Heart-lung machine for cardiopulmonary bypass: modular roller and centrifugal pump console with safety monitoring, gas blender, cardioplegia delivery, battery back-up and heater-cooler, for adult and paediatric open-heart surgery

A heart-lung machine takes over the pumping of the heart and, through a single-use oxygenator, the work of the lungs while the surgeon operates on a stopped heart. It is bought by cardiac surgery centres in public tertiary, teaching, paediatric and private hospitals for coronary bypass, valve, aortic and congenital heart surgery. The key tender decisions are the number and type of pumps, the safety detectors, the heater-cooler and its infection-control routine, battery back-up, and the 5-year cost of oxygenators, tubing packs and pump heads.

What this equipment will do

  • Patients: adults and children having open-heart surgery on cardiopulmonary bypass, such as coronary bypass, valve, aortic and congenital heart operations (IEC 60601-1).
  • Blood pumping: 4 or more independently controlled pumps for arterial flow, suction, venting and cardioplegia, with speed and flow shown for each pump.
  • Six or more pumps on one console for complex aortic and congenital cases. (with the optional module)
  • Arterial flow from a centrifugal pump with measured flow on the arterial line, with a roller pump kept as back-up. (with the optional module)
  • Safety monitoring: reservoir level and arterial air-bubble detection with alarms, at least 2 circuit pressures and 2 temperatures, with stated automatic pump-stop actions.
  • Extra safety: two-point (warning and stop) level detection, 3 or more pressure and temperature channels and a second bubble detector. (with the optional module)
  • Gas supply to the oxygenator: air/oxygen blending from 21% to 100% oxygen with sweep flow up to at least 10 L/min.
  • Very low gas flows for neonates and infants. (with the optional module)
  • Carbon dioxide blending into the sweep gas for pH-stat management. (with the optional module)
  • Cardioplegia delivery by a dedicated pump with line pressure monitoring and alarm.
  • Cardioplegia dose and total volume counted by the machine. (with the optional module)
  • Patient cooling and rewarming and cardioplegia temperature control through the heat exchangers, with maker-validated cleaning and disinfection instructions. (with the optional module)
  • Venous drainage control with an electrically operated venous line occluder. (with the optional module)
  • Continuous in-line venous oxygen saturation and haematocrit/haemoglobin during bypass. (with the optional module)
  • Pulsatile arterial flow mode. (with the optional module)
  • Power cuts: built-in battery keeps pumps and alarms running for at least 20 minutes, with manual back-up drives for total power failure.
  • Perfusion data sent to an electronic perfusion, anaesthesia or hospital record. (with the optional module)

Not covered: It does not oxygenate blood by itself: the membrane oxygenator, tubing pack and cardioplegia set are single-use consumables bought and priced separately for every case.; It is not meant for support lasting days in the ICU or for transport: that needs an ECMO system.; It does not replace laboratory blood gas and electrolyte tests or clotting-time (ACT) tests during bypass: those need an arterial blood gas analyser and an ACT analyser; in-line monitors show trends only.; It does not collect, wash and return shed blood: that needs a cell-salvage (autotransfusion) system.

5 international manufacturers checkedMarket checked 2026-107 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety and essential performance of medical electrical equipment, for the heart-lung machine and heater-cooler.
  • IEC 60601-1-2 Electromagnetic compatibility, so diathermy, defibrillators and phones do not disturb pumps, sensors or alarms.
  • IEC 62304 Software life-cycle for the software that controls pumps, alarms and interlocks.
  • ISO 14971 Risk management by the manufacturer, covering hazards such as pumping air, reservoir run-dry and power failure.
  • ISO 13485 Quality management system of the manufacturing site.
  • IEC 62353 Electrical safety test at installation and after repair, for acceptance and preventive maintenance in the hospital.
  • DRAP Medical Devices Rules, 2017 Medical devices must be registered or enlisted, and importers licensed, before import and sale in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one maker's brochure figures (exact pump raceway sizes, speed range, number of timers, micrometre roller tolerances) into the tender, which only that maker can meet; ask for functions and minimum performance instead.
  • Buying a model whose maker has announced the end of its heart-lung machine and heater-cooler line; at least one major maker did this in 2025. Ask for a written statement of current production and a 10-year support undertaking.
  • Ignoring heater-cooler infection risk: no maker-validated disinfection instructions, filling with tap water, exhaust blowing toward the open chest, and no hygiene log.
  • Not pricing oxygenators, tubing packs, cardioplegia sets, centrifugal pump heads and in-line monitor sensors for 5 years, so a cheap machine becomes expensive to run.
  • Writing features met by only one or two makers (integrated electronic perfusion record, two-point level detection, pulsatile mode, CO2 blending, mast pumps) into a general tender without justification.
  • Not asking which pumps stop automatically on bubble, level and pressure alarms, and not testing these interlocks and the hand cranks at installation.
  • Buying a single heart-lung machine with no loan-unit clause and no manufacturer-trained engineer in Pakistan, so one breakdown cancels all cardiac surgery.
  • Accepting refurbished, demonstration or grey-market machines without manufacturer authorisation and spare-parts undertaking.