Specification builder · Tier A
Stationary medical oxygen concentrator, 5-10 L/min, mains-powered, for hospital wards, primary care and home oxygen
An oxygen concentrator takes in room air, removes nitrogen with a molecular sieve and delivers 87-96% oxygen at up to 5-10 L/min through a flowmeter, with no cylinders or pipeline. Public hospitals, BHUs/RHCs and home oxygen programmes in Pakistan buy them for pneumonia, COPD, newborn care and surge capacity. The key tender decisions are rated flow, outlet pressure, oxygen-purity monitoring and alarms, power use and back-up during load-shedding, and spare parts and filters to keep units running for 7 years.
What this equipment will do
- Patients: adults, children and newborns with low blood oxygen (pneumonia, COPD, newborn breathing problems), by nasal cannula or face mask, on wards, in primary care and at home (ISO 80601-2-69).
- Makes oxygen from room air: at least 87% oxygen (90-96% typical) at every setting up to at least 5 L/min, adjustable in 0.5 L/min steps.
- Higher flows for sicker adults: up to 8 or 10 L/min from one unit. (with the optional module)
- Two patients from one unit, through two separately adjustable outlets within the rated total flow. (with the optional module)
- Several children or newborns from one 10 L/min unit through a low-flow flowmeter stand with 4 outlets of 0-2 L/min. (with the optional module)
- Bubble CPAP for newborns and multi-outlet flowmeter stands over long tubing, using high outlet pressure (about 138 kPa / 20 psi). (with the optional module)
- Safety: oxygen purity monitored all the time with an alarm below 82%, a power-failure alarm that works without mains, a fault alarm and a fire-break at the outlet (IEC 60601-1-8).
- Runs 24 hours a day in rooms of 10-40 deg C, using not more than 70 W per L/min, with washable filters.
- Checking oxygen purity and flow at the outlet with a portable oxygen analyser. (with the optional module)
- Checking patients' SpO2 with a handheld pulse oximeter with adult, child and newborn probes (ISO 80601-2-61). (with the optional module)
- Oxygen continues during power cuts or breakdown from a back-up oxygen cylinder set. (with the optional module)
- Nebuliser treatments driven by the concentrator. (with the optional module)
Not covered: It has no internal battery and stops in a power cut; patients then need a back-up cylinder or a UPS.; It cannot fill oxygen cylinders or supply a hospital pipeline; that needs a PSA oxygen generation plant.; It gives only low-flow oxygen up to 10 L/min; high-flow nasal oxygen (up to 60 L/min) needs a high-flow nasal oxygen system, and breathing support needs a non-invasive ventilator or CPAP.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General basic safety and essential performance of medical electrical equipment, including electric shock, overheating and fire risks in an oxygen-rich unit.
- IEC 60601-1-2 Electromagnetic compatibility, so the concentrator and its oxygen sensor and alarms work near other ward equipment and mobile phones.
- IEC 60601-1-8 Alarm system design for low-oxygen, power-failure and fault alarms.
- IEC 60601-1-11 Extra requirements for equipment used in the home, relevant to home oxygen programmes and dusty, humid settings.
- ISO 80601-2-69 Particular standard for oxygen concentrator equipment: oxygen concentration and flow accuracy, low-oxygen alarm, outlet pressure and fire-break protection.
- ISO 14971 Risk management by the manufacturer, covering hazards such as low oxygen output, fire and power failure.
- ISO 13485 Quality management system of the legal manufacturer.
- IEC 62353 Recurrent and after-repair electrical safety testing of medical equipment, used at acceptance and preventive maintenance.
- WHO Technical specifications for oxygen concentrators (2015) WHO reference specification for selecting, buying and maintaining oxygen concentrators in low-resource settings.
- WHO-UNICEF Technical specifications and guidance for oxygen therapy devices (2019) Updated WHO-UNICEF minimum specifications for concentrators, analysers, flowmeter stands and pulse oximeters used together.
- DRAP Medical Devices Rules 2017 Legal requirement for registration/enlistment of medical devices imported or sold in Pakistan.
Common mistakes in tenders for this equipment
- Copying one brand's figures, such as exact 20 psi outlet pressure, a 0.5 L/min minimum flow or a fixed weight, which only one model meets. Use 'not less than' values and accept any design that meets the clinical need.
- Accepting '93% +/-3%' on the datasheet without measuring purity at the rated maximum flow on delivery. Measure every unit with a calibrated analyser after 30 minutes of running and reject units below 87%.
- Ignoring power: no back-up for load-shedding, no stabiliser for low voltage that burns compressors, and no budget for electricity, which for a 10 L/min unit running all day costs more each year than the unit itself.
- Buying dual-flow units and assuming each outlet gives the full rated flow, or using them for patients who need more than half the unit's flow.
- Not buying filters, sieve beds, compressor kits, service manuals or technician training, so units are abandoned after the first fault, as happened with many units bought during COVID-19.
- Specifying concentrators to drive ventilators, high-flow nasal cannula or anaesthesia machines; these need pipeline or cylinder oxygen at about 4 bar.
- Using adult flowmeters for newborns and small children instead of a low-flow flowmeter stand, leading to over- or under-dosing of oxygen.
- Accepting DRAP registration of a different model or of the brand only; the exact offered model must be covered.