Specification builder · Tier C

Mobile surgical C-arm for intra-operative fluoroscopy with image intensifier or flat-panel detector, high-frequency generator, pulsed fluoroscopy, digital image system with storage, dose display and DICOM connectivity

A mobile X-ray unit shaped like a C, with the X-ray tube at one end and an image intensifier or flat detector at the other, that gives live X-ray images during surgery. Orthopaedic and general surgery theatres at THQ, DHQ, tertiary and private hospitals use it to reduce and fix fractures, place screws and nails, and guide urology, pain, ERCP and vascular procedures. The main tender decisions are image intensifier versus flat detector, generator power and tube cooling for long cases, C-arm geometry, dose display, sterile covers and staff radiation protection, and making sure the theatre is covered by the hospital's PNRA licence. Mini C-arms for hand surgery and fixed angiography systems are different products.

What this equipment will do

  • Patients: adults and children in theatre for fracture reduction and fixation, hip and long-bone nailing and general surgery (IEC 60601-2-54).
  • Live pulsed fluoroscopy at 4-8 pulses/s with low-dose mode, automatic brightness control, last-image hold, digital spot images and a 5-minute screening alarm (IEC 60601-1-3).
  • AP, lateral and oblique views without moving the patient: orbital movement not less than 120 deg and angulation +/-180 deg, for orthopaedic, urology, pain and ERCP work.
  • Long cases such as spine, pelvis and complex trauma: continuous low-dose screening for not less than 60 minutes without tube overheating, by active tube cooling. (with the optional module)
  • Obese patients and lateral hip and spine views: generator of not less than 12 kW with rotating anode and high-current pulsed fluoroscopy. (with the optional module)
  • Larger anatomy in one view (pelvis, spine, long bones): flat detector of about 30 x 30 cm. (with the optional module)
  • Vascular and fistula work: digital subtraction angiography (DSA), roadmap and cine loops (IEC 60601-2-43). (with the optional module)
  • Intra-operative 3D: cone-beam 3D images with multiplanar views in theatre within 2 minutes, and a link to surgical navigation. (with the optional module)
  • Patient dose records: DAP, cumulative air kerma and fluoroscopy time per procedure, and a DICOM radiation dose structured report. (with the optional module)
  • Image storage for not less than 10,000 images, with export to USB or CD/DVD in DICOM with a viewer.
  • Images sent to PACS and patient lists received by DICOM worklist. (with the optional module)
  • Centring on the region of interest by laser aimer, without radiation. (with the optional module)

Not covered: Diagnostic X-ray films in the X-ray room or at the bedside need a fixed DR room or a mobile digital X-ray unit; the C-arm is for live imaging during procedures.; Coronary angiography, cardiac stenting and complex neuro or vascular interventions need a fixed angiography (cath lab) system.; Diagnostic cross-sectional images need a CT scanner; the optional C-arm 3D mode is only for checking anatomy and implant position in theatre.

9 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including the mobile stand, monitor cart, cabling and mechanical hazards of a heavy C-arm moved over the patient.
  • IEC 60601-1-2 Electromagnetic compatibility, so the C-arm neither disturbs nor is disturbed by diathermy, anaesthesia monitors and other theatre equipment.
  • IEC 60601-1-3 Radiation protection in diagnostic X-ray equipment: leakage radiation, beam quality, filtration and dose indication.
  • IEC 60601-2-54 Particular safety and essential performance requirements for X-ray equipment for radiography and radioscopy, including mobile C-arms (the device-defining standard).
  • IEC 60601-2-43 Particular requirements for X-ray equipment for interventional procedures, relevant where the C-arm is used for vascular, pain and other interventional work.
  • IEC 62220-1-3 Method for measuring detective quantum efficiency (DQE) of digital X-ray detectors used for dynamic imaging (fluoroscopy), so flat-detector claims can be compared fairly.
  • DICOM (NEMA PS3) Standard for image transfer, worklist, print and dose structured reports between the C-arm, PACS and HIS.
  • IEC 62304 Software life-cycle for the image system, image processing and generator control software.
  • ISO 14971 Risk management of the medical device by the manufacturer.
  • ISO 13485 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
  • PNRA Regulations PAK/904 and PAK/908 Pakistan radiation protection regulations and licensing of radiation facilities: import NOC for the generator, licence covering C-arm use in theatre, local rules, staff dosimetry and dose limits.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: an exact free space, orbital range, kW rating, detector size or detector technology (for example demanding CMOS only) that only one model has. Use 'not less than' values and accept any detector whose DQE is measured to IEC 62220-1-3.
  • Buying a low-power stationary-anode C-arm for spine, pelvis, nailing or vascular work and comparing only kW: ask for the continuous fluoroscopy time before thermal stop and the heat curves.
  • Refurbished, re-badged or old-stock C-arms offered as new. Require a manufacturer authorisation confirming new, current-production equipment and DRAP registration of the complete unit.
  • Accepting a warranty that excludes the X-ray tube, image intensifier or flat detector, or treats them as consumables. These are the parts that fail and cost most.
  • Paid software options after purchase: DICOM worklist, dose reports, vascular package, laser aimer or extra storage sold as licences. Require every listed function to be fully licensed in the bid price.
  • Forgetting theatre radiation protection: no lead glasses, thyroid shields or personal dosimetry, and no PNRA licence amendment or local rules for theatre use. The C-arm then fails PNRA inspection or is used unsafely.
  • Not budgeting sterile C-arm covers or not fixing their price, so the hospital later pays inflated prices for covers that fit only one model.
  • Buying a C-arm for a theatre whose operating table is not radiolucent, so hip and spine images are blocked by the table frame.