Specification builder · Tier B

Bacterial endotoxin test system: incubating kinetic turbidimetric / chromogenic microplate or tube reader with software, gel-clot and recombinant options

An endotoxin test system measures bacterial endotoxin (the main cause of fever reactions to injections) in injectables, water for injection, dialysis water and medical-device extracts using the LAL (Limulus amebocyte lysate) test or newer recombinant reagents. An incubating plate or tube reader keeps samples at 37 deg C and software calculates results against a standard curve. It is bought by drug and device testing laboratories, sterile-products manufacturers and large dialysis units.

What this equipment will do

  • Samples: injectables and large-volume parenterals, water for injection, raw materials and components (USP <85>, Ph. Eur. 2.6.14, JP 4.01).
  • Samples: dialysis water and dialysis fluid, checked against the endotoxin limits of ISO 23500-3.
  • Samples: extracts of medical devices such as syringes, IV sets, cannulas, catheters and dialysers (USP <161>, ANSI/AAMI ST72).
  • Quantitative kinetic endotoxin test (kinetic turbidimetric or kinetic chromogenic) at 37 +/- 1 deg C, with standard curve, sample concentration, spike recovery and pass/fail (USP <85>, Ph. Eur. 2.6.14).
  • Both kinetic turbidimetric and kinetic chromogenic methods on the same reader, down to 0.01 EU/mL. (with the optional module)
  • Non-animal recombinant reagent test (recombinant factor C or recombinant cascade reagent) (Ph. Eur. 2.6.32, USP <86>). (with the optional module)
  • Gel-clot limit test in a 37 deg C heating block (USP <85>). (with the optional module)
  • Pharmacopoeial validity reports: standard-curve correlation, spike recovery 50-200%, maximum valid dilution, test for interfering factors and initial qualification. (with the optional module)
  • Check for false positives from glucans (samples from cellulose filters or cellulose dialysis membranes). (with the optional module)
  • Rapid single-sample cartridge test with result in 20 minutes or less. (with the optional module)
  • Automated dilutions, spiking, reagent addition and plate loading. (with the optional module)
  • Electronic records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)

Not covered: It does not detect non-endotoxin pyrogens; the rabbit pyrogen test or the monocyte activation test (Ph. Eur. 2.6.30) needs an animal facility or a cell-based MAT laboratory.; It does not test sterility (living bacteria and fungi); that needs the sterility testing pump or isolator and incubators (USP <71>).; It does not count microbes in water or non-sterile products; that needs a membrane filtration manifold and incubators (USP <61>, ISO 8199).; It does not measure particulate matter in injections; that needs a light-obscuration particle counter (USP <788>).

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • USP <85> Bacterial Endotoxins Test: defines the gel-clot, kinetic turbidimetric and kinetic chromogenic methods, 37 +/- 1 deg C incubation and validity criteria the reader and software must meet.
  • Ph. Eur. 2.6.14 European (and BP-harmonised) bacterial endotoxin test chapter used for most injectables registered in Pakistan.
  • JP 4.01 Japanese Pharmacopoeia endotoxin test, harmonised with USP <85> and Ph. Eur. 2.6.14.
  • Ph. Eur. 2.6.32 Test for bacterial endotoxins using recombinant factor C; needed only if a recombinant method is bought.
  • USP <86> USP chapter for bacterial endotoxin tests using recombinant reagents (rFC and recombinant cascade), official from May 2025.
  • USP <161> Medical devices: bacterial endotoxin and pyrogen tests, including extraction and device limits.
  • ANSI/AAMI ST72 Bacterial endotoxin test methods, routine monitoring and alternatives to batch testing for medical devices.
  • ISO 23500-3:2024 Water for haemodialysis and related therapies; sets the endotoxin limits dialysis-water testing must check.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures that the endotoxin software is checked against.
  • ISO/IEC 17025:2017 Accreditation standard for testing labs; requires calibrated equipment and controlled data.
  • IEC 61010-1 Electrical safety standard for laboratory equipment (reader, heating block, oven).
  • IEC 61326-1 EMC requirements for laboratory electrical equipment.

Common mistakes in tenders for this equipment

  • Copying one maker's cartridge, tube-reader or software description (fixed well numbers, run times or wavelengths) so that only one bidder can comply; state methods, temperature and validity calculations instead.
  • Buying the reader without first-year reagents, matched endotoxin standards, LAL reagent water and certified tips, so it stands idle while a second purchase is arranged.
  • Not fixing reagent prices and supply for several years: the reader is cheap compared with the reagents, and a closed system whose reagents later rise in price or go out of stock stops all testing.
  • Accepting a '21 CFR Part 11 compliant' claim without a live demonstration of audit trail, user roles, electronic signatures and protection against deletion of raw kinetic data.
  • Leaving out the depyrogenation oven, vortex and calibrated pipettes, then getting invalid runs from contaminated glassware and poor dilutions.
  • Not checking incubation temperature in edge wells at OQ; a plate reader that is not at 37 +/- 1 deg C everywhere gives failed spike recoveries.
  • Specifying recombinant factor C as the only method for a lab that must follow USP <85> or a BP monograph, before the method has been validated for its products.
  • Asking for DRAP medical-device registration for the reader: it is laboratory equipment, and the demand only confuses bidders.
Medicine brands A–Z:ABCDEFGHIJKLMNOPQRSTUVWXYZ0-9