Specification builder · Tier B

Hospital non-invasive ventilator for ward, HDU and ICU use, with CPAP and bilevel spontaneous/timed modes, leak compensation and oxygen enrichment

A hospital non-invasive ventilator (often called a 'BiPAP machine') supports breathing through a tight-fitting mask or helmet, giving CPAP or two pressure levels with a backup breath rate, while compensating for mask leaks and adding oxygen. Public and private hospitals buy it for medical wards, HDUs, ICUs and chest units to treat COPD exacerbations, pulmonary oedema, obesity hypoventilation and to avoid intubation. It is not the same as a home sleep-apnoea CPAP unit, which has no backup rate, oxygen control or ventilation alarms.

What this equipment will do

  • Patients: adults (from 30 kg) with acute or chronic breathing failure, treated by mask or helmet in hospital; sleep-apnoea-only devices are excluded (ISO 80601-2-80).
  • Patients: children from 13 kg, with child default settings and alarm limits. (with the optional module)
  • Modes: CPAP, bilevel spontaneous (S), spontaneous/timed (S/T) with back-up rate up to 40 bpm, and timed pressure control, with automatic mask-leak compensation; inspiratory pressure up to 30 cmH2O (40 cmH2O with the higher-pressure options).
  • Volume-targeted mode that adjusts pressure support automatically to reach a set tidal volume, for chronic hypercapnic patients. (with the optional module)
  • Monitoring and alarms: pressures, estimated tidal and minute volume, leak, rate and FiO2 on screen, alarms for high/low pressure, disconnection, low volume, apnoea and power failure (IEC 60601-1-8), and at least 30 minutes on battery (2-6 hours with larger battery options).
  • Precise oxygen 21-100% from piped or cylinder high-pressure oxygen, with measured FiO2 and oxygen supply failure alarm (ISO 80601-2-55). (with the optional module)
  • Oxygen from a flowmeter or oxygen concentrator (up to 15 L/min low-pressure inlet), for hospitals without reliable piped oxygen. (with the optional module)
  • Pressure, flow and volume waveforms and 72-hour trends on a large touch screen. (with the optional module)
  • High-flow nasal oxygen 10-60 L/min on the same device. (with the optional module)
  • Heated humidification for NIV and high-flow therapy (ISO 80601-2-74). (with the optional module)
  • SpO2 and pulse on the ventilator screen with trends and alarms (ISO 80601-2-61). (with the optional module)
  • Ventilation of patients with a tracheostomy, using a valve or dual-limb circuit. (with the optional module)

Not covered: Patients who need intubation and full invasive ventilation in ICU need an ICU ventilator (ISO 80601-2-12).; Newborns and infants below 13 kg need a neonatal ventilator or a bubble CPAP system.; It does not produce oxygen; it needs piped oxygen, cylinders or an oxygen concentrator.; Blood gases (PaCO2) to judge whether NIV is working need a blood gas analyser.

8 international manufacturers checkedMarket checked 2026-1012 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment.
  • ISO 80601-2-80 Particular safety and essential-performance standard for ventilatory support equipment for ventilatory insufficiency (bilevel NIV); certification of the same model to ISO 80601-2-12 (critical care) or ISO 80601-2-72 (life-support home ventilator) is accepted as equivalent or higher.
  • ISO 80601-2-79 Particular standard for ventilatory support equipment for ventilatory impairment; relevant to lower-acuity bilevel devices and helps separate them from sleep-apnoea devices.
  • IEC 60601-1-2 Electromagnetic compatibility, so the ventilator is not disturbed by other ward and ICU equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and signals.
  • ISO 80601-2-74 Heated humidifiers supplied with the ventilator or for high-flow therapy.
  • ISO 80601-2-55 Respiratory gas monitors, covering the accuracy of measured FiO2.
  • ISO 17510 Masks and application accessories for positive airway pressure therapy.
  • IEC 62304 Software life-cycle processes for medical device software.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • DRAP Medical Devices Rules, 2017 Registration/enlistment required to import and sell the ventilator in Pakistan.

Common mistakes in tenders for this equipment

  • Accepting home sleep-apnoea CPAP, auto-CPAP or basic bilevel units (built to the sleep-therapy standard) as hospital NIV ventilators: they lack a backup rate, set inspiratory time, oxygen control and ventilation alarms.
  • Copying the brochure numbers of a once-dominant hospital NIV model that was discontinued in 2022 (service only until 2029), so that no current product matches or only one maker can comply.
  • Naming proprietary mode or synchrony feature names (for example a maker's volume-assured or automatic-trigger mode) instead of describing the function, which locks the tender to one brand.
  • Buying low-flow oxygen models for wards where patients need high FiO2, or blender-only models for hospitals that run on concentrators and cylinders.
  • Not specifying masks, sizes and vented/non-vented matching to the circuit exhalation port; a wrong combination causes CO2 rebreathing or huge leaks and the ventilator is blamed.
  • Leaving out the heated humidifier, circuits, filters and oxygen sensors from the bid, so a cheap ventilator cannot be used or becomes expensive to run.
  • Demanding very long battery time or a large screen without clinical need, which few makers meet and inflates price.
  • No written spare-parts and service guarantee, no local trained engineer and no post-warranty CMC price.