Fully electric multi-function intensive care bed with height, backrest, knee and Trendelenburg / reverse Trendelenburg movements, cardiac chair position, CPR release, integrated scale, split side rails, battery back-up and ICU mattress

An electric ICU bed holds the critically ill patient and lets staff change height, backrest, knee and head-down / head-up tilt at the touch of a button, flatten the bed for CPR in seconds and weigh the patient without lifting. Public DHQ, teaching and private hospital ICUs, CCUs and HDUs in Pakistan buy them. The key tender decisions are safe side rails tested with the supplied mattress, a reliable CPR release, a built-in battery for load-shedding, an accurate scale and the right mattress for pressure-injury risk.

What this equipment will do

  • Patients: critically ill adults in ICU, CCU and HDU, safe working load at least 200 kg (IEC 80601-2-52, intensive care environment).
  • Heavier patients and patients at risk of falls: safe working load 230-250 kg and lowest platform height of 450 mm or less. (with the optional module)
  • Electric positioning: height, backrest 0-65 deg, knee section, Trendelenburg and reverse Trendelenburg at least 12 deg, with angle indicators.
  • One-button cardiac chair position and one-button shock (Trendelenburg) position. (with the optional module)
  • Electric side-to-side tilt of at least 15 deg to help turn the patient. (with the optional module)
  • CPR: backrest flattened from either side of the bed within seconds, also during a power cut.
  • Split side rails tested for entrapment with the supplied mattress, removable head and foot boards for airway access, and lockout of patient controls (IEC 80601-2-52).
  • Patient weighing in bed to +/-1 % or +/-1 kg without lifting the patient (OIML R 76-1).
  • Bed-exit alarm with selectable sensitivity. (with the optional module)
  • Built-in battery for load-shedding: CPR, lowering and at least 10 movement cycles without mains.
  • Patients at high risk of pressure injury: high-specification foam or active alternating-pressure / low-air-loss mattress. (with the optional module)
  • Moving the occupied bed to CT or theatre with steer lock or a powered drive. (with the optional module)

Not covered: Patient monitoring (ECG, SpO2, blood pressure) needs a bedside patient monitor; the bed only positions and weighs the patient.; Newborns and small children need an infant incubator, radiant warmer or paediatric cot; this bed is for adults.; Continuous rotation (kinetic) therapy and air-fluidised therapy for severe burns need specialist therapy beds.; Lifting the patient onto or off the bed needs a patient hoist or transfer board.

7 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including protection against shock, leakage current and moving-part hazards.
  • IEC 80601-2-52 Particular requirements for medical beds: safe working load, stability, side-rail height and entrapment gaps, mattress compatibility, CPR and movement safety. Published May 2026, replacing IEC 60601-2-52:2009; a bidder whose bed was tested to the earlier edition must state the edition and the test report date.
  • IEC 60601-1-2 Electromagnetic compatibility, so the bed does not move unexpectedly near other equipment and does not disturb monitors and ventilators.
  • IEC 62304 Software life-cycle for the bed's programmable control, scale and alarm electronics.
  • IEC 62366-1 Usability engineering, so controls, lockouts, CPR release and alarms are hard to misuse.
  • ISO 14971 Risk management, covering hazards such as entrapment, falls, crushing and failure of CPR release.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Electrical safety testing at installation, after repair and at preventive maintenance.
  • IEC 60529 Ingress protection (IP code) of actuators, control box and hand controls against fluids and cleaning.
  • OIML R 76-1 Metrological requirements for non-automatic weighing instruments, used to judge the accuracy of the integrated bed scale (accuracy class III or better).
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: exact height range to the millimetre, a named lateral-tilt or chair technology, or a particular number of motors. Use 'not less than / not more than' ranges and IEC 80601-2-52 load and entrapment tests.
  • Buying the bed and mattress from different suppliers without checking side-rail height: a thick mattress or air overlay can leave the rail too low and the patient can roll over it.
  • Asking only for 'load 250 kg' without the maximum patient weight for intensive care; the safe working load includes the mattress and accessories, so the patient limit is lower.
  • Accepting electric-only CPR with no mechanical release, or not testing CPR flattening on battery at installation; during load-shedding a dead battery means no CPR position.
  • Leaving out IV poles, oxygen holders, spare covers and hand controls so the cheapest bid arrives as a bare bed; these can add 10-20 % later.
  • Accepting a scale with no stated accuracy or calibration; many low-cost scales are off by several kilograms, which misleads fluid balance and drug dosing.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.
  • No fixed price for batteries, actuators, control boxes and hand controls, which are the usual failures after warranty.