Specification builder · Tier C

Laparoscopy video tower (full HD or 4K UHD, optional ICG near-infrared fluorescence and 3D) with light source, CO2 insufflator, monitor, suction-irrigation, recorder, trolley, telescopes and basic instruments

A complete laparoscopy tower: camera and camera control unit, LED light source, CO2 insufflator, surgical monitor, suction-irrigation, recorder and trolley, with telescopes and a basic instrument set. DHQ, THQ, teaching and private hospitals buy it for keyhole cholecystectomy, appendicectomy, hernia, gynaecology, paediatric and cancer surgery. The key tender decisions are full HD or 4K, whether ICG fluorescence or 3D will really be used, insufflator flow, and protecting the telescopes and camera head through correct reprocessing, warranty and maintenance.

What this equipment will do

  • Patients and operations: adult laparoscopic cholecystectomy, appendicectomy, hernia repair, diagnostic laparoscopy and gynaecology, with 10 mm 0 and 30 degree telescopes (IEC 60601-2-18, ISO 8600 series).
  • Children and neonates: 5 mm and 3-4 mm paediatric telescopes with short 3/5 mm instruments and trocars, for appendix, pyloromyotomy, undescended testis and fundoplication. (with the optional module)
  • Bariatric, long and energy-heavy cases: insufflation flow not less than 45 L/min with CO2 gas heating and smoke evacuation that keeps the set pressure. (with the optional module)
  • Live video: full HD (1920 x 1080) or 4K UHD (3840 x 2160) camera image on a medical-grade surgical monitor, with white balance, zoom and not less than 4 stored surgeon presets.
  • Light: LED cold light through an autoclavable fibre-optic cable, with automatic intensity control from the camera and standby to limit heating at the telescope tip (IEC 60601-2-18).
  • Abdominal gas, suction and irrigation: electronic CO2 insufflation of not less than 30 L/min at 3-25 mmHg with over-pressure venting and alarms, plus saline irrigation and suction through a 5 mm cannula.
  • Recording: video and still images with patient name/ID, exported to USB in standard formats.
  • ICG near-infrared fluorescence: white-light and fluorescence images together, for bowel perfusion at anastomoses, bile duct anatomy and lymph node mapping. (with the optional module)
  • 3D stereoscopic imaging with 2D/3D switching and passive 3D glasses, for suturing in bariatric and pelvic surgery. (with the optional module)
  • Electrosurgery: monopolar and bipolar cutting and coagulation, and vessel sealing up to 7 mm (IEC 60601-2-2). (with the optional module)
  • Images and video sent to the hospital PACS with DICOM worklist. (with the optional module)
  • Teaching and assisting: assistant monitor for the second surgeon and/or live video link to a teaching room with two-way audio. (with the optional module)

Not covered: Flexible endoscopy of the stomach, colon or airways (gastroscopy, colonoscopy, bronchoscopy) needs a flexible video endoscopy system or a video bronchoscope.; Hysteroscopy, cystoscopy and arthroscopy need their own telescopes, sheaths and fluid-management pumps, which are not in this pack.; Cutting, coagulation and vessel sealing are not done by the tower itself; they need an electrosurgical (diathermy) unit, offered here only as an optional add-on.; Sterilising the telescopes, camera head and instruments needs the hospital CSSD steam autoclave or low-temperature sterilizer.

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including medical electrical systems on one trolley.
  • IEC 60601-1-2 Electromagnetic compatibility, so the camera image and insufflator are not disturbed by diathermy and other theatre equipment.
  • IEC 60601-2-18 Particular safety requirements for endoscopic equipment, including light-source heating at the telescope tip, insufflator pressure safety and HF compatibility of endoscopes.
  • ISO 8600 series Optics and photonics standards for medical endoscopes: field of view, direction of view, insertion portion width and optical performance.
  • ISO 17664-1 Manufacturer's information for processing (cleaning, disinfection, sterilisation) of reusable medical devices such as telescopes, camera heads and instruments.
  • IEC 60601-2-2 Particular safety requirements for high-frequency surgical equipment and its accessories, including insulation of monopolar laparoscopic instruments.
  • ISO 14971 Manufacturer's risk management for hazards such as over-pressure, gas embolism, burns from light, and image freeze during surgery.
  • ISO 13485 Certified quality management system of the manufacturers of the camera, insufflator, telescopes and instruments.
  • IEC 62304 Software life-cycle control for camera control unit, recorder and insufflator software.
  • IEC 62353 Electrical safety testing at installation, after repair and during preventive maintenance, including the complete trolley system.
  • DICOM (NEMA PS3) Standard for sending laparoscopy images and video to PACS and receiving patient worklists.
  • DRAP Medical Devices Rules 2017 Pakistan's legal requirement for registration/enlistment and import of medical devices.

Common mistakes in tenders for this equipment

  • Requiring the camera, light source, insufflator, monitor, recorder, suction-irrigation and instruments all from one manufacturer. Only the camera, light source and telescopes need to match; the rest can be specified as compatible from any registered maker.
  • Copying one brand's brochure: proprietary fluorescence mode names, a specific sensor count, exact focal lengths or a fixed lumen figure that only one system quotes. Specify resolution, display modes and functions with 'not less than' values.
  • Asking for 4K or ICG fluorescence for district hospitals that do routine cholecystectomy, and then having no budget for repairs, ICG dye or a 4K monitor.
  • Leaving out telescope and camera head reprocessing: no validated autoclave or low-temperature method, no sterilisation trays and no CSSD training, so optics are damaged within months and claims are refused as misuse.
  • Buying only one 30 degree telescope and one instrument set, so a single breakage or a second case cancels the list.
  • Warranty and CMC that exclude telescopes, light cables and the camera head, which are the parts that fail most.
  • Forgetting CO2 cylinders, central-gas hoses, insufflation filters, drapes and first-year tubing, so the tower cannot be used on delivery.
  • Accepting DRAP registration of the brand or of only the camera; every offered device, including telescopes, insufflator and instruments, must be covered.