Handheld (pocket) ultrasonic fetal Doppler with 2-3 MHz obstetric probe and fetal heart rate display for antenatal clinics, outreach and intermittent auscultation in labour

A handheld fetal Doppler is a small battery-powered ultrasound device that lets health workers hear the unborn baby's heartbeat and read the heart rate on a display. BHUs, RHCs, lady health worker programmes, hospital antenatal clinics, labour rooms and private clinics buy them in large numbers. In Pakistan the key tender decisions are the probe frequency (2 MHz for labour and later pregnancy, 3 MHz for early pregnancy), battery type for outreach, and avoiding cheap home-use listeners that are not tested for clinical accuracy.

What this equipment will do

  • Patients: pregnant women from about 12 weeks with a 3 MHz probe, or from the second trimester and in labour with a 2 MHz probe (IEC 61266).
  • Fetal heart sounds on a loudspeaker, so the user and the mother can hear the heartbeat (IEC 60601-2-37 acoustic output limits).
  • Fetal heart rate number on a display, 50 to 210 bpm or wider, accuracy +/-2 bpm.
  • Intermittent auscultation in labour: counting the fetal heart rate during and after contractions at set intervals.
  • Both early and late pregnancy with one unit by changing between 2 MHz and 3 MHz probes. (with the optional module)
  • Short FHR trend on a colour screen to see slowing of the heart rate through a contraction. (with the optional module)
  • Use in wet labour-room conditions and cleaning of the probe by immersion. (with the optional module)
  • Listening with headphones or recording the heart sounds. (with the optional module)
  • Simple peripheral arterial and venous blood-flow checks with a separate vascular probe. (with the optional module)

Not covered: Continuous recording of the fetal heart rate with contractions on paper (non-stress test, high-risk labour) needs a cardiotocograph (CTG).; Dating the pregnancy, checking fetal growth, position, placenta and amniotic fluid needs an obstetric ultrasound scanner.; Umbilical and fetal blood-flow (pulsed and colour Doppler) studies need an ultrasound scanner with Doppler; a handheld fetal Doppler only gives sound and heart rate.; Mothers' blood pressure and pulse oximetry need a BP apparatus and pulse oximeter; the Doppler does not measure maternal vital signs.

6 international manufacturers checkedMarket checked 2026-109 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment.
  • IEC 60601-2-37 Particular safety requirements for ultrasonic diagnostic and monitoring equipment (edition 3, 2024), including acoustic output limits of the Doppler probe.
  • IEC 61266 Performance requirements and methods of measurement and reporting for hand-held probe Doppler foetal heartbeat detectors (1994, still valid).
  • IEC 61157 Standard means for the reporting of acoustic output of medical diagnostic ultrasonic equipment.
  • IEC 60601-1-2 Electromagnetic compatibility, so the unit is not disturbed by phones and other equipment.
  • IEC 60601-1-6 Usability of medical electrical equipment, important for use by outreach workers.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Buying consumer 'home baby heartbeat' listeners with no heart-rate display or clinical accuracy data because they are cheapest; specify IEC 61266 performance data and professional-use approval.
  • Copying one brand's brochure: exact screen colours, a named noise-reduction feature or a particular probe shape. Specify probe frequency, FHR range and accuracy and battery life instead.
  • Choosing only a 3 MHz probe for labour rooms or heavy women, where the 2 MHz probe is needed for depth, or only 2 MHz for clinics seeing mainly early pregnancies.
  • Rechargeable-only units for outreach workers without reliable charging; the units stand idle during load-shedding.
  • No gel, spare batteries or carry case in the bid, so units are useless on outreach visits.
  • No training on telling the mother's pulse from the baby's heart rate, leading to false reassurance when the baby is in distress or has died.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.