Specification builder · Tier A

Laboratory refrigerator (+2 to +8 deg C) and -20 deg C laboratory freezer for reagents, reference standards and samples, 90-750 L, upright or under-counter

Purpose-built laboratory refrigerators (+2 to +8 deg C) and -20 deg C freezers keep reference standards, reagents, culture media, vaccines under test and retained samples at their labelled storage temperature, with alarms and a temperature record. Drug testing labs, pharma QC, biologics and university labs buy them. In Pakistan the key choices are uniformity class, manual versus frost-free defrost for freezers, alarms that work through load-shedding, voltage protection and temperature mapping at installation.

What this equipment will do

  • Products stored: pharmacopoeial reference standards, analytical reagents and kits, culture media and sera, vaccines and biologics under test, retained and reserve samples.
  • Refrigerated storage at +2 to +8 deg C (USP <659> 'refrigerator'). (with the optional module)
  • Frozen storage at about -20 deg C, within -25 to -10 deg C (USP <659> 'freezer'). (with the optional module)
  • Both temperatures in one cabinet with two independent compartments. (with the optional module)
  • High and low temperature alarms, audible and visual, with a dry contact to send the alarm to a remote siren, dialler or monitoring system.
  • Door-open and power-failure alarms. (with the optional module)
  • Alarms that keep working during load-shedding on a backup battery. (with the optional module)
  • Tight temperature uniformity and fast recovery after door openings with forced-air circulation (DIN 13277 test method). (with the optional module)
  • Continuous electronic temperature record downloadable by USB. (with the optional module)
  • Independent temperature monitoring with SMS/e-mail alarms and audit trail for GMP records (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)
  • Paper chart temperature record. (with the optional module)
  • Storage of flammable solvents and reagents in closed containers in a spark-free (ATEX-classified) interior. (with the optional module)

Not covered: Blood and red cell units: need a blood bank refrigerator (+2 to +6 deg C) built to blood bank standards and registered with DRAP.; Storage at -40 deg C or -80 deg C (plasma, cell lines, nucleic acids): needs a -40 deg C laboratory freezer or an ultra-low (-80 deg C) freezer.; Stability studies at controlled temperature and humidity (ICH Q1A conditions): need a stability chamber, not a refrigerator.; Incubation of BOD or microbiology samples at a set temperature: needs a BOD or cooled incubator.

10 international manufacturers checkedMarket checked 2026-109 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1:2010+AMD1:2016 Basic electrical and fire safety of laboratory equipment, including refrigerators and freezers.
  • IEC 61010-2-011:2019 Particular safety requirements for laboratory refrigerating equipment, including flammable hydrocarbon refrigerant circuits.
  • IEC 61326-1:2020 Electromagnetic compatibility of laboratory equipment, so controllers and alarms are not disturbed by nearby equipment or generators.
  • DIN 13277:2022-05 Requirements and test methods for refrigerators and freezers for laboratory and medical use, including temperature uniformity and stability tests.
  • USP <659> Packaging and Storage Requirements Defines the 'refrigerator' (+2 to +8 deg C) and 'freezer' (-25 to -10 deg C) storage conditions used on reference standard and reagent labels.
  • WHO TRS 961 Annex 9, Supplement 8: Temperature mapping of storage areas Method for temperature mapping of storage areas and equipment at installation and periodically.
  • ISO/IEC 17025:2017 Laboratory accreditation standard requiring equipment and environmental conditions (including storage of reference materials) to be controlled and records kept.
  • 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and electronic signature expectations for GMP temperature records inspected by DRAP.
  • IEC 60079-0 / Directive 2014/34/EU (ATEX) Explosion protection rules used to classify a spark-free interior for storing flammable substances in closed containers.

Common mistakes in tenders for this equipment

  • Buying domestic refrigerators or freezers for reference standards and reagents: no uniformity, no alarms, no remote contact and frost cycles that warm the contents.
  • Specifying gross volume instead of net usable volume, or a cabinet that does not pass through the laboratory door.
  • Copying one maker's brochure: exact uniformity decimals, named defrost or monitoring systems, or a single maker's capacity in litres instead of a range.
  • No independent temperature record or calibration certificate, so the lab cannot prove storage conditions to auditors or ISO/IEC 17025 assessors.
  • Ignoring power quality: no stabiliser or UPS, and compressor failures blamed on voltage are then refused under warranty.
  • Asking for an explosion-proof or spark-free interior for every cabinet when only a few store solvents, which narrows the field to very few makers.
  • Asking for DRAP medical device registration, which does not apply to laboratory refrigerators and freezers and wrongly excludes valid bidders.
  • No temperature mapping at installation, so standards are stored in warm spots near the door or evaporator.
Medicine brands A–Z:ABCDEFGHIJKLMNOPQRSTUVWXYZ0-9