Specification builder · Tier B
Automated endoscope reprocessor (endoscope washer-disinfector) for flexible endoscopes: single-basin, two-endoscope and pass-through models with leak test, channel monitoring, chemical high-level disinfection, treated rinse water, drying and cycle traceability (ISO 15883-4)
An automated endoscope reprocessor (AER) is a washer-disinfector that leak-tests, cleans, disinfects, rinses and air-purges flexible endoscopes after each patient in a closed, recorded cycle. Public and private hospitals with GI, bronchoscopy or urology endoscopy buy it to replace manual soaking, which depends on staff discipline. The key tender decisions are capacity, monitoring of every channel, the disinfectant and its running cost, rinse water treatment, traceability and proof that the machine is validated for the hospital's own endoscopes.
What this equipment will do
- Endoscopes handled: flexible endoscopes of the makes and models listed in the tender, one per cycle in a single-basin machine (ISO 15883-4).
- Two flexible endoscopes per cycle, in two basins or one large chamber. (with the optional module)
- Pass-through use between a dirty loading room and a clean unloading room, with interlocked doors. (with the optional module)
- Leak test of the endoscope in every cycle, with alarm if the endoscope is damaged.
- Cleaning of all channels and the outside of the endoscope with detergent, followed by chemical high-level disinfection with a validated disinfectant (ISO 15883-4).
- Check of flow or pressure in each channel, with alarm when a channel is blocked or not connected. (with the optional module)
- Peracetic-acid-based (aldehyde-free) disinfection. (with the optional module)
- Rinsing with treated (filtered or disinfected) water and self-disinfection of the machine's own pipes.
- Rinse water through a 0.2 um or finer bacteria-retentive filter that is itself disinfected by the machine. (with the optional module)
- Heat (thermal) self-disinfection of the machine's chamber, pipes and filters. (with the optional module)
- Air purge of all endoscope channels with filtered air at the end of the cycle.
- Alcohol flush of the channels before the final air purge. (with the optional module)
- Drying of the channels and outside of the endoscope inside the machine. (with the optional module)
- Record of every cycle with endoscope serial number, operator and pass/fail result.
- Automatic reading of endoscope and operator ID (barcode or RFID) with electronic records; software link of each cycle to the patient where chosen. (with the optional module)
- Dry storage of reprocessed endoscopes in a filtered-air cabinet (EN 16442). (with the optional module)
- Cleaning and disinfection of TEE ultrasound probes. (with the optional module)
Not covered: Sterilisation of heat-stable surgical instruments needs a steam sterilizer; heat-sensitive critical devices that must be sterile need a low-temperature sterilizer.; Manual pre-cleaning and brushing straight after the procedure, as written in the endoscope maker's instructions, is still required; the machine does not replace it.; Microbiological culture testing of reprocessed endoscopes and rinse water needs a microbiology laboratory.; Repair of leaking or damaged endoscopes needs the endoscope maker's authorised repair service.
Loading the builder...
Standards this specification draws on
- ISO 15883-4:2018 Requirements and tests for washer-disinfectors that use chemical disinfection for heat-sensitive endoscopes: the device-defining standard (a new edition is at draft stage).
- ISO 15883-1:2024 General requirements, terms and tests for all washer-disinfectors, used together with Part 4 for validation and routine testing.
- ISO 15883-5:2021 Performance requirements and test method criteria for showing that a washer-disinfector really cleans (test soils).
- IEC 61010-1 and IEC 61010-2-040 Electrical safety of the machine; Part 2-040 covers sterilizers and washer-disinfectors used to treat medical materials.
- IEC 61326-1 Electromagnetic compatibility of electrical equipment for measurement, control and laboratory use.
- EN 16442:2015 Controlled-environment drying and storage cabinets for processed heat-sensitive endoscopes.
- ISO 17664-1:2021 Reprocessing information the endoscope manufacturer must give, including which automated washer-disinfector processes are validated.
- IEC 62304 Life-cycle process for medical device software, which controls the cycle, alarms and channel monitoring.
- ISO 14971:2019 Risk management of the reprocessor, including undetected channel blockage, chemical underdosing and software faults.
- ISO 13485:2016 Quality management system of the reprocessor manufacturer.
- DRAP Medical Devices Rules 2017 Legal requirement for registration (Class B-D) or enlistment (Class A) of medical devices imported or sold in Pakistan; the rules cover devices for disinfection of medical devices.
Common mistakes in tenders for this equipment
- Copying one brochure's numbers (an exact cycle time, chamber volume or number of connectors) or a trademarked process name, so only one maker can comply. Specify capacity, monitoring and results instead.
- Accepting a machine tied to one imported disinfectant without asking for cost per cycle and a fixed 5-year chemical price; the chemicals then cost more than the machine.
- Not listing the hospital's own endoscope makes and models, so the machine arrives without validated programs and connectors and some channels are never flushed.
- Treating the reprocessor as a replacement for manual brushing at the bedside and in the sink; the endoscope maker's manual cleaning steps are still required.
- Forgetting utilities: water pre-treatment, drains, compressed air for some models and, for some two-endoscope machines, a 400 V three-phase supply.
- No installation validation, no yearly revalidation and no testing of final rinse water, so a machine that recontaminates scopes is never detected.
- Ignoring software: no free updates and no safety notice history, although channel monitoring depends on software that has needed corrections.
- Accepting DRAP registration of the chemicals only, or of an older model; the exact offered reprocessor must be registered or enlisted.