Flexible video gastroscopes and colonoscopes (standard, large-channel, slim and paediatric) with leak testers and reprocessing accessories, for a same-maker video processor
Flexible video gastroscopes and colonoscopes are the reusable camera scopes used for upper GI endoscopy and colonoscopy: diagnosis, biopsy, variceal banding, bleeding control and polyp removal. DHQ, teaching, children's and private hospitals buy them with a new video processor or to add to an existing one. The key decisions are which scope sizes and channels, how many scopes, how they will be leak-tested and disinfected, and the fact that scopes normally only work on the same maker's processor.
What this equipment will do
- Patients: adults and older children for upper GI endoscopy of the oesophagus, stomach and duodenum, with a standard gastroscope of not more than 9.9 mm insertion tube and a 2.8 mm or larger channel (IEC 60601-2-18).
- Patients: adults for colonoscopy, with an adult colonoscope of not more than 13.2 mm insertion tube, a 3.7 mm or larger channel and a working length of 1650 mm or more. (with the optional module)
- Patients: children, narrow or angulated bowels and strictures, with a paediatric / slim colonoscope of not more than 11.8 mm insertion tube. (with the optional module)
- Patients: infants, tight strictures and transnasal examination, with a slim gastroscope of not more than 6.0 mm outer diameter. (with the optional module)
- Live video inspection of the gut lining on the matching video processor, with field of view and diameters declared and measured the same way by every maker (ISO 8600-1, ISO 8600-3, ISO 8600-4).
- Tissue sampling: biopsy forceps passed through the working channel for histopathology, H. pylori and cancer diagnosis.
- Treatment through the scope: injection, haemostatic clips, polyp removal with snares and foreign-body retrieval. (with the optional module)
- Band ligation of bleeding oesophageal varices with multi-band ligators fitted to the gastroscope. (with the optional module)
- Fast suction of blood and use of larger accessories through a gastroscope with a 3.7 mm or larger channel. (with the optional module)
- Image-enhancement mode that improves the contrast of the mucosa and blood vessels to help spot early cancers and flat polyps. (with the optional module)
- Close-up view of the mucosa from 2 mm or less with a near-focus or magnification mode. (with the optional module)
- Forward water jet to wash blood and stool off the bowel wall during the procedure. (with the optional module)
- Leak testing of every scope before cleaning, and manual cleaning and high-level disinfection as per the manufacturer's validated instructions (ISO 17664-1).
- Single-use gastroscope for emergencies when reusable scopes are being reprocessed. (with the optional module)
Not covered: ERCP (bile duct and pancreas procedures) needs a side-viewing duodenoscope; forward-viewing gastroscopes cannot do it.; Endoscopic ultrasound (EUS) needs an ultrasound endoscope and ultrasound processor, which are not part of this lot.; Automated cleaning and disinfection needs an automated endoscope reprocessor (washer-disinfector to ISO 15883-4), specified separately; this lot supplies only leak testers and cleaning accessories.; The video processor, light source, monitor, insufflator and recorder are specified under the GI video endoscopy tower; small-bowel examination needs capsule endoscopy or an enteroscope.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including the processor and light source the scopes connect to.
- IEC 60601-1-2 Electromagnetic compatibility of the processor and light source, so the image is not disturbed by diathermy and other theatre equipment.
- IEC 60601-2-18 Particular safety and essential performance requirements for endoscopic equipment, covering the scopes and the equipment that powers them.
- ISO 8600-1 General requirements and definitions for medical endoscopes and endotherapy devices.
- ISO 8600-3 How field of view and direction of view are measured, so every maker's declared field of view is comparable.
- ISO 8600-4 How the maximum width of the insertion portion is measured, so declared scope diameters are comparable.
- ISO 17664-1 Information the manufacturer must give for reprocessing critical and semi-critical devices, so each scope has validated cleaning and disinfection instructions.
- ISO 15883-4 Requirements and tests for washer-disinfectors with chemical disinfection for thermolabile endoscopes, against which scope-compatible machine programs are named.
- ISO 14971 Risk management for medical devices, covering hazards such as cross-infection, fluid invasion and image loss.
- ISO 13485 Quality management system of the scope manufacturer.
- DRAP Medical Devices Rules 2017 Legal requirement for registration or enlistment of medical devices imported or sold in Pakistan.
Common mistakes in tenders for this equipment
- Buying scopes in an open tender without naming the processor: scopes of another maker, or another generation, will not run on it. Specify processor and scopes together, or state the existing processor and require a written compatibility statement.
- Copying one brand's data sheet: an exact diameter to 0.1 mm, a trade-named image mode or a unique angulation figure that only one scope quotes. Use 'not more than' and 'not less than' ranges such as insertion tube not more than 9.9 mm and channel not less than 2.8 mm.
- Buying a single gastroscope and a single colonoscope for a busy unit, so every reprocessing cycle or repair stops the list.
- Leaving out leak testers, brushes, adaptors and reprocessing training, so scopes suffer fluid invasion within months and patients are exposed to cross-infection.
- Warranty and maintenance contracts that exclude scopes or treat every scope fault as misuse, with no fixed price list for minor, medium and major scope repairs.
- Accepting dimensions and field of view from a sales leaflet only; require the original data sheet page for each clause and check diameters and angulation at acceptance.
- Accepting DRAP registration of the processor only; every scope model must be registered or enlisted.