Specification builder · Tier B

Closed infant incubator for NICU and nursery with air and skin servo control, humidity, oxygen and weighing options (including hybrid incubator-warmer)

A closed infant incubator is a heated, humidified, see-through cabinet that keeps premature, low-birth-weight and sick newborns at the right body temperature, with controlled humidity and oxygen and often a built-in scale. Public district nurseries, teaching-hospital NICUs and private hospitals buy them. In Pakistan the key tender decisions are servo humidity and oxygen for very small babies, power back-up for the heater during load-shedding, room cooling in summer, and easy, safe cleaning.

What this equipment will do

  • Patients: newborns from about 500 g to 5 kg, nursed in a closed forced-air cabinet with front door, at least 4 hand portholes and tube entries, low air speed over the baby and limited noise and CO2 build-up (IEC 60601-2-19).
  • Patients: extremely preterm and surgical babies who need many bedside procedures: the unit converts to an open radiant warmer without moving the baby (IEC 60601-2-21). (with the optional module)
  • Temperature: air mode or skin (baby) servo mode using a skin probe, 0.1 deg C display, and an independent cut-out that stops the heater before the air exceeds 38 deg C (IEC 60601-2-19; IEC 60601-1-10).
  • Humidity: built-in humidifier filled from outside the hood without opening the baby compartment, with humidity shown on the display.
  • Servo-controlled humidity set from 40 to 85 % RH for very small babies who lose water through their skin, with low-water and deviation alarms. (with the optional module)
  • Oxygen: oxygen enrichment of the cabinet from pipeline or cylinder through an oxygen inlet.
  • Servo-controlled oxygen set from 21-25 % up to at least 65 %, continuously measured and displayed, with high and low oxygen alarms. (with the optional module)
  • Bedside care: mattress tilt of at least +/-10 deg without opening the hood, and an X-ray cassette tray so X-rays are taken without moving the baby.
  • Weighing the baby inside the incubator without taking it out, with stored weight history. (with the optional module)
  • Full prioritised alarm set for air and skin temperature, over-temperature, probe failure, fan failure, power failure and, where fitted, humidity and oxygen (IEC 60601-1-8). (with the optional module)
  • Keeps heating and running for at least 30 minutes during load-shedding on a dedicated online UPS. (with the optional module)
  • Integrated SpO2 and pulse rate monitoring shown on the incubator screen (ISO 80601-2-61). (with the optional module)

Not covered: Phototherapy for newborn jaundice needs a separate phototherapy unit; the incubator only holds and warms the baby.; Breathing support (ventilation or CPAP) needs a neonatal ventilator or bubble CPAP; the incubator only enriches the oxygen inside the hood.; Moving a baby between hospitals or departments needs a transport incubator; this unit runs on mains or UPS and only moves within the nursery on castors.; Full vital-signs monitoring (ECG, breathing, blood pressure) needs a neonatal patient monitor.

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical, mechanical and thermal safety of medical electrical equipment.
  • IEC 60601-2-19 Particular safety and essential performance of infant incubators (2020 edition with Amendment 1:2023): temperature control, over-temperature cut-out, air flow, CO2, oxygen, humidity and noise.
  • IEC 60601-2-21 Particular requirements for infant radiant warmers; applies to the open-warmer mode of hybrid units.
  • IEC 60601-1-2 Electromagnetic compatibility, so control and alarms work near phones, monitors and other NICU equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and visual signals.
  • IEC 60601-1-10 Physiologic closed-loop controllers; applies to skin-servo (baby-controlled) temperature control.
  • IEC 62304 Medical device software life-cycle for the temperature, humidity and oxygen control software.
  • IEC 62366-1 Usability engineering, so settings, modes and alarms are hard to misuse.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Electrical safety testing at installation and after repair.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: exact humidity or oxygen ranges to the percent, a particular screen size, a rotating mattress or a named air-curtain technology that only one model has. Use 'not less than' ranges and performance limits from IEC 60601-2-19.
  • Assuming the internal battery keeps the baby warm. It only powers the alarm; without a dedicated online UPS or instant generator change-over the heater stops during load-shedding.
  • Not buying skin probes, oxygen sensors, filters and porthole sleeves: incubators then stand idle within a year. Ask for first-year quantities and a fixed 5-year price list.
  • Accepting the incubator at installation without measuring temperature, humidity and oxygen with a calibrated incubator analyser, so there is no proof of performance and no baseline for warranty claims.
  • Ignoring room heat: placing incubators near windows, in direct sun or in rooms without air-conditioning, causing overheating alarms and dangerous air temperatures in summer.
  • Asking for hybrid incubator-warmers, dual skin probes or integrated SpO2 for every bed when only a few beds need them; this cuts competition and doubles the price.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.
  • No written cleaning procedure or list of approved disinfectants: hoods crack from wrong chemicals and humidifier water becomes an infection source.
Medicine brands A–Z:ABCDEFGHIJKLMNOPQRSTUVWXYZ0-9