Heated humidified high-flow nasal cannula (HFNC) oxygen therapy system for adult, paediatric and neonatal use

A high-flow nasal oxygen system delivers warmed, fully humidified air-oxygen mixture at up to 60-80 L/min (adults) or 2-25 L/min (children and newborns) through soft nasal cannulas, with oxygen set from 21% to 100%. Pakistani public and private hospitals buy it for ICUs, HDUs, emergency, paediatric wards and NICUs to treat hypoxaemic respiratory failure, bronchiolitis and neonatal respiratory distress and to avoid or follow intubation. The key tender decisions are patient range, integrated blower versus air-oxygen blender (which needs piped medical air), humidifier performance, alarms and battery, and the long-term price of single-patient circuits and cannulas.

What this equipment will do

  • Patients: adults with low-oxygen respiratory failure, before intubation or after extubation, and children with bronchiolitis or pneumonia (flow up to 60 L/min adult, 2-25 L/min child). (with the optional module)
  • Patients: preterm and term newborns and infants, with flow from 2 L/min and neonatal cannulas that leave the nostrils partly open. (with the optional module)
  • Patients with a tracheostomy, through a direct tracheostomy interface. (with the optional module)
  • Therapy: set flow and oxygen (FiO2 21-100%) delivered within stated accuracy, with pressure limited for patient safety (ISO 80601-2-90).
  • Humidification: servo-controlled heated humidifier and heated tube, 34-37 deg C, at least 33 mg/L humidity, gas never above 43 deg C (ISO 80601-2-74).
  • Self-filling water chamber from a sterile water bag, so staff do not open the circuit to refill. (with the optional module)
  • Works without piped medical air, from pipeline or cylinder oxygen, and also from an oxygen concentrator. (with the optional module)
  • Monitoring and alarms: measured flow, delivered FiO2 and gas temperature, with alarms for wrong FiO2, blocked or disconnected circuit, oxygen supply failure, water out and power failure (IEC 60601-1-8, ISO 80601-2-55).
  • Battery backup through power cuts (30 minutes), or 2 hours with an oxygen-cylinder trolley for transport between wards. (with the optional module)
  • SpO2 and pulse with ROX index (ISO 80601-2-61), and automatic FiO2 adjustment to keep SpO2 in a set range. (with the optional module)
  • Bubble CPAP mode for newborns and infants up to 10 kg (4-10 cmH2O). (with the optional module)
  • 72-hour trends and alarm / event log, and data or alarm output to a central station or nurse call. (with the optional module)

Not covered: It is not a ventilator: patients who need pressure support, BiPAP / CPAP by mask or intubation need a non-invasive or ICU ventilator.; It does not produce oxygen; the hospital needs a pipeline from a PSA oxygen plant or liquid oxygen, cylinders or oxygen concentrators.; It does not measure carbon dioxide; CO2 retention needs a blood gas analyser or capnograph.

7 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including electric shock, heating and fire risk in oxygen-enriched use.
  • ISO 80601-2-90 Particular safety and essential performance standard for respiratory high-flow therapy equipment: flow accuracy, oxygen delivery, pressure limits and alarms.
  • ISO 80601-2-74 Particular standard for respiratory humidifying equipment: humidity output, gas temperature limits and humidifier alarms.
  • IEC 60601-1-2 Electromagnetic compatibility, so the device keeps working near other ward and ICU equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds, signals and silencing limits.
  • ISO 80601-2-55 Accuracy and alarms of the built-in or supplied oxygen concentration measurement.
  • ISO 80601-2-61 Pulse oximeter accuracy and alarms where SpO2 is integrated.
  • ISO 11195 Gas mixers (air-oxygen blenders) for medical use, for blender-based systems.
  • IEC 62304 Software life-cycle processes for the device software.
  • ISO 14971 Risk management by the manufacturer for hazards such as overheating, oxygen supply failure and wrong FiO2.
  • ISO 13485 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules 2017 Registration or enlistment required to import and sell the device in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brochure's exact figures, such as a 2-70 or 10-80 L/min range, a screen size or a named auto-fill chamber, so only one model complies. Use 'not less than' values and describe functions, not product names.
  • Demanding a blender-based system for wards with no reliable medical air plant, or an integrated blower when the NICU already runs on blenders, instead of letting the bidder state the design that meets the flow and FiO2 requirement.
  • Not pricing single-patient tube, chamber and cannula sets: the device is often cheap and the proprietary consumables expensive and single-source. Ask for a fixed 3-year price list and cost per patient episode in the evaluation.
  • Assuming any oxygen source will do: high flow at FiO2 0.6-1.0 needs 20-60 L/min of oxygen per bed, which concentrators cannot supply and which can exhaust small cylinder manifolds within hours.
  • Ignoring the device's ambient operating range (some units are rated only up to about 28 deg C); in non-air-conditioned Pakistani wards in summer humidifier control and alarms may not perform as tested.
  • Buying for NICUs without preterm and neonatal cannula sizes, or using adult cannulas on babies, which raises airway pressure and nasal injury risk.
  • Leaving out reprocessing kits, filters and training on disinfection between patients, so devices are either taken out of use or passed between patients without proper cleaning.
  • Accepting DRAP registration of a different model, or of the device without its humidifier, blender or consumables; the exact offered system must be covered.