Specification builder · Tier B
Automatic laboratory glassware washer-disinfector (undercounter or freestanding, injector racks, thermal disinfection up to 93-95 deg C, purified-water final rinse, automatic dosing, hot-air drying)
A laboratory glassware washer is a programmable machine that washes, rinses with purified water, thermally disinfects and dries laboratory glassware such as volumetric flasks, pipettes, beakers, HPLC vials and test tubes. Drug testing labs and pharma QC labs buy it to replace hand washing, which leaves detergent and residues that spoil analyses and cannot be documented. The key choices are size, injector racks for narrow-neck glassware, a purified-water supply and pump, conductivity check of the final rinse and how each cycle is recorded.
What this equipment will do
- Open glassware: beakers, conical flasks, reagent bottles, petri dishes, test tubes and HPLC/autosampler vials in baskets and inserts.
- Narrow-neck glassware: volumetric flasks, measuring cylinders and narrow-neck bottles, cleaned inside by injector nozzles. (with the optional module)
- Glass pipettes (volumetric and graduated) on a dedicated pipette injector rack. (with the optional module)
- Automatic wash cycle: pre-rinse, hot alkaline detergent wash, acid neutralisation and rinses, with detergent and neutraliser dosed automatically.
- Final rinse with purified / deionised water so no tap-water minerals or detergent remain on the glass.
- Thermal disinfection of glassware at 90-95 deg C with a set hold time (ISO 15883-1).
- Hot-air drying of the load inside the washer; HEPA-filtered drying air for particle-free glassware (EN 1822-1).
- Check of the final-rinse conductivity in every cycle, with automatic extra rinse or alarm if residues remain. (with the optional module)
- Record of every cycle (temperatures, times, dosing, alarms) on paper, as an electronic file or in a central database for GMP and ISO/IEC 17025 records.
- Cleaning efficacy checks with test soils or washing indicators during OQ/PQ and routine monitoring (ISO 15883-5). (with the optional module)
Not covered: Sterilisation of glassware and media: needs a laboratory autoclave (steam) or a hot air oven (dry heat).; Depyrogenation of glassware for endotoxin testing (about 250 deg C): needs a depyrogenation-rated hot air oven.; Reprocessing of surgical instruments or other medical devices: needs a CSSD washer-disinfector registered for medical use.; Producing purified water: the washer needs a separate RO/DI laboratory water purification system.
Loading the builder...
Standards this specification draws on
- IEC 61010-1:2010+AMD1:2016 Basic electrical, mechanical and fire safety of laboratory equipment, including a washer with high-power heaters and hot water.
- IEC 61010-2-040 Particular safety requirements for sterilizers and washer-disinfectors, covering hot water, steam, chemical and door-interlock hazards.
- IEC 61326-1:2020 Electromagnetic compatibility of laboratory electrical equipment.
- ISO 15883-1:2006+A1:2014 General requirements and tests for washer-disinfectors, including thermal disinfection, temperature control and cycle documentation.
- ISO 15883-5:2021 Test soils and methods for demonstrating the cleaning efficacy of washer-disinfectors, useful for OQ/PQ.
- EN 1822-1:2019 Classification of HEPA filters (H13/H14) used in the drying air path.
- 21 CFR 211.67 US GMP requirement that equipment and utensils be cleaned by written, documented procedures, the basis of glassware cleaning records in GMP labs.
- EU GMP Annex 15 Qualification and validation expectations (IQ/OQ/PQ) for equipment used in GMP manufacture and QC.
- 21 CFR Part 11 / EU GMP Annex 11 Electronic records and audit trail expectations when cycle records are kept electronically in GMP labs.
- USP <1231> Pharmacopoeial guidance on purified water quality, relevant to the final-rinse water.
- ISO/IEC 17025:2017 Competence of testing and calibration laboratories: clean glassware and calibrated equipment underpin valid results and calibration certificates.
Common mistakes in tenders for this equipment
- Copying one maker's brochure: exact chamber dimensions, a named drying or dosing system, or a fixed number of injector positions that only one model has, instead of performance requirements.
- Buying a washer with a single basket and no injector racks, so volumetric flasks and pipettes still have to be washed by hand.
- Specifying a DI final rinse when the lab has no purified-water supply at the washer, or no DI pump for a non-pressurised tank, so the washer cannot run its final rinse.
- Ignoring the site: no three-phase supply, no drain, hard tap water without softening, or no ventilation for steam, discovered only on delivery.
- Leaving out detergent, neutraliser and cleaning verification tools, so the washer cannot be qualified or used after installation.
- Accepting no record of each cycle; GMP and ISO/IEC 17025 labs need temperatures, times and alarms recorded for every load.
- Asking for DRAP medical device registration or ISO 13485, which do not apply to a laboratory glassware washer and wrongly exclude valid bidders.
- Requiring ISO 15883-1 certification when only chemistry glassware is washed; many suitable laboratory models are certified to IEC 61010-2-040 instead.