GI video endoscopy tower (full HD or 4K video processor with LED light source and image enhancement, medical monitor, air or CO2 insufflation, water-jet pump, image capture and reporting, trolley, with matching gastroscopes and colonoscopes)
A GI video endoscopy tower is the trolley of equipment that runs flexible gastroscopes and colonoscopes: video processor and light source, monitor, air or CO2 insufflation, water-jet pump, recorder and reporting. Gastroenterology and surgical units in DHQ, teaching, cancer and private hospitals buy it for upper GI endoscopy, colonoscopy, biopsy, banding, polypectomy and bleeding control. The key decisions are full HD versus 4K, which enhancement modes, CO2 and water jet, and buying the scopes together with the processor, because scopes work only with the same maker's processor.
What this equipment will do
- Patients: adults (and larger children where the hospital's scope specification allows) having upper GI endoscopy through the mouth or colonoscopy, with flexible video scopes of the same maker as the processor (IEC 60601-2-18).
- Live full HD video of the oesophagus, stomach, duodenum and colon, with freeze, still images and electronic zoom (IEC 60601-1, IEC 60601-2-18).
- 4K UHD picture for the finest mucosal detail, on a 4K medical monitor. (with the optional module)
- Digital enhancement of surface structure and colour on the live image to help see subtle lesions.
- Narrowed-spectrum light mode that makes mucosal patterns and small vessels stand out, for early cancer, Barrett's oesophagus and flat polyp detection. (with the optional module)
- Inflation of the stomach and bowel with room air from the built-in air pump, and lens washing. (with the optional module)
- Inflation with CO2 from a cylinder or pipeline for colonoscopy and long therapeutic procedures. (with the optional module)
- Foot-switch water jet to wash blood, stool and mucus off the lining during bleeding cases and poorly prepared colonoscopy. (with the optional module)
- Biopsy, polypectomy, injection, clipping, banding and foreign-body removal with instruments passed through the scope channel (instruments in the offered set or bought separately).
- Cutting and coagulation for polypectomy and sphincterotomy, and argon plasma coagulation for bleeding and ablation (IEC 60601-2-2). (with the optional module)
- Computer-aided marking of suspected polyps on the live colonoscopy image. (with the optional module)
- Patient data, image and video recording, USB export and a printed procedure report.
- Patient worklist from the hospital system and image storage in PACS (DICOM), with structured endoscopy reports. (with the optional module)
Not covered: ERCP (bile duct and pancreas work) needs a side-viewing duodenoscope and X-ray (C-arm) screening; the tower may drive a duodenoscope of the same maker, but the duodenoscope is bought separately.; Endoscopic ultrasound (EUS) needs a dedicated ultrasound endoscope and ultrasound processor, which are not part of this tower.; Small-bowel examination of the whole small intestine needs capsule endoscopy or a balloon-assisted enteroscopy system.; Cleaning and high-level disinfection of the scopes between patients needs an endoscope washer-disinfector (automated endoscope reprocessor) and a drying cabinet; the tower only includes a leak tester.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including the medical electrical system formed by all devices on one trolley.
- IEC 60601-1-2 Electromagnetic compatibility, so the endoscopic image is not disturbed by electrosurgery, argon plasma and other equipment in the room.
- IEC 60601-2-18 Particular basic safety and essential performance requirements for endoscopic equipment, covering the processor, light source and scopes used inside the patient.
- ISO 8600-1 General requirements and definitions for medical endoscopes (current 2025 edition), so scope properties such as field of view and insertion width are stated the same way by every maker.
- IEC 60601-2-2 Particular safety requirements for high-frequency surgical equipment, applying to the electrosurgical and argon plasma unit used through the scopes.
- ISO 14971 Risk management for medical devices, covering hazards such as image loss, over-insufflation, cross-infection and burns from electrosurgery.
- ISO 13485 Certified quality management system of the device manufacturers.
- IEC 62304 Software life-cycle process for processor, recorder and detection software.
- IEC 62353 Electrical safety testing at installation, after repair and at preventive maintenance, including the complete trolley system.
- DICOM (NEMA PS3) Standard for receiving patient worklists and sending endoscopy images to PACS.
- DRAP Medical Devices Rules 2017 Pakistan's legal requirement (S.R.O. 32(I)/2018) for enlistment or registration of medical devices before import and sale.
Common mistakes in tenders for this equipment
- Writing the specification around one maker's processor and buying scopes later in open tender: scopes from another maker will not run on it. Specify processor and scopes together and require a written compatibility statement and fixed 5-year scope prices.
- Copying one brand's brochure: trademarked image-enhancement mode names, a unique number of LEDs or an exact output connector list that only one processor has. Ask for the function (for example 'narrowed-spectrum light mode') and a minimum resolution instead.
- Asking for 4K without a 4K monitor, or accepting a bid where 4K output, DICOM or detection software needs a paid software key that is not in the price.
- Leaving out the CO2 regulator, cylinders and tubing, the flushing-pump tubing, valves and water bottles, so the tower cannot be used on delivery.
- Buying only one gastroscope and one colonoscope for a busy unit, so one scope repair or one long cleaning cycle stops the list.
- Warranty and CMC that cover the processor but exclude the scopes, or treat every scope fault as misuse, with no fixed price list for scope repairs.
- Accepting DRAP registration of the processor only; every scope, CO2 unit, pump and any electrosurgical unit must be registered or enlisted.
- Buying the tower without a plan for scope reprocessing (leak tester, washer-disinfector, drying cabinet, trained staff), which leads to scope damage and cross-infection.