Specification builder · Tier A
Flame (emission) photometer for sodium, potassium, lithium, calcium and barium with air compressor, calibration standards and optional dilutor, autosampler and PC software
A flame photometer sprays a liquid sample into a low-temperature gas flame and measures the coloured light that sodium, potassium, lithium, calcium (and barium) atoms emit, which gives their concentration directly. In Pakistan it is bought by pharma QC labs and drug testing labs for oral rehydration salts, infusion fluids and dialysis solutions, by small hospital and clinical labs for serum and urine electrolytes, and by water, soil and fertiliser labs. The key choices are single-channel versus simultaneous multi-channel measurement, the detection limits and precision proven at installation, the fuel gas, and whether a dilutor, autosampler and PC software with data-integrity controls are needed.
What this equipment will do
- Products: oral rehydration salts (ORS) powders and solutions, sodium chloride / glucose / compound sodium lactate / potassium chloride infusion fluids, oral electrolyte solutions, haemodialysis and peritoneal dialysis concentrates (BP/Ph. Eur. Appendix II D / 2.2.22 atomic emission; USP ORS monograph flame photometer assay).
- Samples: serum, plasma and urine after dilution with the lithium or caesium diluent; drinking water, waste water and irrigation water (ISO 9964-3, Standard Methods 3500-Na B / 3500-K B); soil extracts, fertilisers, plant digests and process liquids.
- Sodium (Na) assay by flame emission at about 589 nm, result in ppm, mg/L, mmol/L or mEq/L.
- Potassium (K) assay by flame emission at about 766 nm, result in ppm, mg/L, mmol/L or mEq/L.
- Lithium (Li) assay at about 671 nm, including lithium carbonate products and serum lithium for therapy monitoring (after dilution). (with the optional module)
- Calcium (Ca) assay at about 622 nm in solutions free of interfering phosphate, at the higher detection limit typical of a low-temperature flame. (with the optional module)
- Barium (Ba) assay at about 515 nm on a propane/butane flame. (with the optional module)
- Simultaneous reading of two or more elements from a single aspiration of the sample. (with the optional module)
- Automatic in-line dilution of serum, urine or concentrated samples before the flame. (with the optional module)
- Unattended batch analysis from a sample tray. (with the optional module)
- Multi-point calibration curves, stored calibrations and result storage with printout or USB export. (with the optional module)
- Electronic records with user log-ins, audit trail and tamper-evident reports for GMP data integrity. (with the optional module)
Not covered: Magnesium, zinc, iron, copper, lead and other heavy metals: a flame photometer cannot measure them; use the atomic absorption spectrophotometer or ICP-OES.; Chloride, bicarbonate, ionised calcium, pH and blood gases in patient samples: use the ion-selective electrolyte analyser or blood gas analyser.; Elemental impurities screening per ICH Q3D / USP <232>/<233>: needs ICP-MS or ICP-OES.; Trace calcium and magnesium (water hardness below about 2 ppm) and calcium in phosphate-rich samples: use atomic absorption or complexometric titration.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical safety of laboratory equipment, covering the instrument and its air compressor next to an open gas flame.
- IEC 61326-1 Electromagnetic compatibility of laboratory measurement equipment.
- Ph. Eur. 2.2.22 Atomic emission spectrometry chapter under which flame photometric assays in Ph. Eur. and BP monographs are performed.
- BP Appendix II D British Pharmacopoeia atomic spectrophotometry (emission and absorption) appendix cited by infusion, ORS and electrolyte monographs for sodium and potassium assays.
- USP Oral Rehydration Salts monograph (flame photometer assay) USP monograph that assays sodium and potassium with a flame photometer zeroed on a lithium diluent at about 589 nm and 766 nm.
- USP <1058> Framework for analytical instrument qualification (DQ, IQ, OQ, PQ).
- 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and e-signature requirements where PC software stores release data.
- ISO 9964-3 Water quality: determination of sodium and potassium by flame emission spectrometry.
- Standard Methods 3500-Na B / 3500-K B APHA/AWWA/WEF flame emission photometric methods for sodium and potassium in water and waste water.
- ISO 17034 / ISO/IEC 17025 Accreditation of reference material producers and calibration laboratories supplying the certified standards and calibration certificates.
Common mistakes in tenders for this equipment
- Copying one brand's detection limits or precision (for example 0.02 ppm potassium or a 15-second stabilisation time) so only that model complies; specify the pharmacopoeial-grade figures instead and verify them at installation.
- Demanding a hardware internal-standard channel, Part 11 software or an Ethernet port by name when the lab only runs ORS and infusion assays; each of these narrows the field to one or two makers.
- Leaving out the air compressor, gas regulator, hoses and leak test, so the instrument sits unconnected for months after delivery.
- Forgetting certified calibration standards, diluent and spare nebulisers; the first blocked nebuliser or expired standard stops all testing while imports are awaited.
- Asking for calcium at the same detection limit as sodium; a low-temperature flame measures calcium only at tens of ppm, and phosphate-rich samples need pretreatment or atomic absorption.
- Specifying natural gas operation in a lab with unreliable piped gas pressure, or specifying LPG without a safe cylinder location and ventilation above the chimney.
- Buying a flame photometer for routine patient electrolytes in a busy hospital where an ion-selective electrolyte analyser (chloride, ionised calcium, no dilution, no gas) is the safer and faster choice.
- Ignoring post-warranty service: a cheap instrument with no local nebuliser and compressor spares costs more over 5 years than a slightly dearer one with a local engineer.