Neuronavigation (image-guided surgery) system for cranial and spine surgery with optical (and optional electromagnetic) tracking, CT/MRI planning and fusion software, patient registration, navigated instruments, DICOM/PACS interface and optional microscope, ultrasound and functional modules

An image-guided surgery platform that matches the patient's CT and MRI scans to the patient on the table and shows the position of the surgeon's instruments in real time during brain and spine surgery. Teaching, tertiary and private hospitals with neurosurgery and spine units buy it for tumour resection, biopsy, shunts and navigated screw fixation. The key tender decisions are cranial, spine or both, optical or optical plus electromagnetic tracking, avoiding lock-in to one maker's instruments and disposables, and securing case support, accuracy testing and long-term software support.

What this equipment will do

  • Brain surgery: localisation of tumours and other intracranial lesions and planning of the craniotomy. (with the optional module)
  • Spine surgery: navigated pedicle screw placement. (with the optional module)
  • Images handled: the patient's CT and MRI scans from the scanner, PACS or portable media, with the instrument position shown on them and target or trajectory planning (DICOM).
  • Real-time instrument tracking with an optical camera, with accuracy stated by the bidder and tested on a phantom (ASTM F2554).
  • Electromagnetic tracking of stylets for ventricular catheter placement, without line of sight. (with the optional module)
  • Patient registration by landmarks, fiducials and/or surface matching, with an accuracy check.
  • Automatic registration from intra-operative CT or cone-beam CT, mainly for spine. (with the optional module)
  • Frameless navigated brain biopsy. (with the optional module)
  • Diffusion MRI fibre tracking to show motor and language pathways near a tumour. (with the optional module)
  • Navigation shown inside the operating microscope view (focal point and outlines). (with the optional module)
  • Navigated intra-operative ultrasound to check brain shift and residual tumour. (with the optional module)
  • Deep brain stimulation and stereo-EEG electrode trajectory planning. (with the optional module)

Not covered: It does not take any images; CT, MRI and any intra-operative CT or cone-beam CT scanner are separate purchases.; It does not hold or drive instruments; robotic guidance systems and stereotactic frames for deep brain stimulation are separate equipment.; It does not give the magnified view; that comes from the operating microscope or a neuro-endoscope (see Neurosurgical microscope).; Nerve and spinal cord function monitoring during surgery needs an intra-operative neurophysiological monitoring system.

5 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of the navigation cart, tracking camera and EM field generator used next to the patient.
  • IEC 60601-1-2 Electromagnetic compatibility, so tracking is not disturbed by diathermy, microscopes and other theatre equipment.
  • IEC 62304 Software life-cycle control for the navigation and planning software, whose errors directly misplace instruments.
  • IEC 62366-1 Usability engineering, so registration, accuracy warnings and screens are designed to prevent user error.
  • IEC 81001-5-1 Security activities in the health software life-cycle, for a networked computer holding patient scans.
  • ASTM F2554 Standard practice for measuring and reporting the positional accuracy of computer-assisted surgical systems, so accuracy claims can be compared.
  • ISO 14971 Manufacturer's risk management for hazards such as registration error, loss of tracking and wrong image data.
  • ISO 13485 Certified quality management system of the navigation system and instrument manufacturer.
  • IEC 62353 Electrical safety testing at installation, after repair and during preventive maintenance.
  • DICOM Standard for importing CT and MRI scans from scanners and PACS into the planning and navigation software.
  • DRAP Medical Devices Rules 2017 Pakistan's legal requirement for registration/enlistment and import of medical devices.

Common mistakes in tenders for this equipment

  • Specifying one maker's navigated instruments, implants or screw system, so only that maker can bid and every later instrument must come from it. Ask for universal instrument adapters and a compatibility list for the hospital's own screw systems.
  • Copying one maker's brochure: camera model, proprietary tracking or registration names, exact monitor size or number of screens. Specify tracking type, a stated accuracy tested by a stated method, and the required functions.
  • Buying automatic spine registration without a compatible intra-operative CT or cone-beam CT, so the system is used only with slower manual registration.
  • Leaving out disposables (marker spheres, drapes, EM stylets) and their prices, which are sold only by the system maker and can cost more than the service contract over 5 years.
  • Training for two days and no case support, so the system sits unused; require an application specialist for the first 20 cases and a refresher.
  • No accuracy test at acceptance or during maintenance; require a phantom accuracy test at installation, after repair and at each preventive maintenance.
  • Time-limited software licences, annual subscriptions or an operating system near end of support; require perpetual licences, security patches and a 10-year support undertaking.
  • Accepting DRAP registration of the cart only; every software application and the navigated instruments must also be covered.