Specification builder · Tier B
Advanced haemodynamic (cardiac output) monitoring system for ICU and cardiac OT: calibrated and uncalibrated arterial pulse contour, transpulmonary thermodilution, non-invasive methods and ScvO2, as stand-alone monitor or plug-in module
A monitor or plug-in module that measures how much blood the heart pumps (cardiac output) and whether a patient will respond to fluids, using an arterial line, a thermodilution catheter or non-invasive sensors. Tertiary ICUs, cardiac surgery centres and high-risk OTs buy it to guide fluids, inotropes and vasopressors in shock and major surgery. The key tender decisions are the measurement method and controlling the long-term cost of single-use sensors that usually come from only one maker.
What this equipment will do
- Patients: adults in septic shock or ARDS, after cardiac surgery and during high-risk major surgery (goal-directed fluid therapy).
- Patients: children, with paediatric catheters or sensors and the lowest validated body weight stated (not more than 15 kg). (with the optional module)
- Continuous cardiac output, stroke volume and SVR from an ordinary radial or femoral arterial line without calibration, with stroke volume and pulse pressure variation for ventilated patients (IEC 60601-2-34). (with the optional module)
- Transpulmonary thermodilution (cold bolus) to calibrate continuous cardiac output, plus preload volume (GEDI), lung water (ELWI) and heart pump function (GEF, CFI). (with the optional module)
- Non-invasive cardiac output with no catheter, by finger cuff, bioimpedance or pulse wave transit time (ISO 81060-3 for finger-cuff methods). (with the optional module)
- Pulmonary artery catheter cardiac output, bolus or continuous, with mixed venous oxygen saturation (SvO2) and right heart values. (with the optional module)
- Oxygen delivery and use: continuous central venous oxygen saturation (ScvO2) by fibre-optic catheter, and DO2 / VO2 calculated from blood gas results. (with the optional module)
- Regional / cerebral oxygen saturation by near-infrared sensors on 2 or more channels. (with the optional module)
- Early warning of hypotension (MAP below 65 mmHg) minutes before it happens, from the arterial waveform. (with the optional module)
- 72-hour trends, user target ranges and alarms with adjustable limits for every parameter (IEC 60601-1-8, IEC 80601-2-49).
- Guided fluid challenge and passive leg-raise test that reports the change in stroke volume against a set threshold. (with the optional module)
- Values shown on the bedside monitor and central station, and sent to the hospital EMR in HL7. (with the optional module)
Not covered: It does not show heart valves, wall movement or pericardial fluid; that needs echocardiography or a point-of-care ultrasound machine.; Blood gases and lactate need a blood gas analyser; the monitor only uses results typed or sent into it.; A stand-alone unit is not a full patient monitor: routine ECG, SpO2, NIBP and respiratory monitoring need a multiparameter patient monitor.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical equipment connected to catheters near the heart.
- IEC 60601-1-2 Electromagnetic compatibility near diathermy, defibrillators and phones.
- IEC 60601-1-8 Alarm system design: priorities, sounds and visual signals.
- IEC 80601-2-49 Particular requirements for multifunction patient monitoring equipment, covering both stand-alone haemodynamic monitors and modules in host monitors.
- IEC 60601-2-34 Invasive blood pressure monitoring, on which pulse contour and thermodilution methods depend.
- ISO 81060-3 Clinical validation of continuous non-invasive blood pressure (finger-cuff) methods, where offered.
- IEC 62304 Software life-cycle for the cardiac output and prediction algorithms.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer.
- IEC 62353 Electrical safety testing at installation and after repair.
- DRAP Medical Devices Rules 2017 Registration/enlistment needed to import and sell the monitor and its disposables in Pakistan.
Common mistakes in tenders for this equipment
- Writing the tender around one maker's trademarked method or parameter names, so only that maker qualifies. Specify the measurement principle, displayed parameters and validation evidence, and accept any validated technology.
- Accepting a 'free' or low-priced monitor in return for buying the maker's single-use sensors: the disposables then cost several times the monitor over 5 years and the hospital is locked to one maker. Evaluate monitor plus 5-year disposable cost together.
- Not fixing disposable prices or local stock: proprietary sensors and catheter kits are single-source, so price rises or import delays leave the monitor idle for months.
- Buying a plug-in module without naming the host monitor make and software version; the module does not fit or needs a paid host upgrade.
- Choosing transpulmonary thermodilution for every bed when most patients only need uncalibrated or non-invasive monitoring; it needs femoral or axillary catheters and trained staff and costs much more per patient.
- Leaving out clinical application training and bedside support: the monitor is installed but rarely used because staff cannot set up lines, calibrate or interpret the numbers.
- Accepting DRAP registration of the monitor only; the single-use sensors and catheters must also be registered or enlisted, or they cannot be legally imported.
- Accepting accuracy claims without published validation against thermodilution in the intended patient group (adult shock, cardiac surgery, paediatric).