Mobile electro-hydraulic universal operating table with battery back-up, radiolucent top for C-arm imaging, and orthopaedic, neurosurgical and bariatric options

An electro-hydraulic operating table holds and positions the patient during surgery, with powered height, head-down/head-up tilt, side tilt and back movements, and a top that lets X-ray or a C-arm see through it. Every hospital theatre from THQ to tertiary and private hospitals buys them. In Pakistan the key tender decisions are a safe load rating for every position, a built-in battery that keeps the table moving through load-shedding, C-arm access, and the right specialty extensions.

What this equipment will do

  • Patients and uses: adults up to at least 225 kg for general, gynaecological, urological, orthopaedic and neurosurgical operations, locked firmly to the floor with no tipping at rated load (IEC 60601-2-46).
  • Heavier patients: safe working load at least 300 kg (450 kg in the bariatric class). (with the optional module)
  • Powered positioning: height about 720-1,000 mm, head-down and head-up tilt 25 deg, side tilt 18 deg, back section, flex and reflex, and split leg plates that drop and spread for lithotomy (gynae and urology).
  • Steeper tilt (30 deg), more side tilt (20 deg) and a lower table for laparoscopy and neurosurgery. (with the optional module)
  • X-ray during surgery: radiolucent back and seat with a side-loading X-ray cassette tunnel; head and leg plates swap ends for reverse patient position.
  • C-arm imaging from pelvis to head with a powered top slide of at least 300 mm. (with the optional module)
  • Carbon-fibre imaging zone at least 900 mm long with no metal, for full 360 deg C-arm views. (with the optional module)
  • Keeps moving through load-shedding on its built-in battery: at least 30 movement cycles or one operating day without mains.
  • Emergency control that still raises, tilts and levels the table if the hand control or electronics fail. (with the optional module)
  • Hip, femur and tibia fracture fixation with an orthopaedic traction extension and C-arm access to hip and femur. (with the optional module)
  • Cranial surgery in a 3-pin skull clamp, and prone and sitting neuro and spine positions with special headrests. (with the optional module)
  • Kidney surgery: flank raised at least 100 mm with a kidney elevator. (with the optional module)

Not covered: It does not take X-rays; intra-operative imaging needs a mobile C-arm, the table only lets X-rays pass through.; It is not a transport trolley: patients are brought to theatre on a patient trolley and moved across with a transfer board.; Dedicated open spine frames for long prone spine fusion and fixed imaging tables for hybrid theatres are separate specialist tables.; Keeping the patient warm during surgery needs a forced-air patient warming system.

8 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including protection against shock, crushing and moving-part hazards.
  • IEC 60601-2-46 Particular requirements for operating tables (2016 edition): safe working load, stability against tipping, crushing points, movement stopping and load testing.
  • IEC 60601-1-2 Electromagnetic compatibility, so the table does not move unexpectedly near diathermy units and does not disturb monitors.
  • IEC 62304 Software life-cycle for the programmable control electronics.
  • IEC 62366-1 Usability engineering, so hand-control keys, lock and load limits are hard to misuse.
  • ISO 14971 Risk management, covering hazards such as patient fall, crushing, tipping and unintended movement.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Electrical safety testing at installation, after repair and at preventive maintenance.
  • IEC 60529 Ingress protection rating (IP code) of the base, column and hand control against cleaning fluids and blood.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one brand's brochure: exact height range to the millimetre, a particular number of motorised movements, a named cordless-control or drive technology, or a specific slide length that only one model has. Use 'not less than / not more than' ranges and IEC 60601-2-46 load and stability tests.
  • Asking only for 'maximum load 450 kg' without the load allowed in every position; many tables carry the headline load only when level and not slid. Ask for the safe working load with all movements permitted.
  • Buying a mains-only table or ignoring battery capacity: during load-shedding a stuck table in steep head-down or with an open abdomen is dangerous. Specify the built-in battery and number of movement cycles.
  • Leaving out accessories (arm boards, anaesthesia screen, leg holders, rail clamps, traction or neuro modules) so the cheapest bid arrives as a bare table; accessories can add 20-40 % later.
  • Ordering specialty modules (traction, neuro, width extensions) from a third party that the table maker has not approved; the load rating and warranty may then not apply.
  • Accepting the table without testing every movement, the floor lock and battery operation against the specification, so there is no baseline for warranty claims.
  • Accepting DRAP registration of a different model or of the brand only; the exact offered model must be registered or enlisted.
  • No fixed price for batteries, hand controls and hydraulic seals, which are the usual failures after warranty.