Specification builder · Tier D

Inductively coupled plasma optical emission spectrometer (ICP-OES) with axial/radial viewing, autosampler and argon supply for elemental impurities (USP <232>/<233>, ICH Q3D) at higher limits, metals in medical device extracts, raw materials and water, with optional closed-vessel microwave digestion

An ICP-OES measures dozens of metals at once by heating the dissolved sample in an argon plasma and reading the light each element emits. In Pakistan it is bought by pharma QC and drug testing labs for elemental impurities in oral products and raw materials, by device testing labs for metals in device extracts, and by water, food and university labs. The key choices are axial/radial viewing, matrix tolerance, data-integrity software, argon supply and sample digestion; where limits fall to a few ug/L or below (for example parenteral class 1 elements), an ICP-MS is needed instead.

What this equipment will do

  • Medicines, APIs, excipients and herbal products prepared as solutions or digests, for elemental impurities at oral-product and higher limits (USP <232>/<233>, Ph. Eur. 2.4.20, ICH Q3D).
  • Extracts of medical devices and their materials (syringes, IV sets, catheters, gloves, polymers) for metals and elemental extractables (ISO 10993-18).
  • Raw materials, purified water, drinking water and effluent for major and trace metals (ISO 11885).
  • Solid samples (tablets, capsules, polymers, food, soil) after closed-vessel microwave digestion (USP <233>, Ph. Eur. 2.4.20). (with the optional module)
  • Preparing device extracts at 37, 50 or 70 deg C (ISO 10993-12 conditions). (with the optional module)
  • Simultaneous multi-element measurement of up to about 70 elements per sample, including the ICH Q3D class 2B and 3 elements (for example Cu, Ni, Cr, Li, Sb, Ba, Mo, V) and majors such as Na, K, Ca, Mg, Fe and Zn (Ph. Eur. 2.2.57, USP <730>).
  • Radial-view measurement of high-matrix samples and of elements at mg/L levels.
  • Axial-view measurement for low ug/L levels of lead, arsenic, cadmium, mercury and nickel. (with the optional module)
  • Arsenic, selenium, antimony and mercury below 1 ug/L by hydride and cold-vapour generation. (with the optional module)
  • Samples containing hydrofluoric acid (silica, talc, glass, soils). (with the optional module)
  • Direct analysis of organic solvent extracts and oils. (with the optional module)
  • Brines, effluent and digests with up to 10 % dissolved solids. (with the optional module)
  • GMP-compliant electronic records with audit trail and e-signatures (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)

Not covered: Ultra-trace (ng/L) levels, such as class 1 elements (Cd, Pb, As, Hg) at parenteral or inhalation limits, need an ICP-MS.; Chemical forms of an element (speciation, for example inorganic arsenic or methylmercury) need HPLC coupled to ICP-MS.; Organic extractables and leachables from devices and packaging need GC-MS and LC-MS; the ICP-OES only measures the elements.; Composition of metal parts (needle steel, implant alloys) without dissolving them needs XRF or a spark emission spectrometer; single elements at very low levels can also be done by graphite furnace AAS.

7 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety requirements for laboratory equipment, covering the high-power RF generator, torch box and accessories.
  • IEC 61326-1 Electromagnetic compatibility (EMC) of laboratory equipment; an ICP-OES contains a high-power RF source next to sensitive detector electronics.
  • ICH Q3D(R2) International guideline setting permitted daily exposures for 24 elemental impurities in medicines, which decide whether ICP-OES is sensitive enough for a given product.
  • USP <232> / USP <233> USP elemental impurity limits and the ICP-OES / ICP-MS procedures and validation criteria (J value, drift, spike recovery, precision) used to test them.
  • Ph. Eur. 2.4.20 European Pharmacopoeia general method for determination of elemental impurities, including closed-vessel digestion and procedure validation, also followed by BP monographs.
  • Ph. Eur. 2.2.57 European Pharmacopoeia general chapter on inductively coupled plasma-atomic emission spectrometry: apparatus, interferences, calibration and system qualification.
  • USP <730> USP general chapter on plasma spectrochemistry: instrument qualification and procedure requirements for ICP-OES and ICP-MS.
  • ISO 10993-18 Chemical characterisation of medical device materials, including metals and elemental extractables measured by plasma spectrometry.
  • ISO 11885 International standard method for determination of selected elements in water by ICP-OES, used by water and environmental labs.
  • USP <1058> Framework for analytical instrument qualification (DQ, IQ, OQ, PQ).
  • 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and electronic signature requirements for software in GMP labs.
  • ISO 17034 / ISO/IEC 17025 Accreditation of reference material producers and testing/calibration labs; standards used for calibration and qualification must come from accredited producers.

Common mistakes in tenders for this equipment

  • Buying ICP-OES for parenteral or inhalation elemental impurity testing without calculating the J value: the class 1 limits (Cd, Pb, As, Hg) are often below what ICP-OES can validate, and the lab then needs an ICP-MS as well.
  • Copying one maker's datasheet (a particular RF frequency, detector type, optical design, torch orientation or argon figure) so that only one model complies; state the measurable detection limits, wavelength range and stability the methods need and accept equivalent designs.
  • Using a maker's trade names for the plasma, viewing or interference-correction system instead of describing the function (for example 'dual view, axial and radial, selectable per wavelength').
  • Specifying axial detection limits on a radial-only instrument, or radial-only viewing for a lab that must reach trace lead, arsenic and mercury limits.
  • Leaving out closed-vessel microwave digestion, certified standards, trace-metal acids, a chiller, exhaust and introduction spares, so the instrument arrives but cannot produce valid results.
  • Ignoring argon logistics: budgeting cylinder argon for an instrument that empties a cylinder in less than a day, instead of a liquid argon dewar with back-up and a supply contract.
  • Accepting 'Part 11 ready' software without requiring the audit trail, unique logins and e-signatures to be demonstrated at installation, or finding later that the compliance module is a separately priced licence.
  • Not asking for a fixed post-warranty CMC price that includes the RF generator and detector, which are the expensive failures after the warranty.