Blue-LED phototherapy unit for neonatal jaundice (mobile overhead unit, with under-body pad, double phototherapy and radiometer options)
An LED phototherapy unit shines narrow-band blue light (about 450-490 nm) on a jaundiced newborn's skin to break down bilirubin and avoid exchange transfusion and brain damage (kernicterus). District nurseries, NICUs, children's hospitals and maternity units buy them in numbers. In Pakistan the key tender decisions are a proven intensive irradiance of at least 30 uW/cm2/nm over a large area, a radiometer to check the dose, eye shields, and power back-up so treatment does not stop during load-shedding.
What this equipment will do
- Patients: term and preterm newborns with jaundice (raised unconjugated bilirubin), lying in an open cot, under a radiant warmer or inside an incubator (IEC 60601-2-50).
- Light: narrow-band blue LED light peaking at 450-475 nm, with no meaningful ultraviolet and little heat, so the baby is not burned or dried out (IEC 60601-2-50).
- Dose: intensive phototherapy of at least 30 uW/cm2/nm over an area of at least 800 cm2, with a lower setting for standard phototherapy (AAP 2022 guideline; IEC 60601-2-50).
- Large treatment area of at least 1200 cm2, covering the whole body of a term baby with one lamp. (with the optional module)
- High irradiance of at least 45 uW/cm2/nm for babies with bilirubin near exchange-transfusion level. (with the optional module)
- Positioning: height-adjustable, tilting lamp head on a mobile stand that reaches over a cot, warmer or incubator.
- Lamp placed directly over an incubator hood or on a rail or flexible arm, saving floor space. (with the optional module)
- Treatment timer showing hours of therapy given, kept through short power cuts.
- Stored record of treatment time and lamp operating hours for audit and LED replacement planning. (with the optional module)
- Second light surface from an under-body LED pad, for intensive double phototherapy or treatment while the mother holds the baby. (with the optional module)
- Double phototherapy with two overhead lamps on one baby. (with the optional module)
- Checking the delivered dose at the baby's position with a phototherapy radiometer (uW/cm2/nm). (with the optional module)
- Keeps treating for at least 60 minutes during load-shedding on a dedicated UPS. (with the optional module)
Not covered: It does not measure bilirubin; use a transcutaneous bilirubinometer for screening and a laboratory or bedside bilirubin analyser for serum levels to start and stop treatment.; It does not keep the baby warm; a radiant warmer or incubator is needed for small or sick babies, and temperature must still be checked.; It cannot replace exchange transfusion for very high bilirubin or haemolysis; that needs a blood bank and an exchange-transfusion set.; It is not a UV phototherapy cabinet for adult skin diseases such as psoriasis or vitiligo; those use UVB/UVA dermatology units.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical, mechanical and thermal safety of medical electrical equipment.
- IEC 60601-2-50 Particular safety and essential performance of infant phototherapy equipment (2020 edition with Amendment 1:2023): how irradiance, effective area and uniformity are measured and declared, UV and heat limits.
- IEC 60601-1-2 Electromagnetic compatibility, so the LED drivers do not disturb monitors and other NICU equipment.
- IEC 60601-1-6 Usability collateral standard for medical electrical equipment (applies IEC 62366-1).
- IEC 62366-1 Usability engineering process for medical devices.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer.
- IEC 62353 Electrical safety testing at installation, after repair and at preventive maintenance.
- AAP 2022 Hyperbilirubinemia Guideline (Pediatrics 2022;150(3):e2022058859) Clinical practice guideline for newborns of 35 weeks or more; defines intensive phototherapy as at least 30 uW/cm2/nm in the blue-green band over as much body surface as possible.
- UNICEF Supply Catalogue specification S0002032 (LED overhead phototherapy unit) Public UN procurement specification for a mobile overhead LED phototherapy unit: blue band, standard and intensive irradiance, minimum area, UV limits, timer and accessories.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Asking for irradiance without a measurement distance, band or area: a lamp can show a high number at 20 cm over a tiny spot. Specify at least 30 uW/cm2/nm at the stated treatment distance over a stated area, measured as per IEC 60601-2-50.
- Copying one brand's brochure: exact LED counts, a named light technology, a particular head shape or a built-in meter that only one model has. Use performance limits instead.
- Buying lamps without a radiometer, so the dose is never checked and failing units stay in service for years.
- Accepting broad-band 'blue' lamps or white LEDs; only narrow-band blue light around 450-490 nm treats jaundice efficiently.
- Forgetting eye shields, curtains and spare fuses, so units stand idle or babies are treated without eye protection.
- Placing lamps over incubators or warmers without checking the dose on the mattress and the fit of the stand base under the cot or incubator.
- Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.