Specification builder · Tier B

Laminar (unidirectional) air-flow clean bench, horizontal or vertical flow, ISO 14644-1 Class 5 / EU GMP Grade A, with H14 HEPA filtration, for product protection only

A laminar air-flow (LAF) clean bench blows HEPA-filtered air in a steady, one-direction stream over the work surface so that sterile media, plates, preparations and components stay free of particles and microbes (ISO Class 5 / Grade A air). It protects only the product and gives NO protection to the operator or the room, so it must never be used for pathogens, clinical samples or hazardous drugs; those need a Class II biological safety cabinet. Buyers are drug testing laboratories, pharma QC microbiology units, hospital pharmacies doing non-hazardous aseptic preparation, and university teaching labs.

What this equipment will do

  • Work handled: pouring and dispensing sterile culture media, plates and reagents.
  • Work handled: non-pathogenic microbiology such as microbial limit test samples, water and environmental-monitoring samples and media growth promotion.
  • Work handled: hospital-pharmacy sterile compounding of non-hazardous IV admixtures, TPN and eye drops as an ISO Class 5 work area (USP <797>).
  • Work handled: teaching aseptic technique and plant tissue culture with non-pathogenic material.
  • Clean air: work zone meets ISO Class 5 for particles of 0.5 um and larger, checked at installation (ISO 14644-1, ISO 14644-3).
  • EU GMP Grade A airflow: 0.36-0.54 m/s at the working position, every reading within +/-20% of the mean, flow pattern shown by smoke test (EU GMP Annex 1). (with the optional module)
  • Air speed held automatically within +/-10% as filters load and mains voltage varies. (with the optional module)
  • H14 HEPA (99.995%) or better main filter, individually certified and leak-scanned on site (EN 1822 / ISO 29463, ISO 14644-3). (with the optional module)
  • Audible and visual alarm when airflow falls or the blower fails. (with the optional module)
  • UV-C lamp for surface treatment between sessions, with timer and interlock (supplement to chemical disinfection only). (with the optional module)
  • Qualification pack: factory test report, filter certificate and IQ/OQ protocols referencing ISO 14644 and EU GMP Annex 1. (with the optional module)
  • Settle and contact plates for weekly environmental monitoring of the work zone. (with the optional module)

Not covered: Pathogens and clinical samples (sputum, blood cultures, TB work) need a Class II biological safety cabinet, which also protects the operator.; Chemotherapy and other hazardous drugs need a Class II biological safety cabinet or containment isolator (USP <800>).; Pharmacopoeial sterility testing for GMP batch release needs a sterility-testing isolator or a Grade A zone inside a Grade B cleanroom.; Work with solvents, acids or toxic fumes needs a chemical fume hood.

8 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Electrical safety of laboratory equipment.
  • IEC 61326-1 Electromagnetic compatibility (EMC) of laboratory electrical equipment.
  • ISO 14644-1:2015 Classification of air cleanliness by particle concentration; the work zone must reach ISO Class 5.
  • ISO 14644-3:2019 Test methods for installed filter leak, airflow velocity and uniformity, particle count and flow visualisation used at installation and recertification.
  • ISO 14644-2:2015 Monitoring to prove continued cleanliness performance and set requalification intervals.
  • EN 1822-1 / ISO 29463-1 Classification and testing of HEPA/ULPA filters (H14 = not less than 99.995% at MPPS).
  • EU GMP Annex 1 (2022) Grade A requirements for sterile work: unidirectional airflow 0.36-0.54 m/s guidance value at the working position, particle limits and requalification at least every 6 months.
  • WHO TRS 1044, Annex 2 (2022) - WHO GMP for sterile pharmaceutical products WHO sterile-product GMP, aligned with EU Annex 1, used by WHO-prequalification and public-sector programmes.
  • USP <797> Pharmaceutical compounding of sterile preparations: ISO Class 5 primary engineering control and certification every 6 months (hazardous drugs fall under USP <800> and need a BSC or isolator instead).
  • IEST-RP-CC002 Recommended practice for design and testing of unidirectional-flow clean-air devices (clean benches).
  • ISO/IEC 17025:2017 Competence of testing and calibration laboratories; certification instruments must have traceable calibration.

Common mistakes in tenders for this equipment

  • Buying a laminar-flow clean bench as a "biosafety cabinet", or using one for clinical samples, pathogens or cytotoxic drugs; it protects only the product and blows contaminated air at the operator.
  • Copying one brochure's exact figures (a single velocity value, ULPA 99.999%, ISO Class 3, a noise value to one decimal, exact dimensions in mm) so only one brand can comply; specify ranges and minimums from ISO 14644 and Annex 1 instead.
  • Accepting a factory test report without on-site certification by ISO 14644-3 methods (installed HEPA leak scan, velocity and uniformity, particle count) done with calibrated instruments.
  • Making a UV lamp essential or relying on it instead of cleaning with alcohol and sporicides.
  • Placing the bench next to a door, window or air-conditioner outlet that breaks the unidirectional flow, and not checking door widths for 5-6 ft benches.
  • Leaving out prefilter spares, main filter replacement price and 6-monthly recertification, so the bench is never re-tested after delivery and the filter clogs with city dust.
  • Not stating 220-240 V, 50 Hz and voltage tolerance; on unstable supply, benches without constant-airflow control drop below the required velocity unnoticed.