Specification builder · Tier C
Mobile / portable digital radiography (DR) unit, battery-powered and motor-driven, with high-frequency generator, collapsible or fixed column, wireless flat-panel detector(s), integrated acquisition console, dose reporting and DICOM / wireless connectivity
A battery-powered X-ray unit on wheels with a wireless flat-panel detector that takes chest, abdomen and limb films at the bedside in ICUs, CCU, emergency, theatres, NICU and wards, and sends them to PACS within seconds. Public hospitals at DHQ and tertiary level, THQs and private hospitals buy it for patients who cannot be moved to the X-ray room. The main tender decisions are generator power, column type, number and quality of detectors, battery and drive, dose reporting, ward radiation protection and making sure the unit is covered by the hospital's PNRA licence. Hand-held and backpack ultra-portable units for community TB screening are a different product.
What this equipment will do
- Patients: adults in ICU, CCU, wards and emergency who cannot be moved, for chest, abdomen, pelvis and limb X-rays at the bedside (IEC 60601-2-54).
- Trauma and theatre patients: cross-table and on-trolley views with a free wireless 35 x 43 cm detector and clip-on grid.
- Babies in incubators and children: dose-optimised paediatric protocols, incubator detector tray, immobilisation aids and a small 24 x 30 cm detector. (with the optional module)
- Large adults, supine abdomen and pelvis: generator of not less than 30 kW for shorter exposures. (with the optional module)
- Large adult chests in either orientation on a 43 x 43 cm detector. (with the optional module)
- Busy rounds: two wireless detectors, so imaging continues while one is in use, being cleaned or charging. (with the optional module)
- Battery operation: motorised drive and exposures on battery, with not less than 100 exposures per charge and image preview within 5 seconds.
- Images sent to PACS/RIS from the ward with DICOM worklist over wired or wireless network; images taken out of range are sent automatically later. (with the optional module)
- Bedside images without a grid: scatter correction software giving contrast close to a grid image. (with the optional module)
- Bone-suppressed soft-tissue chest image from a single exposure. (with the optional module)
- Dose records: DAP, exposure index and deviation index (IEC 62494-1), DICOM radiation dose structured report and reject / repeat analysis. (with the optional module)
- AI reading of chest X-rays (for example pneumothorax, effusion, TB or tube position) with results sent to PACS. (with the optional module)
Not covered: Live X-ray (fluoroscopy) during surgery or procedures needs a mobile C-arm.; Full spine and full leg stitched images, tomosynthesis and dual-energy chest need a fixed DR X-ray room.; Bedside imaging of fluid, soft tissue and the heart without radiation needs a point-of-care ultrasound.; Community TB screening in camps and villages uses hand-held or backpack ultra-portable X-ray units, a different product.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including the battery-powered generator, drive and moving column.
- IEC 60601-1-2 Electromagnetic compatibility, so the unit neither disturbs nor is disturbed by ICU monitors, ventilators and phones.
- IEC 60601-1-3 Radiation protection in diagnostic X-ray equipment: leakage radiation, beam quality, filtration and dose indication.
- IEC 60601-2-54 Particular safety and essential performance requirements for X-ray equipment for radiography and radioscopy, including mobile units (the device-defining standard).
- IEC 62220-1-1 Method for measuring detective quantum efficiency (DQE) of digital X-ray detectors, so detector claims can be compared fairly.
- IEC 62494-1 Exposure index of digital X-ray imaging systems, so radiographers can see over- and under-exposure on every image.
- DICOM (NEMA PS3) Standard for image transfer, worklist, print and dose structured reports between the unit, PACS and RIS.
- IEC 62304 Software life-cycle for the console, image processing and generator control software.
- ISO 14971 Risk management of the medical device by the manufacturer.
- ISO 13485 Quality management system of the manufacturer.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
- PNRA Regulations PAK/904 and PAK/908 Pakistan radiation protection regulations and licensing of radiation facilities: import NOC for the generator, licence covering the mobile unit, local rules for ward use and dose limits.
Common mistakes in tenders for this equipment
- Copying one brand's brochure: an exact detector pixel size, battery exposure count, column height or unit weight that only one model has. Use 'not less than / not more than' values and accept any detector that meets DQE measured to IEC 62220-1-1.
- Buying a low-power portable or ultra-portable unit (a few kW, stationary anode) for ICU and emergency work, or comparing only kW: ask for the mA available, the shortest exposure time and exposures per battery charge.
- Accepting a warranty that excludes dropped or liquid-damaged wireless detectors, or treats drive batteries and detector batteries as consumables. On mobile units the detector and batteries are the parts that fail most and cost most.
- Analogue mobile units converted with a retrofit wireless detector offered as a 'mobile DR'. Require a factory-integrated unit with DRAP registration of the complete system, not the detector alone.
- Paid software options after purchase: Modality Worklist, MPPS, RDSR, wireless DICOM, scatter correction or exposure index sold as extra licences. Require every listed function to be fully licensed in the bid price.
- Forgetting ward radiation protection: no mobile lead shield, aprons or personal dosimetry, and no PNRA licence amendment or local rules for ward use. The unit then fails PNRA inspection or is used unsafely next to other patients.
- Not checking the hospital Wi-Fi and PACS at acceptance: images stuck on the unit, worklist not configured, or the unit not reachable from ICU and ward network points.
- Ignoring infection control: no detector covers and no list of approved disinfectants, so detectors are damaged by harsh cleaning agents or carry infection between ICU patients.