Infant radiant warmer (open care unit) for newborn resuscitation and nursing, with skin servo and manual control, alarms and optional Apgar timer, resuscitation, scale and SpO2

An infant radiant warmer is an open bed with an overhead infrared heater that keeps newborns warm while staff dry, resuscitate, examine and treat them. Labour rooms, operating theatres, newborn nurseries and NICUs in public and private hospitals buy them; they are the first piece of equipment a baby meets after birth. In Pakistan the key tender decisions are reliable skin-servo control with safe manual-mode alarms, power back-up during load-shedding, skin probes and spare heaters, and whether to add integrated resuscitation, weighing and SpO2.

What this equipment will do

  • Patients: newborns at birth in the labour room or operating theatre who need drying, warming and resuscitation, and sick or preterm newborns who need open access for procedures (IEC 60601-2-21).
  • Warming by an overhead radiant heater controlled by the baby's own skin temperature (skin servo mode, 0.1 deg C steps) through a skin probe (IEC 60601-2-21; IEC 60601-1-10).
  • Pre-warming of the bed before the baby is born, and manual heater mode with a check alarm and automatic heater cut-back after about 15 minutes to prevent overheating.
  • Alarms for high or low skin temperature, over-temperature, probe failure, manual-mode check, heater failure and power failure (IEC 60601-1-8).
  • Bedside care: bed tilt of at least +/-10 deg, fold-down transparent side panels and chest X-ray without moving the baby.
  • Dedicated X-ray cassette tray that slides in and out under the baby. (with the optional module)
  • Height-adjustable bed so staff can work standing or seated. (with the optional module)
  • Central and peripheral skin temperatures shown together to spot cold stress early. (with the optional module)
  • Heated mattress as a second heat source. (with the optional module)
  • Apgar timer with signals at 1, 5 and 10 minutes after birth. (with the optional module)
  • Newborn resuscitation on the warmer: T-piece breathing support with set pressures, oxygen/air mixing or flow, and suction. (with the optional module)
  • Weighing the baby on the warmer without lifting it off. (with the optional module)
  • SpO2 and pulse rate shown on the warmer screen (ISO 80601-2-61). (with the optional module)
  • Blue-light phototherapy for jaundice from a lamp built into the warmer (IEC 60601-2-50). (with the optional module)
  • Keeps heating for at least 20 minutes during load-shedding on a dedicated online UPS. (with the optional module)

Not covered: Long-term care of very small preterm babies (below about 1.5 kg or 30 weeks) who need controlled humidity is done in a closed infant incubator; an open warmer increases their water loss.; Ongoing breathing support (CPAP or ventilation) needs bubble CPAP or a neonatal ventilator; the warmer only supports short resuscitation.; Moving a baby between hospitals needs a transport incubator; the warmer runs on mains or UPS and only moves on castors within the unit.; Routine treatment of newborn jaundice needs a separate phototherapy unit unless the phototherapy option is bought.

7 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical, mechanical and thermal safety of medical electrical equipment.
  • IEC 60601-2-21 Particular safety and essential performance of infant radiant warmers: irradiance limits, skin servo accuracy, manual-mode alarm and over-temperature protection.
  • IEC 60601-1-2 Electromagnetic compatibility, so control and alarms work near phones, monitors and other equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and visual signals.
  • IEC 60601-1-10 Physiologic closed-loop controllers; applies to skin-servo heater control.
  • IEC 62304 Medical device software life-cycle for the heater control and alarm software.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Electrical safety testing at installation and after repair.
  • ISO 80601-2-61 Pulse oximeter safety and accuracy; applies when integrated SpO2 is bought.
  • IEC 60601-2-50 Infant phototherapy equipment; applies when built-in phototherapy is bought.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: a named heater technology, exact heater wattage, a particular screen size or a heated mattress that only one model has. Use 'covering at least' ranges and the safety limits of IEC 60601-2-21.
  • Assuming the internal battery keeps the baby warm. It only powers the alarm; without a dedicated online UPS or instant generator change-over the heater stops during load-shedding.
  • Not buying spare skin probes and heater elements; when they fail, staff use manual mode without a probe, which risks overheating and burns.
  • Accepting the warmer without checking skin temperature accuracy, heater output and the manual-mode alarm at installation, so there is no proof of performance and no baseline for warranty claims.
  • Asking for integrated resuscitation, SpO2 and a scale on every labour-room warmer when only a few beds need them; this cuts competition and doubles the price.
  • Buying integrated resuscitation where there is no pipeline oxygen, air or vacuum and no cylinder holders, so the module is never used.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.
  • Ignoring cleaning: no written procedure or approved disinfectant list, so side panels crack and mattresses become an infection source.