Low-temperature hydrogen peroxide (gas plasma / vaporized) sterilizer for heat-sensitive medical devices

A cabinet sterilizer that uses hydrogen peroxide vapour under vacuum, often with a plasma phase, to sterilise reusable devices that steam would damage, such as telescopes, cameras, flexible endoscopes, electronic and plastic instruments. Hospitals buy it for the CSSD, operating theatres and endoscopy units alongside their steam sterilizers. Cycles run at low temperature and, as makers state, the sterilant breaks down mainly to water and oxygen, but each machine is validated only for the loads and lumen sizes in its maker's claim.

What this equipment will do

  • Loads: heat- and moisture-sensitive reusable devices that cannot go in a steam sterilizer, such as rigid telescopes, camera heads, light cables, electronic and powered instruments and plastic devices, at low temperature (maximum stated by the maker) (ISO 14937, ISO 22441).
  • Loads: lumened instruments within the maker's validated claim for lumen material, diameter, length and number per load (ISO 22441).
  • Loads: packed in non-cellulose pouches, wraps or trays validated for hydrogen peroxide.
  • Sterilises at a maximum process temperature of 55 deg C or less. (with the optional module)
  • Single-channel flexible endoscopes on a dedicated programme, and non-lumened items on a fast programme. (with the optional module)
  • Dual-channel flexible endoscopes within the maker's claim. (with the optional module)
  • Rigid stainless steel lumens of 1 mm internal diameter and 500 mm length; single-channel flexible endoscopes of 1 mm x 850 mm. (with the optional module)
  • Records every cycle (programme, process parameters, pass/fail) for load release and audit, with failed cycles marked as failed.
  • Prints the cycle record automatically at the end of each cycle. (with the optional module)
  • Routine monitoring with biological indicators (ISO 11138-1) and chemical indicators (ISO 11140-1) for the hydrogen peroxide process. (with the optional module)
  • Measures hydrogen peroxide concentration in the chamber during each cycle. (with the optional module)
  • Pass-through barrier layout: loads on the packing side and unloads on the sterile side. (with the optional module)
  • Links each cycle to instrument-set or endoscope tracking software. (with the optional module)

Not covered: Linen, paper or other cellulose items, liquids, powders and oils cannot be processed; textiles, wrapped heat-resistant instrument sets and most metal instruments go in the steam sterilizer (CSSD or bench-top).; Cleaning and high-level disinfection of used instruments and endoscopes before sterilisation need a washer-disinfector or an automated endoscope reprocessor; the sterilizer does not clean.; Long or multi-channel lumens outside the maker's validated claim need another method approved by the device maker (for example ethylene oxide sterilisation or high-level disinfection in an endoscope reprocessor).; It is not validated for prion inactivation, and it does not decontaminate rooms or isolators (that needs a room hydrogen peroxide decontamination system).

7 international manufacturers checkedMarket checked 2026-108 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1:2010+AMD1:2016 General electrical, mechanical and fire safety of the sterilizer as laboratory-type equipment (the basis for the sterilizer particular standard).
  • IEC 61010-2-040:2020 Particular safety requirements for sterilizers and washer-disinfectors used to treat medical materials, including door, chamber and toxic-agent hazards.
  • ISO 14937:2009 General requirements for characterising a sterilizing agent and developing, validating and routinely controlling a sterilization process; the basis on which hydrogen peroxide processes have long been validated.
  • ISO 22441:2022 Requirements for development, validation and routine control of low-temperature vaporized hydrogen peroxide sterilization processes for medical devices.
  • ISO 11138-1:2017 General requirements for the biological indicators used to validate and monitor the process.
  • ISO 11140-1:2014 General requirements for the chemical indicators used on packs and inside loads.
  • ISO 13485:2016 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules, 2017 Pakistan registration / enlistment and import requirements for medical devices.

Common mistakes in tenders for this equipment

  • Requiring a plasma phase, one sterilant concentration or one container type, which excludes validated vaporized hydrogen peroxide systems; specify validated claims, temperature and cycle time instead.
  • Comparing chamber sizes without asking usable load volume: makers quote total chamber and usable volume differently, and the gap can be large.
  • Accepting 'suitable for lumens' or 'endoscope compatible' without the maker's claims table (lumen material, diameter, length, channels, number per load) from the instructions for use.
  • Judging price without sterilant cost per cycle: containers give anywhere from one cycle to about thirty, some have a limited life once loaded, and only the maker sells them.
  • Leaving non-cellulose pouches, trays, hydrogen peroxide biological and chemical indicators out of the bid; paper wraps and linen absorb the sterilant and cause aborted cycles.
  • Asking for certification to EN 17180: in September 2026 it was still a draft (FprEN) at formal vote, so no maker can show it. Ask for ISO 14937 or ISO 22441 validation evidence instead.
  • Not checking that each endoscope and camera maker's instructions approve the offered sterilizer and cycle, so scopes are damaged or their warranty is lost.
  • Ignoring power quality: a 3-phase unit on an unstable supply without stabiliser and UPS aborts cycles and wastes sterilant.