Whole-body 1.5 tesla superconducting magnetic resonance imaging (MRI) system with zero boil-off or sealed low-helium magnet, gradient and RF systems, receive coils, patient table, operator console, post-processing, RF-shielded room and site works

A 1.5 T MRI scanner uses a strong magnet and radio waves, with no X-rays, to image the brain, spine, joints, abdomen, heart and breast in fine detail. Tertiary, teaching, cancer and large divisional hospitals and private diagnostic centres in Pakistan buy it as their main MRI. The key tender decisions are bore size, helium technology (zero boil-off or sealed low-helium), gradient strength, receive channels and coils, the advanced software packages, and a site with a tested RF cage, chiller and back-up power that keeps the magnet cold through load-shedding.

What this equipment will do

  • Patients: adults and children for brain, spine, joint and body MRI on a table rated not less than 200 kg, with SAR and gradient stimulation limits monitored (IEC 60601-2-33).
  • Large, anxious, claustrophobic, uncooperative or child patients: 70 cm wide bore, 250 kg table, motion-robust and quiet sequences. (with the optional module)
  • Routine examinations with the supplied coils: head and neck, whole spine, abdomen and pelvis (torso), knee, shoulder and flexible coils for other parts.
  • Breast, foot and ankle, wrist and hand, cardiac and head-to-toe whole-body coverage with dedicated coils. (with the optional module)
  • Advanced neuro: diffusion tensor tractography, perfusion with and without contrast, susceptibility-weighted imaging and functional MRI. (with the optional module)
  • MR spectroscopy of brain, prostate or breast, with metabolite maps and ratios. (with the optional module)
  • Heart: ECG-triggered cine, perfusion, late gadolinium enhancement, T1/T2 mapping and flow measurement. (with the optional module)
  • Breast cancer: dynamic contrast breast MRI, silicone implant imaging and MR-guided biopsy. (with the optional module)
  • Liver, pancreas and prostate: fat and iron quantification, diffusion, dynamic contrast imaging and MRCP. (with the optional module)
  • Cancer staging and follow-up: whole-body diffusion imaging. (with the optional module)
  • Blood vessels: MR angiography with and without contrast, including peripheral run-off. (with the optional module)
  • Patients with MR-conditional metal implants: metal artefact reduction and implant-conditional scan limits (IEC 60601-2-33). (with the optional module)

Not covered: Fast emergency trauma scanning and detailed lung imaging need a CT scanner.; Patients with MR-unsafe implants or metal fragments cannot be scanned at all and need CT or ultrasound instead.; PET-CT and other nuclear medicine imaging need a PET-CT or gamma camera system.; High-resolution research neuro imaging and some advanced spectroscopy need a 3 T MRI scanner.

7 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including the gradient and RF amplifiers, motorised table and cryogen system.
  • IEC 60601-1-2 Electromagnetic compatibility, so the scanner neither disturbs nor is disturbed by other hospital equipment.
  • IEC 60601-2-33 Particular safety and essential performance requirements for MR equipment: static field, RF heating (SAR), gradient stimulation, acoustic noise, operating modes, quench and emergency field shut-down (the device-defining standard).
  • IEC 62464-1 Methods for measuring MR image quality (SNR, uniformity, distortion, slice thickness, resolution, ghosting) used for acceptance and constancy testing.
  • IEC 62304 Software life-cycle for scan control, reconstruction and safety-limit software.
  • ISO 14971 Risk management of the medical device by the manufacturer.
  • ISO 13485 Quality management system of the manufacturer.
  • ASTM F2503 Marking of items as MR safe, MR conditional or MR unsafe, needed for the injector, monitor, trolleys, cylinders and other items used in the scan room.
  • DICOM (NEMA PS3) Standard for image transfer, worklist and print between the scanner, PACS and RIS.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: an exact helium volume (for example 7 litres), a magnet weight limit, a single bore length, an inflated channel count or named techniques that only one model has. Use 'not less than / not more than' values and require gradient amplitude and slew rate per axis at the same time.
  • Counting coil elements or 'total channels' instead of independent receive channels usable in one field of view; some offers quote 128 or more by adding all coil elements. Ask for both figures and evaluate on independent channels, accepting digital-in-coil designs as equivalent.
  • Buying the magnet before the room is ready: no tested RF cage, no quench pipe route, weak floor, no chiller or precision air conditioning, or no 3-phase supply. The magnet sits in crates or warms up while the warranty runs out.
  • Forgetting that the cold head and chiller must run 24 hours a day: no generator back-up during load-shedding, no standby chiller and no commitment on cold-head repair time, so helium is lost and the magnet may quench.
  • Warranty or CMC that excludes coils, cold head, compressor or helium refill. These are the commonest expensive MRI failures and can cost more than a year of service fees.
  • Leaving out MR-conditional items (injector, patient monitor, trolley, wheelchair, oxygen cylinders) and the ferromagnetic detector, then using ordinary steel equipment near the magnet, which causes projectile accidents.
  • Paid software after purchase: diffusion tensor, perfusion, spectroscopy, cardiac, compressed sensing or deep-learning reconstruction sold as extra licences, or only on the console and not the workstation. Require every listed function to be fully and permanently licensed in the bid price.
  • Accepting DRAP registration of a different model or magnet configuration, or refurbished or end-of-production systems, and skipping acceptance tests of every coil to IEC 62464-1 methods before final payment.