Specification builder · Tier B
Battery-powered modular orthopaedic power tool system with cannulated drill/reamer, sagittal and reciprocating saws, wire/pin driver, attachments, aseptic or sterilisable batteries and chargers for trauma and arthroplasty
An orthopaedic power tool system is a set of battery-powered drills, reamers, saws and wire drivers that orthopaedic surgeons use to fix fractures and replace hips and knees. Every orthopaedic theatre needs one, from THQ fracture lists to tertiary arthroplasty units. The key tender decisions are which handpieces are really needed, the battery method (aseptic or sterilisable), attachments that fit existing instruments, and the long-term price of saw blades and batteries.
What this equipment will do
- Patients: adults and children having fracture fixation, osteotomy and joint replacement in the operating theatre (IEC 60601-1).
- Bones: long bones, pelvis and acetabulum, and small bones of the hand and foot (with the small-bone option).
- Drilling screw holes at variable speed, including cannulated drilling over guide wires, with speed and cannulation as stated in the maker's data sheet.
- Reaming of the intramedullary canal and acetabulum with a reaming attachment, with declared reaming speed and torque.
- Very high-torque acetabular and large intramedullary reaming, not less than 15 Nm. (with the optional module)
- Bone cutting with a saw attachment for fracture and small-bone osteotomy. (with the optional module)
- Bone cutting with a dedicated sagittal saw for osteotomy and knee or hip replacement cuts. (with the optional module)
- Reciprocating saw cuts for femoral neck, tibial and humeral cuts and implant removal. (with the optional module)
- Driving Kirschner wires and pins with a wire driver attachment or collet.
- Fine wire and small-bone work in hand, foot and paediatric surgery with a dedicated small handpiece set. (with the optional module)
- Median sternotomy with a guarded sternum saw, so cardiac surgery can share the system. (with the optional module)
- Digital measurement of drilled depth during plating. (with the optional module)
- Batteries used in the sterile field by aseptic transfer or as sterilisable batteries; handpieces reprocessed by validated steam sterilisation (ISO 17664-1, ISO 17665).
Not covered: High-speed burring, craniotomy and spine bone work at tens of thousands of rpm needs a high-speed neurosurgical/spine drill system.; Arthroscopic shaving and burring inside the knee or shoulder needs an arthroscopy shaver console.; Removing plaster casts needs a separate plaster (cast) cutter, not a surgical saw.; Implants, drill bits, reamer heads and guide wires are bought with the implant system or as consumables, not with the power tools.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General basic safety and essential performance of medical electrical equipment, applied to battery handpieces and chargers.
- IEC 60601-1-2 Electromagnetic compatibility so motors and chargers do not disturb monitors, navigation or diathermy and are not disturbed by them.
- ISO 17664-1 Information the manufacturer must give for cleaning, disinfection and sterilisation of handpieces, attachments and sterilisable batteries.
- ISO 17665 Validation and routine control of moist-heat (steam) sterilisation used for handpieces and attachments.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer.
- IEC 62353 Recurrent and after-repair electrical safety testing of chargers and other mains-powered parts.
- DRAP Medical Devices Rules 2017 Registration required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Copying one maker's handpiece weights, exact rpm or cycle figures, trademarked system names or proprietary blade hub names; use minimum speeds, declared torque and blade compatibility instead.
- Requiring a single battery method (only aseptic or only sterilisable batteries) or only lithium-ion chemistry without a clinical reason; this removes good bids.
- Forgetting attachments that fit the hospital's existing reamers and implant drill bits (AO-type, Hudson-type couplings), so the new drill cannot be used with existing instrument sets.
- Buying too few batteries or no multi-bay charger, so lists stop when batteries run flat or are still being charged.
- Not pricing saw blades: proprietary blade hubs can make blades the largest running cost over 5 years.
- No validated steam sterilisation instructions or CSSD training; wrong cleaning of cannulations and over-long autoclave cycles destroy seals, batteries and motors.
- Accepting a universal single handpiece with attachments for a joint replacement service where dedicated saw and drill handpieces are needed.
- Accepting DRAP registration of one handpiece only; all offered handpieces, batteries and chargers must be covered.