optional module)
- Lithium 0.2-3.0 mmol/L for patients on lithium therapy. (with the optional module)
- Total CO2 (bicarbonate) with calculated anion gap. (with the optional module)
- Ionised magnesium 0.2-1.5 mmol/L. (with the optional module)
- Dialysis fluid sodium and potassium in a validated dialysate mode. (with the optional module)
- Unattended batches on an autosampler with 10 or more positions and an emergency (STAT) position. (with the optional module)
- Internal quality control stored for 2-3 levels with mean, SD, CV and Levey-Jennings chart (ISO 15189). (with the optional module)
- Results sent to the LIS/HIS, one-way or bidirectional by barcode (CLSI LIS1/LIS2, HL7 v2.x). (with the optional module)
Not covered: Blood gases (pO2, pCO2), lactate, glucose and haemoglobin need a blood gas analyser; the pH option here measures pH only.; Total calcium, phosphate, total magnesium, urea and creatinine need the clinical chemistry analyser.; Drug levels other than lithium (for example digoxin or antiepileptics) need an immunoassay analyser.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical, mechanical and fire safety for laboratory equipment, including IVD analysers.
- IEC 61010-2-101 Particular safety requirements for IVD medical equipment such as electrolyte analysers (biohazardous samples, probes, labelling).
- IEC 61326-2-6 EMC requirements for IVD equipment, so millivolt-level electrode signals are not disturbed by mains noise and nearby equipment.
- ISO 13485:2016 Quality management system standard for the manufacturer of the analyser, electrodes and reagents.
- ISO 14971:2019 Risk management for medical devices, including IVD analysers and their software.
- ISO 15189:2022 Medical laboratory accreditation standard; requires verification of performance, internal QC and EQA before patient use.
- CLSI EP15-A3 Protocol a lab uses to verify the manufacturer's precision and trueness claims at installation.
- CLSI EP09c Method comparison and bias estimation using patient samples, used to check agreement with the lab's chemistry analyser.
- CLSI LIS1/LIS2 (ASTM E1381/E1394) and HL7 v2.x Open communication protocols between analysers and laboratory information systems.
- DRAP Medical Devices Rules 2017 Pakistan's legal framework for registration / enlistment of medical devices and IVDs, including analysers and reagents.
Common mistakes in tenders for this equipment
- Copying one model's exact sample volume, analysis time or electrode list (for example '95 uL' or '35 seconds') into the tender, which excludes other makers without any clinical benefit; set maximum limits instead.
- Comparing only the analyser price: electrodes (replaced every 6-12 months) and reagent packs (used during calibration and standby even with few samples) decide the real cost. Evaluate cost per reportable sample over 5 years at the lab's real workload.
- Excluding electrodes from the warranty as 'consumables' without a stated life, so the buyer pays for electrodes that fail after a few weeks in hot labs.
- Ordering lithium, pH or TCO2 'for completeness' when nobody will request them; each extra electrode still needs calibration, controls and replacement, and fewer makers can bid.
- Placing the analyser in a room without air-conditioning; most ISE analysers are specified for about 15-32 deg C, and summer temperatures in Pakistan cause drift, failed calibrations and short electrode life.
- Ignoring sample handling for ionised calcium and potassium: wrong heparin, delayed or haemolysed samples, and tourniquet or fist clenching give false results that no analyser can correct.
- Asking for 'maintenance-free electrodes' without defining a guaranteed life and free replacement, so the claim cannot be enforced.
- Not checking DRAP enlistment of the reagent packs and electrodes as well as the analyser, which leads to stock-outs at customs after installation.
