Specification builder · Tier C
Benchtop or reach-in temperature and humidity test chamber for accelerated ageing of sterile barrier systems and medical devices and for conditioning of packages and specimens before testing (humidity control about 20-60 deg C, 30-90% RH; walk-in chambers excluded)
An environmental test chamber is a refrigerated, heated and humidified cabinet that holds samples at a set temperature and relative humidity. Medical-device manufacturers and testing laboratories use it to age sterile packs and devices at 50-60 deg C so that a shelf-life claim can be supported before real-time data exist (ASTM F1980, ISO 11607-1), and packaging and plastics labs use it to bring samples to standard atmospheres such as 23 deg C / 50% RH before seal, burst, leak or mechanical tests. In Pakistan, measured uniformity, purified water for the humidifier, temperature-humidity mapping and power backup decide whether weeks-long ageing studies can be defended.
What this equipment will do
- Sample types: sterile barrier systems (pouches, header bags, sealed trays), packaged single-use devices such as syringes, IV sets, catheters, gloves and sutures, packaging materials (films, foils, medical paper, nonwovens, cartons, labels) and plastic test specimens.
- Accelerated ageing of sterile packs and devices at a constant raised temperature, usually 50-60 deg C, with humidity controlled where the protocol requires, to support an expiry date until real-time data are available (ASTM F1980, ISO 11607-1 stability testing).
- Real-time ageing samples held at a controlled reference condition, for example 25 deg C, alongside the accelerated samples (ISO 11607-1).
- Conditioning of plastic specimens and packaging materials at the standard atmospheres 23 deg C / 50% RH or 27 deg C / 65% RH before testing (ISO 291).
- Conditioning of packages and packaging components at standard and special temperature-humidity atmospheres within the chamber's range, for example humid tropical conditions, before transit or performance tests (ASTM D4332, ISO 2233).
- Pre-conditioning of aged and unaged packs before seal-strength, burst and leak tests carried out on other instruments (ASTM F1980 test plan, ISO 11607-1).
- Frozen conditioning at about -18 deg C in the same chamber (ASTM D4332). (with the optional module)
- Extreme-cold conditioning at about -30 deg C (ASTM D4332). (with the optional module)
- Humid conditions above 60 deg C, up to 85 deg C / 85% RH, for research and materials testing. (with the optional module)
- Programmed temperature and humidity holds, ramps and timed runs, with an independent over-temperature cut-out that protects the samples.
- Even conditions on every shelf, with temperature and humidity deviation and fluctuation measured by the IEC 60068-3-5 / IEC 60068-3-6 method.
- Temperature and humidity record for every study, exported as CSV or PDF, with remote alarms. (with the optional module)
- Tamper-evident electronic ageing records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)
Not covered: Seal strength, burst and creep, dye-penetration and bubble leak tests of the aged packs (ASTM F88, ASTM F1140 / F2054, ASTM F1929, ASTM F2096) need a package seal-strength tester, a burst / leak tester or a dye-penetration kit.; ICH stability storage of medicines (for example 30 deg C / 65% RH and 40 deg C / 75% RH for registration dossiers) needs a qualified stability chamber, and light exposure to ICH Q1B needs a photostability chamber.; Transport simulation (drop, vibration and compression to ASTM D4169 or ISTA) needs a drop tester, vibration table and compression tester; whole pallets or large shipping cases need a walk-in chamber.; Sterility and bioburden testing of the aged devices needs a microbiology laboratory with incubators and a sterility-test isolator or clean room.
Loading the builder...
Standards this specification draws on
- IEC 61010-1:2010+AMD1:2016 Basic electrical, mechanical and fire safety for laboratory equipment; the safety baseline for any test chamber.
- IEC 61010-2-012:2019 Particular safety requirements for climatic and environmental testing and other temperature-conditioning equipment, including over-temperature and refrigerant hazards.
- IEC 61326-1:2020 Electromagnetic compatibility (EMC) of laboratory equipment, so the controller and alarms work reliably near other equipment.
- ASTM F1980-21 Standard guide for accelerated ageing of sterile barrier systems and medical devices: elevated-temperature ageing (Arrhenius / Q10 method) to support shelf-life claims until real-time data are available.
- ISO 11607-1:2019+A1:2023 Requirements for materials, sterile barrier systems and packaging systems for terminally sterilised medical devices, including stability testing (accelerated and real-time ageing) to show the pack stays sterile until its expiry date.
- ASTM D4332-22 Standard practice for conditioning containers, packages and packaging components at standard and special temperature-humidity atmospheres before testing.
- ISO 291:2008 Standard atmospheres for conditioning and testing plastics (23 deg C / 50% RH and 27 deg C / 65% RH) with class 1 and class 2 tolerances.
- ISO 2233:2000 Conditioning of complete, filled transport packages and unit loads for testing at defined temperature and humidity.
- IEC 60068-3-5:2018 Guidance on confirming the performance of temperature test chambers (how deviation and fluctuation are measured).
- IEC 60068-3-6:2018 Guidance on confirming the performance of temperature and humidity test chambers, including humidity deviation and fluctuation.
- ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories, including traceable calibration of conditioning equipment.
- 21 CFR Part 11 / EU GMP Annex 11 Requirements for trustworthy electronic records, audit trails and signatures for ageing and conditioning records.
Common mistakes in tenders for this equipment
- Buying an ICH drug stability chamber for package ageing (or the reverse) without checking the range: package ageing needs about 50-60 deg C, often above a stability chamber's qualified range, while conditioning needs 23 deg C / 50% RH held tightly.
- Copying one maker's exact figures (a fixed litre size, a -70 to +180 deg C range or a named software) so that only one brand qualifies; use 'not less than' sizes and ranges that match the real protocols.
- Accepting controller 'accuracy' or 'setting resolution' instead of measured temperature and humidity deviation and fluctuation at 23 deg C / 50% RH and the ageing temperature, with a test report.
- Asking for an over-wide range (-40 to +150 deg C, 10-98% RH) for ageing work that only needs 20-60 deg C, which doubles the price and narrows the field to a few makers.
- Forgetting humidifier water: no purified-water supply in the bid, so tap water is used and the humidifier and humidity sensor fail within months.
- Skipping temperature and humidity mapping, accredited calibration of the humidity sensor and an independent logger, so shelf-life data are challenged by auditors or notified bodies.
- Not budgeting for power protection: a chamber that restarts after every load-shedding cycle can run a 4-8 week ageing study out of tolerance without anyone noticing.
- Accepting 120 V / 60 Hz models without a factory-built 230 V / 50 Hz version, which then run on transformers and void the warranty.