Laboratory muffle (chamber) furnace, 1100-1200 deg C class, 3-15 L and larger, with programmable controller, over-temperature protection and exhaust outlet, for sulphated ash, residue on ignition, total ash, loss on ignition and ashing before metals analysis
A muffle furnace is an electrically heated, insulated chamber that holds crucibles at 500-1200 deg C to burn off organic matter and leave the inorganic ash. Pharmaceutical QC and drug testing laboratories use it for sulphated ash / residue on ignition and loss on ignition; herbal, food, feed and medical-device materials laboratories use it for total ash, ash of plastics and ashing of samples before metals analysis. In Pakistan the choice is decided by chamber size, a programmable controller for slow charring, an independent over-temperature cut-out, a usable exhaust outlet for sulphur trioxide and smoke, and spare heating elements, more than by the maximum temperature.
What this equipment will do
- Sample types: pharmaceutical raw materials, excipients and finished products, herbal drugs and extracts, food, flour and animal feed, plastics and rubber from medical devices, soils and minerals, in porcelain, silica or platinum crucibles of 30-50 mL.
- Sulphated ash / residue on ignition of medicines and raw materials at 600 +/- 50 deg C (USP <281>, Ph. Eur. 2.4.14), including the slow charring and sulphuric-acid steps before ignition.
- Total ash of herbal drugs at 600 +/- 25 deg C (Ph. Eur. 2.4.16) and total ash of articles of botanical origin at 675 +/- 25 deg C (USP <561>), with acid-insoluble ash after filtration.
- Loss on ignition of inorganic raw materials and excipients at the temperature and time given in the monograph, ignited to constant mass (USP <733>).
- Ash of flour, cereals, food and animal feed at 550-600 deg C (AOAC 923.03, AOAC 942.05).
- Ash of plastics and rubber from medical devices and packaging at 600-950 deg C (ISO 3451-1, Method A direct calcination or sulphated ash).
- Dry ashing of samples before dissolving the ash for metals analysis by atomic absorption or ICP (only for metals that do not volatilise at the ashing temperature, such as Fe, Ca, Mg, Zn, Cu).
- Programmed heating: a stored programme with slow ramps and dwells (for example 2 deg C/min to 250 deg C, hold, then to 600 deg C) so that samples char without spitting and the same cycle is run every time. (with the optional module)
- Ashing of large or fatty organic loads (feed, oil, polymer) with forced air flow through the chamber so that combustion is complete and faster. (with the optional module)
- Work at up to 1200 deg C (for example ignition of refractory oxides or ceramics research). (with the optional module)
- Temperature record of every run for the test report: logged chamber temperature, programme and alarms, exported as CSV or PDF. (with the optional module)
- Removal of acid fumes and smoke through a chimney or duct to a fume extraction system outside the room. (with the optional module)
Not covered: Weighing the ash to constant mass needs an analytical balance (0.1 mg) and a desiccator; the furnace only ignites the sample.; Loss on drying or moisture content at 105 deg C needs a drying oven or moisture analyser, not a furnace.; Wet acid digestion of samples for arsenic, mercury, cadmium and other volatile elements needs a microwave digestion system; dry ashing loses these elements.; Melting, sintering or heat treatment above about 1200-1300 deg C needs a high-temperature chamber furnace with silicon-carbide or molybdenum-disilicide elements.
Loading the builder...
Standards this specification draws on
- IEC 61010-1:2010+AMD1:2016 Basic electrical, mechanical and fire safety for laboratory equipment; the safety baseline for any furnace.
- IEC 61010-2-010:2019 Particular safety requirements for laboratory equipment for the heating of materials: over-temperature protection, hot surfaces, insulation and fire containment.
- IEC 61326-1:2020 Electromagnetic compatibility (EMC) of laboratory equipment, so the controller and limiter work reliably near other equipment.
- USP <281> / Ph. Eur. 2.4.14 (sulphated ash, residue on ignition) Defines the ignition at 600 +/- 50 deg C to constant mass, with charring and sulphuric-acid treatment, that the furnace and its accessories must support.
- USP <733> (loss on ignition) Loss on ignition to constant mass at the temperature given in the monograph, for inorganic raw materials such as excipients and minerals.
- Ph. Eur. 2.4.16 / USP <561> (total ash of herbal drugs) Total ash of herbal drugs at 600 +/- 25 deg C (Ph. Eur.) and 675 +/- 25 deg C (USP), plus acid-insoluble ash, for herbal and alternative medicines.
- AOAC 923.03 / AOAC 942.05 (ash of flour and animal feed) Reference ash methods for food and feed at 550-600 deg C that food, feed and nutrition laboratories follow.
- ISO 3451-1:2019 Determination of ash of plastics (direct calcination or sulphated ash) used for polymer parts of medical devices and packaging.
- IEC 60519-2 Safety in electroheat installations, particular requirements for resistance heating equipment; defines the classes of temperature limiting device quoted by furnace makers.
- DIN 17052-1 Defines how temperature uniformity of a furnace working space is measured, so that uniformity figures in bids are comparable.
- ISO/IEC 17025:2017 General requirements for the competence of testing laboratories, including traceable calibration of furnaces used for tests.
- ISO 9001:2015 Quality management system of the manufacturer.
Common mistakes in tenders for this equipment
- Specifying the maximum temperature (1200 or 1300 deg C) as the main criterion when every ash and ignition method runs at 500-950 deg C; the higher class costs more, may need platinum-type thermocouples and shortens element life.
- Copying one maker's exact figures (a fixed litre size such as 9 L or 13 L, a named controller or software, a fixed number of programme segments) so that only one brand qualifies; use 'not less than' sizes and performance.
- Leaving out the exhaust outlet and ducting, so sulphur trioxide fumes and smoke go into the room and corrode the analytical balance within months.
- Accepting a controller software alarm as 'over-temperature protection' instead of an independent limiter with its own sensor and relay, then running programmes overnight.
- Forgetting that sulphated ash per the pharmacopoeias needs silica or platinum crucibles, a charring hot plate in the fume cupboard and a desiccator; the furnace alone cannot run the test.
- No spare heating elements, thermocouple or relay in the bid and no element price list, so the first element failure stops release testing for months while parts are imported.
- No calibration at the pharmacopoeial set points and no accredited certificate, then failing an ISO/IEC 17025 assessment or a DRAP GMP inspection on the furnace record.
- Ignoring the electrical supply: 15 L and larger furnaces often need three-phase power, and a 3-10 kW heater on an undersized stabiliser or a long thin cable trips and burns connectors.