Specification builder · Tier C

Floor-stand operating microscope for eye surgery (cataract, glaucoma, cornea and retina with optional non-contact wide-angle viewing) or for ENT and otology surgery, with LED illumination, motorised controls, assistant microscope and optional HD/4K camera

A floor-stand operating microscope that gives the surgeon a magnified, brightly lit stereo view for eye surgery (cataract, glaucoma, cornea and retina) or for ENT and ear surgery. Eye departments, ENT departments and private eye centres buy it; eye and ENT versions differ in working distance, red-reflex lighting and controls, so the tender must say which. The key decisions are step or motorised zoom and X-Y, red-reflex quality and retinal light safety for eye work, the working distance for ENT, an assistant scope and camera for teaching, and wide-angle viewing for retina.

What this equipment will do

  • Eye operations: small-incision cataract, phacoemulsification, glaucoma, cornea and pterygium surgery at 175-200 mm working distance (ISO 10936-1, ISO 10936-2). (with the optional module)
  • ENT operations: tympanoplasty, mastoidectomy, stapes, microlaryngeal and nasal surgery at 200-430 mm working distance (ISO 10936-1). (with the optional module)
  • Magnified stereo view by step changer or zoom, with the magnification range stated by the maker.
  • Motorised zoom and X-Y movement of not less than 50 x 50 mm, for phaco and retina work. (with the optional module)
  • LED illumination, with retinal light-hazard data for eye use (ISO 10936-2, ISO 15004-2).
  • Red reflex: coaxial illumination producing a red reflex, for seeing the lens capsule during cataract surgery. (with the optional module)
  • Hands-free control of the motorised functions by a foot control with not less than 12 functions (wired or wireless). (with the optional module)
  • Teaching and assisting: assistant microscope showing the same field to a second doctor. (with the optional module)
  • Video: full HD or 4K camera with monitor and recording of video and stills with patient ID, and live teaching display. (with the optional module)
  • Retina and vitrectomy: non-contact wide-angle fundus view with stereo image inversion. (with the optional module)
  • Intraoperative OCT cross-section images of the eye during surgery. (with the optional module)

Not covered: The microscope does not remove the cataract or vitreous: phacoemulsification and vitrectomy need a separate phaco or vitrectomy machine.; Pre-operative eye measurements (biometry and IOL power, keratometry) and clinic slit-lamp examination need an optical biometer and a slit lamp, not this microscope.; Brain and spine surgery need a neurosurgical microscope with longer working distance and fluorescence modules, which is a separate pack.; Endoscopic sinus and ear surgery (FESS, endoscopic ear surgery) need rigid ENT endoscopes with a camera tower.

7 international manufacturers checkedMarket checked 2026-109 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including stability of the floor stand and protection of the patient and staff.
  • IEC 60601-1-2 Electromagnetic compatibility, so motor controls, camera and LED drivers are not disturbed by, and do not disturb, phaco machines, diathermy and other theatre equipment.
  • ISO 10936-1 Requirements and test methods for operation microscopes (ISO 10936-1:2017): optical performance, magnification, field of view and markings.
  • ISO 10936-2 Light hazard from operation microscopes used in ocular surgery (ISO 10936-2:2010; a revision is in progress), limiting retinal light injury during long eye operations.
  • ISO 15004-2 General light-hazard protection requirements for ophthalmic instruments (ISO 15004-2:2024, which replaced the 2007 edition); the microscope-specific ISO 10936-2 takes precedence where they differ.
  • ISO 14971 Manufacturer risk management for hazards such as retinal light injury, stand tipping or arm drift, and loss of view mid-procedure.
  • ISO 13485 Certified quality management system of the microscope manufacturer.
  • IEC 62353 Electrical safety testing at installation, after repair and during preventive maintenance.
  • DRAP Medical Devices Rules 2017 Pakistan's legal requirement for registration/enlistment and import of medical devices.

Common mistakes in tenders for this equipment

  • Writing one specification for both eye and ENT: eye needs a 175-200 mm working distance, red reflex and X-Y; ENT needs 250-300 mm or variable working distance. A combined tender either excludes good bids or buys a microscope that suits neither.
  • Copying one brand's brochure: exact stereo base, a fixed magnification range to one decimal, named illumination modes or a specific number of foot-switch functions that only one maker quotes. Use 'not less than' values that several makers meet.
  • Asking for intraoperative OCT, 3D or 4K heads-up displays for district cataract units, then having no budget for service or training.
  • Leaving out the red reflex, retina-protection filter and light-hazard data (ISO 10936-2), which matter more for cataract outcomes and patient safety than extra magnification.
  • Forgetting sterile drapes, autoclavable handle covers and a dust cover, so the microscope cannot be used sterile on delivery or grows fungus within a year in a humid theatre.
  • No optical alignment, balancing and electrical safety check in the acceptance report, so faults are found only after the warranty starts running.
  • Accepting DRAP registration of the brand only; the exact microscope model and any wide-angle viewing system must be covered.