Specification builder · Tier C

Flexible video bronchoscopy system (video processor with LED light source or portable display unit, reusable or single-use bronchoscopes of standard, therapeutic and slim sizes, optional linear/radial EBUS) with scope reprocessing equipment

A flexible video bronchoscope lets a pulmonologist, anaesthetist or intensivist look inside the airways, suck out secretions or blood, and take washings and biopsies for TB, pneumonia and lung cancer. Chest units, ICUs and operating theatres in DHQ, teaching, children's and private hospitals buy it either as a reusable system (video processor, LED light, scopes of different sizes, leak tester and washer-disinfector) or as a portable single-use system. The key decisions are reusable versus single-use, which scope sizes and channel widths, whether EBUS is really needed, and how scopes will be safely reprocessed; reusable scopes normally work only with the same maker's processor.

What this equipment will do

  • Patients: adults (and older children) through the mouth, nose, breathing tube or tracheostomy, with a standard scope of not more than 5.5 mm outer diameter and a 2.0 mm or larger working channel (IEC 60601-2-18).
  • Patients: small children, narrow airways, stenoses and peripheral lesions, with a slim scope of not more than 4.2 mm outer diameter. (with the optional module)
  • Airway inspection with live video: diagnostic bronchoscopy, checking breathing-tube position, help with difficult intubation and guidance of percutaneous tracheostomy, with still images and video recorded from the scope buttons (IEC 60601-2-18, ISO 8600).
  • Suction of secretions and mucus plugs, and bronchial washing and bronchoalveolar lavage (BAL) samples for TB and pneumonia testing.
  • Suction of blood, clots and thick secretions, and use of larger forceps, baskets and probes, through a therapeutic scope with a 2.8 mm or larger channel. (with the optional module)
  • Tissue and cell sampling and foreign-body removal: biopsy forceps, cytology brushes, transbronchial aspiration needles and grasping forceps passed through the working channel.
  • Linear EBUS: ultrasound view and needle sampling of lymph nodes in the chest for lung cancer staging and TB (IEC 60601-2-37). (with the optional module)
  • Radial EBUS: finding and sampling small lung nodules in the outer parts of the lung with a miniature ultrasound probe. (with the optional module)
  • Treatment inside the airway: cutting, coagulation and argon plasma coagulation to stop bleeding and remove tumour or tissue (IEC 60601-2-2). (with the optional module)
  • Image-enhancement mode that improves the contrast of the airway lining and blood vessels to help spot abnormal areas. (with the optional module)
  • Automated cleaning and high-level disinfection of reusable scopes between patients, with leak test and a printed or electronic record of every cycle (ISO 15883-4, ISO 17664-1). (with the optional module)
  • Patient worklist from the hospital system and storage of images in PACS (DICOM), with a structured bronchoscopy report. (with the optional module)

Not covered: Rigid bronchoscopy for large foreign bodies, massive bleeding and airway stents needs a rigid bronchoscope set; this flexible system does not replace it.; Gastroscopy, colonoscopy and other gut examinations need a GI video endoscopy system; bronchoscopes are not used in the gut.; Cryobiopsy and freezing treatment of airway tumours need a cryotherapy unit with cryoprobes, which is not part of this system.; Testing the samples taken (AFB smear, GeneXpert, culture, cytology and histopathology) needs the microbiology and pathology laboratory.

8 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including systems on one trolley.
  • IEC 60601-1-2 Electromagnetic compatibility, so the video image is not disturbed by diathermy, argon plasma and ICU equipment.
  • IEC 60601-2-18 Particular safety requirements for endoscopic equipment: light and tip temperature, insulation of the insertion tube and HF compatibility of bronchoscopes.
  • IEC 60601-2-37 Particular safety requirements for ultrasound equipment, applying to the EBUS scope transducer and ultrasound processor.
  • ISO 8600 series Optics and photonics standards for medical endoscopes: field of view, direction of view, insertion portion width and channel width, measured the same way by every maker.
  • ISO 15883-4 Requirements and tests for washer-disinfectors with chemical disinfection for thermolabile endoscopes (read with ISO 15883-1 general requirements).
  • ISO 17664-1 Manufacturer's information for processing reusable medical devices, so the reprocessing method for each bronchoscope is validated and written down.
  • ISO 14971 Risk management for medical devices, covering hazards such as cross-infection, image loss, tip heating and fluid invasion.
  • ISO 13485 Quality management system of the device manufacturer.
  • IEC 62304 Software life-cycle process for the processor, display unit and ultrasound software.
  • DRAP Medical Devices Rules 2017 Legal requirement for registration or enlistment of medical devices imported or sold in Pakistan.

Common mistakes in tenders for this equipment

  • Writing the specification around one maker's processor and then buying scopes later in open tender: scopes from another maker will not run on it. Specify processor, scopes and any EBUS scope together, and require a written compatibility statement and fixed 5-year scope prices.
  • Copying one brand's brochure: trademarked image-enhancement mode names, an exact outer diameter to 0.1 mm or a unique bending angle that only one scope quotes. Specify ranges such as 'outer diameter not more than 6.3 mm, channel not less than 2.8 mm'.
  • Buying only one reusable scope, or only a 2.0 mm channel scope, so one repair or one case of massive haemoptysis stops the service.
  • Leaving out reprocessing: no leak tester, no washer-disinfector to ISO 15883-4, no drying cabinet and no staff training, so scopes are damaged by fluid invasion within months and patients are put at risk of TB and resistant-bacteria cross-infection.
  • Comparing a single-use system with a reusable one on purchase price alone. At 300 procedures a year single-use scopes can cost more each year than the whole reusable system; at low volume or in ICU they can be cheaper and safer. Evaluate 5-year cost per procedure.
  • Asking for EBUS in a hospital without trained operators, cytopathology support or budget for needles and scope repairs, so an expensive brand-locked scope sits unused.
  • Warranty and CMC that exclude reusable scopes or treat every scope fault as misuse, and no fixed price list for minor, medium and major scope repairs.
  • Accepting DRAP registration of the processor only; every scope, single-use scope, ultrasound unit and washer-disinfector must be registered or enlisted.