Specification builder · Tier C
Automatic solvent (Soxhlet / hot extraction / continuous-flow) extraction system with solvent recovery, 4-6 or more positions, for crude and total fat/oil in foods, feeds, oilseeds, herbal and pharmaceutical raw materials and solvent extractables from medical-device materials (AOAC 920.39, ISO 6492, ISO 1443, ISO 659, EPA 3540C/3541, ISO 10993-12)
A fat extraction system dissolves the fat or oil out of a weighed, dried sample with a boiling organic solvent (petroleum ether, hexane, diethyl ether), recovers most of the solvent and leaves the extract in a cup that is dried and weighed. In Pakistan it is bought by food authority and PSQCA labs, feed mills and oilseed crushers, dairy and meat processors, pharma and herbal raw-material QC labs, device material testing labs, environmental labs and universities. The main choices are manual Soxhlet glassware against an automatic hot-extraction unit, the number of positions, the extraction modes it can run, how much solvent it recovers, whether an acid-hydrolysis unit is added for total fat, and the safety and data functions.
What this equipment will do
- Samples: animal feeds, feed ingredients, cereals, flours and oilseed meals for crude fat / ether extract (AOAC 920.39, ISO 6492).
- Samples: processed foods, snacks, bakery and confectionery products for crude fat before labelling and compliance checks (AOAC 920.39 type Soxhlet or hot extraction).
- Samples: oilseeds (cotton, sunflower, canola, sesame, soybean) and nuts for oil content (ISO 659).
- Samples: herbal, unani and botanical raw materials for crude fat / solvent-soluble extractive matter (WHO quality control methods for herbal materials).
- Samples: polymers, elastomers, packaging and other medical-device materials for exhaustive solvent extraction before extractables analysis (ISO 10993-12 sample preparation, ISO 10993-18 exhaustive extraction). (with the optional module)
- Samples: soils, sediments and sludges for solvent extraction of semi-volatile organics before GC or GC-MS (US EPA 3540C Soxhlet, EPA 3541 automated Soxhlet). (with the optional module)
- Gravimetric crude fat / oil determination with petroleum ether, hexane or diethyl ether and recovery of most of the solvent for reuse.
- Classical Soxhlet reflux (siphon) extraction as in the reference procedure. (with the optional module)
- Fast hot (submersion) extraction with programmed boiling, rinsing, recovery and drying steps, unattended. (with the optional module)
- Continuous-flow (Twisselmann type) extraction for heat-sensitive or slow-extracting matrices. (with the optional module)
- Total fat (including bound fat) in meat, dairy, bakery and pet-food products by acid hydrolysis before extraction (ISO 1443, AOAC 954.02 type). (with the optional module)
- Larger samples (10 g or more) for low-fat or inhomogeneous materials and for environmental extractions. (with the optional module)
- Automatic calculation and storage of the fat percentage from the entered weights, with export of result reports to a PC or LIMS. (with the optional module)
- Audit-trailed, electronically signed extraction records for regulated labs (21 CFR Part 11 type). (with the optional module)
Not covered: It does not weigh or dry anything: an analytical balance (0.1 mg) and a drying oven at 103 deg C plus a desiccator are needed to finish every determination.; It does not give the fatty-acid profile, trans fat or oil quality (free fatty acid, peroxide value): those need a gas chromatograph (FID) for fatty acid methyl esters and titration glassware.; Rapid fat in milk and dairy liquids is done by Gerber centrifuge or mid-infrared milk analysers; fat in meat by NMR or NIR analysers needs the extraction system only as the reference method.; Identification and quantification of extractables and leachables from device materials needs GC-MS and LC-MS after the extraction; this system only prepares the extract.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical safety requirements for laboratory equipment.
- IEC 61010-2-010 Particular safety requirements for laboratory equipment that heats materials, covering over-temperature and hot-surface hazards of a solvent extractor.
- IEC 61326-1 Electromagnetic compatibility of laboratory equipment with electronic temperature and sequence control.
- AOAC 920.39 / ISO 6492 Reference methods for crude fat (ether extract) in animal feeds by Soxhlet or equivalent solvent extraction; the methods the instrument must be able to perform.
- ISO 1443 / AOAC 954.02 Total fat in meat and pet food by acid hydrolysis followed by solvent extraction, which needs the hydrolysis unit in front of the extractor.
- ISO 659 Oil content of oilseeds by hexane or petroleum ether extraction, used by oilseed crushers and grain exporters.
- AOCS Am 5-04 Rapid crude fat determination by high-temperature solvent extraction in filter bags, the official basis for closed filter-bag extractors.
- US EPA SW-846 Method 3540C / 3541 Soxhlet and automated Soxhlet extraction of semi-volatile organics from soils and sludges in environmental labs.
- ISO 10993-12 / ISO 10993-18 Sample preparation and chemical characterisation of medical-device materials, including exhaustive solvent extraction of polymers before extractables analysis.
- WHO Quality control methods for herbal materials (2011) Determination of extractable matter in medicinal plant materials by solvent extraction, used by herbal and unani manufacturers and drug testing labs.
- ISO/IEC 17025 Competence of testing labs; requires traceable calibration of the extractor's temperature and timer and documented verification with reference materials.
Common mistakes in tenders for this equipment
- Copying one maker's figures (exact solvent recovery percentage, IP ratings, ATEX valves, pressurised electronics, named sensor or dispensing features) so that only one model complies; state the positions, temperature, recovery, sample mass and safety functions needed and accept equivalent designs.
- Using trade names for hot extraction, thimble filters, hydrolysis caps or controllers instead of describing the measurable function (Randall / Soxhlet / continuous modes, program steps, recovery percentage).
- Asking for medical electrical safety standards (IEC 60601) for a laboratory extractor; the correct standards are IEC 61010-1 and IEC 61010-2-010.
- Forgetting that every result also needs an analytical balance, a drying oven and a desiccator; a lab with an extractor but no 0.1 mg balance cannot report fat.
- Leaving out thimbles, seals, spare cups, solvent, a reference material and cooling water (chiller) so the unit arrives but cannot run or cannot be verified.
- Specifying total fat (ISO 1443) without the hydrolysis unit, or specifying a hydrolysis unit whose filters do not fit the offered extractor, so samples must be transferred by hand with loss.
- Ignoring site needs: a fume hood or exhaust, no open flames, a solvent store, a stable earthed single-phase supply and cooling water.
- Not asking for a fixed post-warranty AMC price, traceable annual temperature calibration and a 7-year guarantee of glassware and seal availability.